Abstract
This paper present a new laparoscopic surgical setup (minimally invasive surgery, MIS) with spherical serial surgical robot. The robot was developed at the PPRIME Institute in the ROBIOSS team from Poitiers University. Based on laparoscopic surgery, specification for workspace and constraints are outlined and the architecture selected for the surgical robots is presented and discussed. Analytical models of the inverse and forward kinematics are provided together with a detailed analysis of mobility and constraints. The proposed approach is based on motion capture of an expert gestures during which an anastomosis technique has to be performed. The evaluation of the workspace is based on the use of the Vicon Nexus motion capture system. The medical gestures were analyzed in term of position and velocities; the result has been used in the definition of the kinematics specifications of the proposed manipulator. The effective workspace size is determined through an experimental study on a simulator. The spherical serial robot (3R-T) has been selected because of its characteristics meeting the constraint requirements. The robot and its environment are described.Keywords
This publication has 12 references indexed in Scilit:
- New trends in minimally invasive urological surgery. What is beyond the robot?World Journal of Urology, 2010
- Design requirements for a new robot for minimally invasive surgeryIndustrial Robot: the international journal of robotics research and application, 2004
- Laparoscopic surgery update for gastrointestinal malignancyJournal of Gastroenterology, 2004
- Kinematic error correction for minimally invasive surgical robots2004
- A six degree of freedom micromanipulator for ophthalmic surgery2002
- The Black Falcon: a teleoperated surgical instrument for minimally invasive surgery2002
- Recent advances in minimal access surgeryBMJ, 2002
- A Surface-Matching Technique for Robot-Assisted RegistrationComputer Aided Surgery, 2001
- Retargetting motion to new characters1998
- ISB recommendations for standardization in the reporting of kinematic dataJournal of Biomechanics, 1995