Technique factors and their relationship to radiation dose in pendant geometry breast CT
Authors:
- John M. BooneJBJohn M. Boone
- Alexander L. C. KwanAKAlexander L. C. Kwan
- J. Anthony SeibertJSJ. Anthony Seibert
- Nikula ShahNSNikula Shah
- Karen K. LindforsKLKaren K. Lindfors
- Thomas R. NelsonTNThomas R. Nelson
- 22 November 2005
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 32 (12) , 3767-3776
Abstract
The use of breast computed tomography (CT) as an alternative to mammography in some patients is being studied at several institutions. However, the radiation dosimetry issues associated with breast CT are markedly different than in the case of mammography. In this study, the spectral properties of an operational breast CT scanner were characterized both by physical measurement and computer modeling of the kVp‐dependent spectra, from 40 to 110 kVp (Be window W anode with 0.30 mm added Cu filtration). Previously reported conversion factors, normalized glandular dose for CT‐DgN(ct), derived from Monte Carlo methods, were used in concert with the output spectra of the breast scanner to compute the mean glandular dose to the breast based upon different combinations of x‐ray technique factors (kVp and mAs). The mean glandular dose (MGD) was measured as a function of the compressed breast thickness (2–8 cm) and three different breast compositions (0%, 50%, and 100% glandular fractions) in four clinical mammography systems in our institution. The average MGD from these four systems was used to compute the technique factors for breast CT systems that would match the two‐view mammographic dose levels. For a 14 cm diameter breast (equivalent to a 5 cm thick compressed breast in mammography), air kerma levels at the breast CT scanner's isocenter (468 mm from the source) of 4.4, 6.4, and 9.0 mGy were found to deliver equivalent mammography doses for 0%, 50%, and 100% glandular breasts (respectively) at 80 kVp. At 80 kVp (where air kerma was at the isocenter), 57 mAs (integrated over the entire scan) was required to match the mammography dose for a 14 cm 50% glandular breast. At 50 kVp, 360 mAs is required to match mammographic dose levels. Tables are provided for both air kerma at the isocenter and mAs for 0%, 50%, and 100% glandular breasts. Other issues that impact breast CT technique factors are also discussed.Keywords
Funding Information
- National Cancer Institute (R01 CA 89260)
- California Breast Cancer Research Program (7EB‐0075)
This publication has 10 references indexed in Scilit:
- Mammography “dose creep”: Causes and solutions1Academic Radiology, 2004
- Factors affecting increasing radiation dose for mammography in North Carolina from 1997 through 2001: An analysis of food and drug administration annual surveys1Academic Radiology, 2004
- A comprehensive analysis of DgNCT coefficients for pendant-geometry cone-beam breast computed tomographyMedical Physics, 2004
- Dedicated Breast CT: Radiation Dose and Image Quality EvaluationRadiology, 2001
- Molybdenum, rhodium, and tungsten anode spectral models using interpolating polynomials with application to mammographyMedical Physics, 1997
- An accurate method for computer‐generating tungsten anode x‐ray spectra from 30 to 140 kVMedical Physics, 1997
- A survey of fluoroscopic exposure rates: AAPM Task Group No. 11 ReportMedical Physics, 1993
- Specific Value of Computed Tomographic Breast Scanner (CT/M) in Diagnosis of Breast DiseasesRadiology, 1979
- Computed tomographic evaluation of the breastAmerican Journal of Roentgenology, 1978
- Computed Tomography of the BreastRadiology, 1977