Estimating the Magnitude and Phase Response of a 50 GHz Sampling Oscilloscope Using the "Nose-to-Nose" Method
Authors:
- 1 June 2000
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- Vol. 37, 1-8
Abstract
We describe estimation of the magnitude and phase response of a sampling oscilloscope with 50 GHz bandwidth using the nose-to-nose method. The measurements are corrected for the non-ideal properties of the oscilloscope and calibration apparatus, including mismatch and time-base distortion, drift, and jitter. The mean and standard deviation of repeated measurements of an ensemble of three oscilloscope samplers are reported, along with attempts to verify the magnitude calibration using a swept sine-wave method.Keywords
This publication has 8 references indexed in Scilit:
- Realistic sampling-circuit model for a nose-to-nose simulation2002
- Nose-to-Nose Response of a 20-GHz Sampling Circuit2000
- Analysis of Interconnection Networks and Mismatch in the Nose-to-Nose Calibration2000
- Least-squares estimation of time-base distortion of sampling oscilloscopesIEEE Transactions on Instrumentation and Measurement, 1999
- Signal reconstruction for non-equidistant finite length sample sets: a "KIS" approachIEEE Transactions on Instrumentation and Measurement, 1998
- Broadband sampling oscilloscope characterization with the "Nose-to-Nose" calibration procedure: a theoretical and practical analysisIEEE Transactions on Instrumentation and Measurement, 1995
- Individual characterization of broadband sampling oscilloscopes with a nose-to-nose calibration procedureIEEE Transactions on Instrumentation and Measurement, 1994
- The Measurement and Deconvolution of Time Jitter in Equivalent-Time Waveform SamplersIEEE Transactions on Instrumentation and Measurement, 1983