Automatic patient alignment system using 3D ultrasound
Authors:
- Michael FiglMFMichael Figl
- Rainer HoffmannRHRainer Hoffmann
- Wolfgang BirkfellnerWBWolfgang Birkfellner
- Johann HummelJHJohann Hummel
- Markus StockMSMarkus Stock
- Dietmar GeorgDietmar Georg
- Gregor GoldnerGGGregor Goldner
- 20 March 2013
- journal article
- clinical trial
- Published by Wiley in Medical Physics
- Vol. 40 (4) , 041714
Abstract
Purpose:: Recent developments in radiation therapy such as intensity modulated radiotherapy (IMRT) or dose painting promise to provide better dose distribution on the tumor. For effective application of these methods the exact positioning of the patient and the localization of the irradiated organ and surrounding structures is crucial. Especially with respect to the treatment of the prostate, ultrasound (US) allows for differentiation between soft tissue and was therefore applied by various repositioning systems, such as BAT or Clarity. The authors built a new system which uses 3D US at both sites, the CT room and the intervention room and applied a 3D/3D US/US registration for automatic repositioning.Methods:: In a first step the authors applied image preprocessing methods to prepare the US images for an optimal registration process. For the 3D/3D registration procedure five different metrics were evaluated. To find the image metric which fits best for a particular patient three 3D US images were taken at the CT site and registered to each other. From these results an US registration error was calculated. The most successful image metric was then applied for the US/US registration process. The success of the whole repositioning method was assessed by taking the results of an ExacTrac system as golden standard.Results:: The US/US registration error was found to be 2.99 ± 1.54 mm with respect to the mutual information metric by Mattes (eleven patients) which revealed to be the most suitable of the assessed metrics. For complete repositioning chain the error amounted to 4.15 ± 1.20 mm (ten patients).Conclusions:: The authors developed a system for patient repositioning which works automatically without the necessity of user interaction with an accuracy which seems to be suitable for clinical application.Keywords
Funding Information
- Austrian Science Fund (L625‐N15)
- Oesterreichische Nationalbank (14525)
- Austrian Science Fund (L625-N15)
- Oesterreichische Nationalbank (14525)
This publication has 26 references indexed in Scilit:
- Stability of Markers Used for Real-Time Tumor Tracking After Percutaneous Intrapulmonary PlacementInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2011
- Evaluation of Inter-fraction Prostate Motion using Kilovoltage Cone Beam Computed Tomography during RadiotherapyClinical Oncology, 2011
- Effects of Setup Errors and Shape Changes on Breast RadiotherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2011
- Breast Patient Setup Error Assessment: Comparison of Electronic Portal Image Devices and Cone-Beam Computed Tomography Matching ResultsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2010
- ExacTrac X-ray 6 degree-of-freedom image-guidance for intracranial non-invasive stereotactic radiotherapy: Comparison with kilo-voltage cone-beam CTRadiotherapy and Oncology, 2009
- Variation in target and rectum dose due to prostate deformation: an assessment by repeated MR imaging and treatment planningPhysics in Medicine & Biology, 2008
- Accuracy of Ultrasound-Based (BAT) Prostate-Repositioning: A Three-Dimensional On-Line Fiducial-Based Assessment With Cone-Beam Computed TomographyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2008
- Intra-subject elastic registration of 3D ultrasound imagesMedical Image Analysis, 2006
- Multi-modal volume registration by maximization of mutual informationMedical Image Analysis, 1996
- Closed-form solution of absolute orientation using unit quaternionsJournal of the Optical Society of America A, 1987