ADAPTIVE REAL-TIME CARDIAC MRI USING PARADISE: VALIDATION BY THE PHYSIOLOGICALLY IMPROVED NCAT PHANTOM
Authors:
- Behzad SharifBSBehzad Sharif
- Yoram BreslerYBYoram Bresler
- 1 April 2007
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- No. 19457928,p. 1020-1023
Abstract
Patient-adaptive reconstruction and acquisition dynamic imaging with sensitivity encoding (PARADISE) is a dynamic MR imaging scheme that optimally combines parallel imaging and model-based adaptive acquisition. In this work, we propose the application of PARADISE to real-time cardiac MRI. We introduce a physiologically improved version of a realistic four-dimensional cardiac-torso (NCAT) phantom, which incorporates natural beat-to-beat heart rate and motion variations. Cardiac cine imaging using PARADISE is simulated and its performance is analyzed by virtue of the improved phantom. Results verify the effectiveness of PARADISE for high resolution un-gated real-time cardiac MRI and its superiority over conventional acquisition methods.Keywords
This publication has 4 references indexed in Scilit:
- Spatio-temporal modeling and minimum redundancy adaptive acquisition in dynamic MRI2003
- SENSE: Sensitivity encoding for fast MRIMagnetic Resonance in Medicine, 1999
- A realistic spline-based dynamic heart phantomIEEE Transactions on Nuclear Science, 1999
- Lattice-theoretic analysis of time-sequential sampling of spatiotemporal signals. IIEEE Transactions on Information Theory, 1997