Fast and compact scatterometer for ex vivo qualification of low-scattering biological samples.
- 1 July 2026
- journal article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 97 (7)
Abstract
We present the design, calibration, qualification, and application of a fast and compact scatterometer designed for the characterization of ex vivo human corneas. As an alternative to standard qualification processes and in order to meet the current need for improved reliability of graft qualification, this instrument provides a non-invasive, label-free, and contact-free quantitative means of characterizing corneal grafts, based on back-scattered light diffusometry. The instrument is shown to be capable of detecting BRDF (bidirectional reflectance distribution function) variations of 1.4 × 10-4 sr-1 with a detectivity threshold of 2.0 × 10-3 sr-1, with an overall dynamics >27 dB for the total integrated scattering parameter used for graft quality assessment. We also demonstrate the feasibility of integrating this new technique into tissue bank qualification processes through controlled test experiments reported here.Keywords
This publication has 58 references indexed in Scilit:
- Corneal Endothelial Cell Density Loss after Glaucoma Surgery Alone or in Combination with Cataract SurgeryOphthalmology, 2022
- Reliability and efficiency of corneal thickness measurements using sterile donor tomography in the eye bankCell and Tissue Banking, 2021
- Donor demographics and factors affecting corneal utilisation in Eye Bank of North IndiaInternational Ophthalmology, 2021
- Global Survey of Corneal Transplantation and Eye BankingJAMA Ophthalmology, 2016
- Non-invasive measurement of transparency, arcus senilis, and scleral rim diameter of corneas during eye bankingCell and Tissue Banking, 2013
- Repeatability and validity of lens densitometry measured with Scheimpflug imagingJournal of Cataract & Refractive Surgery, 2009
- An evaluation of image quality and accuracy of eye bank measurement of donor cornea endothelial cell density in the Specular Microscopy Ancillary StudyOphthalmology, 2005
- Corneal endothelium and postoperative outcomes 15 years after penetrating keratoplastyAmerican Journal of Ophthalmology, 2005
- Grading of corneal transparencyContact Lens and Anterior Eye, 2004
- Why the cornea swellsJournal of Theoretical Biology, 1971