DC-Amplifier-Input-Offset-Voltage Control in a Constant-Temperature Thermoresistive-Sensor-Measurement Instrument
Authors:
- Raimundo Carlos Silvrio FreireRFRaimundo Carlos Silvrio Freire
- Jos Felcio da SilvaJSJos Felcio da Silva
- Sebastian Yuri Cavalcanti CatundaSCSebastian Yuri Cavalcanti Catunda
- Glauco FontgallandGFGlauco Fontgalland
- 15 May 2007
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Instrumentation and Measurement
- Vol. 56 (3) , 778-783
Abstract
In this paper, the effect of dc amplifier input offset voltage on the static and dynamic performances of a generic constant-temperature thermoresistive-sensor-feedback circuit (CTC) is reviewed. A negative-feedback automatic control circuit for the input offset voltage is proposed in order to assure a stable operation of the CTC and to obtain optimal response time. Experimental evaluation results of the proposed circuit performance are presentedKeywords
This publication has 6 references indexed in Scilit:
- Frequency Domain Analysis of an Electrical Substitution RadiometerJournal of Solar Energy Engineering, 1999
- Error in the thermoresistive sensor-based feedback measurement configuration due to DC amplifier offset voltageIEEE Transactions on Instrumentation and Measurement, 1999
- Dynamic Response of a Feedback Thermoresistive Electrical Substitution PyranometerJournal of Solar Energy Engineering, 1998
- A constant temperature hot-wire anemometerJournal of Physics E: Scientific Instruments, 1987
- A study of the constant-temperature hot-wire anemometerJournal of Fluid Mechanics, 1971
- Feedback Control Theory for Constant-Temperature Hot-Wire AnemometersReview of Scientific Instruments, 1967