Lateral cavity acoustic transducer as an on-chip cell/particle microfluidic switch
Authors:
- 10 November 2011
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in Lab on a Chip
- Vol. 12 (1) , 139-145
Abstract
A novel on-chip microfluidic switch is demonstrated that utilizes the acoustic microstreaming generated by an oscillating air–liquid interface to switch cells/particles into bifurcating microchannels. The air–liquid interface of the Lateral Cavity Acoustic Transducers (LCATs) can be actuated by an external acoustic energy source causing the interface to oscillate. The oscillating interface results in the generation of vortex-like microstreaming flow within a localized region of the surrounding liquid. This streaming was utilized here to deflect cells/particles into a collection outlet. It was demonstrated that the switching zone could be controlled by varying the actuation time of the LCAT. An LCAT based microfluidic switch is capable of achieving theoretical switching rates of 800 cells/particles per second. It was also demonstrated that K562 cells could be switched into a collection channel with cell viability comparable to that of controls as determined by Trypan blue exclusion assay.Keywords
This publication has 32 references indexed in Scilit:
- Cells on chipsNature, 2006
- Dielectrophoretic manipulation of drops for high-speed microfluidic sorting devicesApplied Physics Letters, 2006
- Microfluidic sorting of mammalian cells by optical force switchingNature Biotechnology, 2005
- An Integrated Microfluidic System for Reaction, High-Sensitivity Detection, and Sorting of Fluorescent Cells and ParticlesAnalytical Chemistry, 2003
- Integrating advanced functionality in a microfabricated high-throughput fluorescent-activated cell sorterLab on a Chip, 2003
- An Integrated Microfabricated Cell SorterAnalytical Chemistry, 2002
- Monolithic Microfabricated Valves and Pumps by Multilayer Soft LithographyScience, 2000
- A microfabricated fluorescence-activated cell sorterNature Biotechnology, 1999
- Rapid prototyping of microfluidic switches in poly(dimethyl siloxane) and their actuation by electro-osmotic flowJournal of Micromechanics and Microengineering, 1999
- Automated single-cell manipulation and sorting by light trappingApplied Optics, 1987