Optimized distortion correction technique for echo planar imaging
Open Access
- 28 February 2001
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 45 (3) , 525-528
Abstract
A new phase‐shifted EPI pulse sequence is described that encodes EPI phase errors due to all off‐resonance factors, including Bo field inhomogeneity, eddy current effects, and gradient waveform imperfections. Combined with the previously proposed multichannel modulation postprocessing algorithm (Chen and Wyrwicz, MRM 1999;41:1206–1213), the encoded phase error information can be used to effectively remove geometric distortions in subsequent EPI scans. The proposed EPI distortion correction technique has been shown to be effective in removing distortions due to gradient waveform imperfections and phase gradient‐induced eddy current effects. In addition, this new method retains advantages of the earlier method, such as simultaneous correction of different off‐resonance factors without use of a complicated phase unwrapping procedure. The effectiveness of this technique is illustrated with EPI studies on phantoms and animal subjects. Implementation to different versions of EPI sequences is also described. Magn Reson Med 45:525–528, 2001.Keywords
This publication has 10 references indexed in Scilit:
- fMRI of visual system activation in the conscious rabbitMagnetic Resonance in Medicine, 2000
- Correction for EPI distortions using multi-echo gradient-echo imagingMagnetic Resonance in Medicine, 1999
- Correction of off resonance‐related distortion in echo‐planar imaging using EPI‐based field mapsMagnetic Resonance in Medicine, 1998
- Mapping of lateral geniculate nucleus activation during visual stimulation in human brain using fMRIMagnetic Resonance in Medicine, 1998
- Reduction of geometric and intensity distortions in echo‐planar imaging using a multireference scanMagnetic Resonance in Medicine, 1997
- Multiple-shot echo-planar imagingConcepts in Magnetic Resonance, 1997
- Correction for geometric distortion in echo planar images from B0 field variationsMagnetic Resonance in Medicine, 1995
- Accurate T1 determination from inversion recovery images: Application to human brain at 4 TeslaMagnetic Resonance in Medicine, 1994
- Image reconstruction for echo planar imaging with nonequidistant k‐space samplingMagnetic Resonance in Medicine, 1992
- Multi-planar image formation using NMR spin echoesJournal of Physics C: Solid State Physics, 1977