Excimer Laser Calibration System

Authors:
  • ERErik V Rencs
    Erik V Rencs
  • JCJames L Cambier
    James L Cambier
  • DHDeborah Hall
    Deborah Hall
  • DADimitri T Azar
    Dimitri T Azar
  • WSWalter J Stark
    Walter J Stark

Abstract
BACKGROUND: Excimer laser photoablation for refractive and therapeutic keratectomies has been demonstrated to be feasible and practicable. However, corneal laser ablations are not without problems, including the delivery and maintenance of a homogeneous beam. We have developed an excimer laser calibration system capable of characterizing a laser ablation profile. METHODS: Beam homogeneity is determined by the analysis of a polymethylmethacrylate (PMMA)based thin-film using video capture and image processing. The ablation profile is presented as a colorcoded map. Interpolation of excimer calibration system analysis provides a three-dimensional representation of elevation profiles that correlates with two-dimensional scanning profilometry. Excimer calibration analysis was performed before treating a monkey undergoing phototherapeutic keratectomy and two human subjects undergoing myopic spherocylindrical photorefractive keratectomy. Excimer calibration analysis was performed before and after laser refurbishing. RESULTS: Laser ablation profiles in PMMA are resolved by the excimer calibration system to .006 µ??/pulse. Correlations with ablative patterns in a monkey cornea were demonstrated with preoperative and postoperative keratometry using corneal topography, and two human subjects using video-keratography. Excimer calibration analysis predicted a central-steep-island ablative pattern with the VISX Twenty/Twenty laser, which was confirmed by corneal topography immediately postoperatively and at 1 week after reepithelialization in the monkey. Predicted central steep islands in the two human subjects were confirmed by videokeratography at 1 week and at 1 month. Subsequent technical refurbishing of the laser resulted in a beam with an overall increased ablation rate measured as microns/pulse with a donut ablation profile. A patient treated after repair of the laser electrodes demonstrated no central island. CONCLUSIONS: This excimer laser calibration system can precisely detect laser-beam ablation profiles. The calibration system correctly predicted central islands after excimer photoablation in a treated monkey cornea and in two treated human subjects. Detection of excimer-laser-beam ablation profiles may be useful for precise calibration of excimer lasers before human photorefractive and therapeutic surgery. [J Refract Surg. 1996;12:401-411.]

This publication has 15 references indexed in Scilit: