Reproducibility of a Triaxial Seismic Accelerometer (DynaPort)
Authors:
- Vincent T. VAN HeesVHVincent T. VAN Hees
- Sander M. SlootmakerSSSander M. Slootmaker
- Gert DE GrootGGGert DE Groot
- Willem VAN MechelenWMWillem VAN Mechelen
- Rob C. VAN LummelRLRob C. VAN Lummel
- 1 April 2009
- journal article
- research article
- Published by Wolters Kluwer Health in Medicine & Science in Sports & Exercise
- Vol. 41 (4) , 810-817
Abstract
To examine the reproducibility of a triaxial seismic accelerometer under controlled conditions and real-life conditions. Instrumental reproducibility was examined using a shaker device. The accelerometers (DynaPort MiniMod; McRoberts B.V., The Hague, The Netherlands) were shaken at four frequencies (0.8, 2.1, 3.6, and 4.6 Hz) in x- and y-directions. The magnitude of acceleration ranged from 0 to 1.277g. Additionally, reproducibility under real-life conditions was examined in 55 adolescents (12-17 yr), with the accelerometer attached to the lower back. Each subject walked four short walking trials on level ground at preferred speed. To make this setting meet real-life conditions, we detached and reattached the accelerometer between trials 2 and 3. Detachment of accelerometer between trials 2 and 3 was done by either the same researcher or different researchers (four in total). Intra- and interobserver reproducibility were calculated. Intra- and interinstrumental intraclass correlation coefficients (ICC) were 0.99 for both x- and y-directions. The intrainstrumental coefficients of variance (CoV) were lower than 1.13%. The interinstrumental CoV were lower than 1.37%. Intraobserver ICC was 0.97, and interobserver ICC was 0.88. The reproducibility of the accelerometer is high under the controlled conditions of a shaker device and in walking at preferred speed.Keywords
This publication has 24 references indexed in Scilit:
- Detection of walking periods and number of steps in older adults and patients with Parkinson's disease: accuracy of a pedometer and an accelerometry-based methodAge and Ageing, 2008
- Role of Low Energy Expenditure and Sitting in Obesity, Metabolic Syndrome, Type 2 Diabetes, and Cardiovascular DiseaseDiabetes, 2007
- Technical Reliability Assessment of Three Accelerometer Models in a Mechanical SetupMedicine & Science in Sports & Exercise, 2006
- Accelerometry based assessment of gait parameters in childrenGait & Posture, 2006
- Occupational, Commuting, and Leisure-Time Physical Activity in Relation to Total and Cardiovascular Mortality Among Finnish Subjects With Type 2 DiabetesCirculation, 2004
- Long-term mobility monitoring of older adults using accelerometers in a clinical environmentClinical Rehabilitation, 2004
- Exercise, body mass index, caloric intake, and cardiovascular mortalityAmerican Journal of Preventive Medicine, 2003
- Validity of 10 Electronic Pedometers for Measuring Steps, Distance, and Energy CostMedicine & Science in Sports & Exercise, 2003
- Reexamination of Validity and Reliability of the CSA Monitor in Walking and RunningMedicine & Science in Sports & Exercise, 2003
- Establishing a standard definition for child overweight and obesity worldwide: international surveyBMJ, 2000