DETERMINATION OF MAGNESIUM BY THE TITAN YELLOW AND AMMONIUM PHOSPHATE METHODS
Author:
- F. W. HeatonFHF. W. Heaton
Open Access
- 1 July 1960
- journal article
- research article
- Published by BMJ in Journal of Clinical Pathology
- Vol. 13 (4) , 358-360
Abstract
The ammonium phosphate and Titan yellow methods for determining magnesium have been investigated, and improvements to both techniques are presented. The Titan yellow method is shown to be subject to interference which makes it unsuitable for the analysis of blood cells, food, and faecal materials, and may also affect its reliability with pathological samples of serum and urine. The relative insusceptibility to interference of the ammonium phosphate method makes it more reliable than the Titan yellow method with pathological specimens, and slight modifications permit the use of the same basic procedure with blood, urine, and solutions of food or faecal ash.Keywords
This publication has 6 references indexed in Scilit:
- STUDIES OF THE RENAL EXCRETION OF MAGNESIUM IN MAN*†Journal of Clinical Investigation, 1959
- Comment on “The Estimation of Magnesium in Serum Using Titan Yellow”Journal of Clinical Pathology, 1957
- On the Determination of Magnesium in Serum and Urine by the Titan Yellow MethodScandinavian Journal of Clinical and Laboratory Investigation, 1957
- The Estimation of Magnesium in Serum using Titan YellowJournal of Clinical Pathology, 1956
- MICROESTIMATION OF MAGNESIUM IN BODY FLUIDSJournal of Biological Chemistry, 1951
- THE COLORIMETRIC DETERMINATION OF PHOSPHORUSJournal of Biological Chemistry, 1925