Hyperconcentrated platelets stored in additive solution: aspects on productivity and in vitro quality

Authors:

Abstract
Background and Objectives  New platelet (PLT) additive solutions (PASs) allow a plasma carryover of < 30% in PLT concentrates. This implicates the need to collect apheresis PLT concentrates at very high PLT concentrations: so‐called dry PLTs (DPs). We used the TRIMA, with software version 4 (TRIMA V4), to collect such DPs and investigated the in vitro quality of these PLTs when stored in the new modified PAS‐III (PAS‐IIIM).Materials and Methods  TRIMA V4 was programmed to collect 6·0 × 1011 PLTs at a concentration of 5000 × 103 PLTs/µl. Two DPs were pooled, split into four equal parts and diluted to obtain secondary pools (SPs) consisting of 70% PAS‐III/30% plasma, 70% PAS‐IIIM/30% plasma, 80% PAS‐IIIM/20% plasma or 100% plasma. In vitro testing was performed on days 0, 1, 5 and 7. Collection efficiency (CE), collection rate (CR) and PLT yield were calculated for each donation.Results  Thirty‐two runs with TRIMA V4 were performed, collecting 6·58 ± 0·74 × 1011 PLTs at a concentration of 4255 ± 914 × 103/µl in 99 ± 19·9 min, resulting in a CE of 65·3 ± 8·2% and a CR of 6·92 ± 1·6 × 109 PLTs/min. On day 0, 34–37% of the PLTs in the units prepared for storage were already activated. PLTs stored in 70% or 80% PAS‐IIIM showed superior in vitro quality compared to PLTs stored in PAS‐III.Conclusions  TRIMA V4 is a suitable device for the collection of DPs. Nevertheless, improvements are desirable to further increase the ability to concentrate PLTs at very high levels. The storage of apheresis‐derived PLTs in PAS III‐M is a very promising approach, even at a plasma carryover of < 30%.