Design and field evaluation of a lateral flow cassette device for point-of-care bilirubin measurement
Open Access
- 8 August 2023
- journal article
- research article
- Published by Public Library of Science (PLoS) in PLOS Global Public Health
- Vol. 3 (8) , e0002262
Abstract
Neonatal jaundice is an important cause of morbidity and mortality worldwide, and neonates born in low and middle-income countries bear a disproportionate burden. We previously developed a low-cost, point-of-care system to measure total serum bilirubin (TSB) in neonates. This device was effective at detecting and monitoring jaundice; however, the disposable strips were difficult to produce at scale. Here, we report a new lateral flow cassette design, called BiliDx, that was produced at scale using traditional manufacturing techniques. We evaluated the performance of BiliDx at sites in Nigeria and Malawi. The lateral flow strip consists of plasma separation membranes, nitrocellulose, and a plastic cassette. We evaluated the performance of the strips and reader at two hospitals located in Nigeria and Malawi compared to reference standard TSB. We also assessed performance for samples with high direct bilirubin (DB) and high hematocrit (HCT). We collected 1,144 samples from 758 neonates (TSB ranged from 0.2 to 45.9 mg/dL). The mean bias of BiliDx measurements in the validation set was +0.75 mg/dL, and 95% limits of agreement were -2.57 to 4.07 mg/dL. The mean bias and limits of agreement were comparable for samples with HCT < 60% and HCT ≥ 60%, and for samples with low and intermediate DB levels; the samples with high DB levels had wider 95% limits of agreement (-4.50 to +3.03 mg/dL). Error grid analysis shows that 96.9% of samples measured with BiliDx would have resulted in the same clinical decision as the reference standard. This performance is comparable to previous results that used a handmade two-dimensional strip. Additionally, error grid analysis shows that all 20 samples with high DB levels would have resulted in the same clinical decision as the reference standard. This evaluation supports the use of BiliDx lateral flow cassettes to provide accurate point-of-care measurements in low-resource settings.Keywords
Funding Information
- John D. and Catherine T. MacArthur Foundation (17-1709-152484-100ANDCHANGE)
- Bill and Melinda Gates Foundation (INV-0077561)
- ELMA Foundation (19-F0012)
- Children's Investment Fund Foundation (R-1810-03159)
- Lemelson Foundation (A19-0120-007)
- Ting Tsung and Wei Fong Chao Family Foundation
- Sall Family Foundation
- Individual donors to NEST 360
This publication has 42 references indexed in Scilit:
- Predischarge Screening for Severe Neonatal Hyperbilirubinemia Identifies Infants Who Need PhototherapyThe Journal of Pediatrics, 2013
- Bilistick: A Low-Cost Point-of-Care System to Measure Total Plasma BilirubinNeonatology, 2012
- An approach to the management of hyperbilirubinemia in the preterm infant less than 35 weeks of gestationJournal of Perinatology, 2012
- Limitations and Opportunities of Transcutaneous Bilirubin MeasurementsPediatrics, 2012
- Hyperbilirubinemia in the Newborn Infant ≥35 Weeks’ Gestation: An Update With ClarificationsPediatrics, 2009
- Hyperbilirubinemia and Transcutaneous BilirubinometryClinical Chemistry, 2009
- Visual assessment of jaundice in term and late preterm infantsArchives of Disease in Childhood - Fetal and Neonatal Edition, 2009
- Evaluating Clinical Accuracy of Systems for Self-Monitoring of Blood GlucoseDiabetes Care, 1987
- Molar Absorption Coefficients for Bilirubins in Adult and Infant Serum with Determination of an Isosbestic PointScandinavian Journal of Clinical and Laboratory Investigation, 1972
- Advancement of Dermal Icterus in the Jaundiced NewbornArchives of Pediatrics & Adolescent Medicine, 1969