Evaluation of a commercially available apparatus for measuring the acute phase protein serum amyloid A in horses
Authors:
- 13 September 2008
- journal article
- research article
- Published by Wiley in Veterinary Record
- Vol. 163 (11) , 327-330
Abstract
The aim of this study was to investigate the reliability of an immunoturbidometric assay for measuring the acute phase protein serum amyloid A ( saa ) in horses in clinical practice. The assay was compared to a previously validated assay, and overlap performance was assessed by measuring the concentration of saa in clinically healthy horses and horses with inflammatory and non‐inflammatory diseases. In pools of serum with low and high saa concentrations the assay's intra‐assay coefficients of variation were 11·7 per cent and 4·6 per cent, and its interassay coefficients of variation were 9·1 per cent and 5·6 per cent, respectively. Slight inaccuracies were observed, but they were negligible in comparison with the range of the saa response. The assay systematically underestimated the concentrations of saa in comparison with the results of the validated assay. The assay detected the expected difference in saa concentrations between the healthy and diseased horses.Keywords
This publication has 15 references indexed in Scilit:
- The acute phase protein serum amyloid A (SAA) as a marker of inflammation in horsesEquine Veterinary Education, 2007
- Method comparison in the clinical laboratoryVeterinary Clinical Pathology, 2006
- Evaluation of a commercially available human serum amyloid A (SAA) turbidometric immunoassay for determination of equine SAA concentrationsThe Veterinary Journal, 2006
- An Investigation of the Ability of the Glutaraldehyde Test to Distinguish between Acute and Chronic Inflammatory Disease in HorsesActa Veterinaria Scandinavica, 2005
- Dose Dependency and Individual Variability of the Lipopolysaccharide-Induced Bovine Acute Phase Protein ResponseJournal of Dairy Science, 2004
- A non-competitive chemiluminescence enzyme immunoassay for the equine acute phase protein serum amyloid A (SAA) – a clinically useful inflammatory marker in the horseVeterinary Immunology and Immunopathology, 1999
- Standardization of Diagnostic Assays for Animal Acute Phase Proteins1999
- Differential plasma clearance of murine acute-phase serum amyloid A proteins SAA1 and SAA2Biochemical Journal, 1997
- Evaluation of Serum Amyloid A Protein as an Acute-Phase Reactive Protein in Horses.Journal of Veterinary Medical Science, 1993
- Assessment of an immunoturbidimetric method for measuring equine serum haptoglobin concentrationsEquine Veterinary Journal, 1991