Co-Oximetry Interference by Hemoglobin-Based Blood Substitutes
Authors:
- 1 April 2001
- journal article
- cardiovascular anesthesia
- Published by Wolters Kluwer Health in Anesthesia & Analgesia
- Vol. 92 (4) , 863-869
Abstract
The blood substitutes now being developed from molecularly modified hemoglobin interfere with a wide variety of clinical analyzers, but their effects on cooximeters are unknown. Therefore, we investigated the effects of five hemoglobin-based blood substitutes on the measurements of eight different oximeters and cooximeters: the AVL Omni 6, the AVOXimeters 1000 and 4000, the Ciba Corning (now Bayer) CC270 CO-Oximeter, the Instrumentation Laboratory Synthesis 35, the IL482 and IL682 CO-Oximeters, and the Radiometer OSM3 Hemoximeter. The five blood substitutes in this study were obtained from Apex Bioscience (Research Triangle Park, NC), Baxter Healthcare Corp. (Deerfield, IL), Biopure Corp. (Cambridge, MA), Hemoglobin Therapeutics, and Hemosol, Inc. (Etobicoke, Ontario, Canada). A cooximeter control was used to compare the eight different instruments' measurements on unaltered human hemoglobin. The instruments yielded measurements of total hemoglobin concentration in undiluted blood substitutes that were generally not more variable than those on the control material. By contrast, when compared with readings on controls, the test instruments yielded measurements of the fractional concentrations of oxy-, deoxy-, carboxy-, and methemoglobin that showed greater instrument-to-instrument disparities and larger standard deviations about the all-instrument means. In some cases, the interference was even more obvious: five of six cooximeters gave negative carboxyhemoglobin readings on one particular product. Our findings indicate that the instruments will give less accurate but clinically useful measurements in the presence of these hemoglobin-based blood substitutes.We investigated the effects of five hemoglobin-based blood substitutes on the measurements of eight different cooximeters. Some blood substitutes caused obvious interference, such as negative carboxyhemoglobin readings; however, the findings indicate that cooximeters will generally give less accurate but clinically useful measurements in the presence of the hemoglobin-based blood substitutes that were tested.Keywords
This publication has 7 references indexed in Scilit:
- A Second-Generation Blood Substitute (Perfluorodichlorooctane Emulsion) Generates Spurious Elevations in Platelet Counts from Automated Hematology AnalyzersAnesthesia & Analgesia, 2000
- Hemoglobin Solutions Come of AgeAnesthesiology, 2000
- Randomized Trial of Diaspirin Cross-linked Hemoglobin Solution as an Alternative to Blood Transfusion after Cardiac SurgeryAnesthesiology, 2000
- Effect of the Hemoglobin-Based Oxygen Carrier HBOC-201 on Laboratory Instrumentation: Cobas Integra, Chiron Blood Gas Analyzer 840, Sysmex™SE-9000 and BCTClinical Chemistry and Laboratory Medicine (CCLM), 1999
- Evaluation of the Avoximeter: precision, long-term stability, linearity, and use without heparinJournal of Clinical Monitoring and Computing, 1997
- Near-patient measurements of methemoglobin, oxygen saturation, and total hemoglobinCritical Care Medicine, 1995
- Evaluation of two oximeters for use in cardiac catheterization laboratoriesCatheterization and Cardiovascular Diagnosis, 1993