Abstract
Background: Strain gauge analysis and finite element (FE) simulations were used to estimate the bone strain and micromovement at the bone–implant interface (BII) for platform switching and different diameters of a single, immediately loaded mandibular implant.Methods: Four models were created, including 5‐mm‐diameter implants assembled with abutments that were 5 or 4 mm in diameter on bonded (delay‐loading treatment) and contact (immediate‐loading treatment) BIIs; a model with an implant diameter of 3.75 mm was also analyzed. Vertical and lateral loads of 130 N were applied to all models.Results: During lateral loading, the strains were highly concentrated on one side of the mandible in the experimental and validation FE models. Bone strains were reduced by Conclusions: Bone strain was reduced more by increasing the diameter of the implant than by using platform switching in the immediately loaded implant. However, neither a wide implant nor platform switching reduced micromotion at the BII for enhancing implant stability.