The Swanson Ulnar Head Prosthesis for Post-Traumatic Disorders of the Distal Radio-Ulnar Joint
Authors:
- 1 December 1992
- journal article
- Published by SAGE Publications in Journal of Hand Surgery (European Volume)
- Vol. 17 (6) , 682-688
Abstract
20 Swanson ulnar head prostheses inserted for post-traumatic disorders of the distal radio-ulnar joint were reviewed at a mean of 44.2 months post-operatively (range 12–104 months). The indication for surgery was painful loss of forearm rotation with or without ulno-carpal impingement and ulno-carpal instability. Symptoms in 16 patients were the result of malunion following severe distal radial fractures (two Frykman 7 fractures and 14 Frykman 8 fractures), and in the remainder were due to unreduced dislocations in the distal radio-ulnar joint (two patients) or unsatisfactory Darrach procedures (two patients). 70% of patients achieved excellent or good clinical results despite the fact that radiographs showed bone resorption in all cases, tilting of the prosthesis in 40% and implant fractures in 15%.Keywords
This publication has 8 references indexed in Scilit:
- A critical analysis of Swanson ulnar head replacement arthroplasty: Rheumatoid versus nonrhematoidThe Journal of Hand Surgery, 1990
- Resection of the distal ulna with and without implant arthroplasty in rheumatoid arthritisThe Journal of Hand Surgery, 1986
- The long-term results of Swanson silicone rubber interpositional wrist arthroplastyThe Journal of Hand Surgery, 1986
- Distal radioulnar joint arthroplasty: The hemiresection-interposition techniqueThe Journal of Hand Surgery, 1985
- Ulnar impingement syndromeThe Journal of Bone and Joint Surgery. British volume, 1985
- Results from Resection of the Distal end of the Ulna After Fractures of the Lower end of the RadiusScandinavian Journal of Plastic and Reconstructive Surgery, 1982
- Indications for and Results With the Swanson Distal Ulnar ProsthesisSouthern Medical Journal, 1976
- Implant Arthroplasty for Disabilities of the Distal Radioulnar Joint Use of a Silicone Rubber Capping Implant Following Resection of the Ulnar HeadOrthopedic Clinics of North America, 1973