This paper presents a feasibility study of the application of giant magneto resistive (GMR) sensors in detecting motion of slow moving fluids. A motivating application for the proposed effort is the development of a smart catheter capable of monitoring the amount of body fluid drained from the ventricles of the brain. Microfabricated ferromagnetic flaps are used to detect motion of the surrounding fluid. The deflection of the flaps is detected by an ultrasensitive GMR sensor placed outside of the lumen of the catheter. Numerical and experimental results are provided demonstrating a resolution of 1.4 mL/h. Numerical analysis of the fluid past the sensing element show an optimal hinge length of the flexible flaps, as well as a significant increase in sensitivity with reduction of the by-pass gap to ~50 μm. The effect of electro-magnetic interference and other sources of low-frequency noise (drift) has also been investigated. The results from the study are used to derive a set of design rules that may lead to the successful development of a smart catheter.Keywords
Funding Information
- National Science Foundation (1446098)
- Rosztoczy Foundation
This publication has 24 references indexed in Scilit:
- A self-bended piezoresistive microcantilever flow sensor for low flow rate measurementSensors and Actuators A: Physical, 2010
- Miniaturized thermal flow sensor with planar-integrated sensor structures on semicircular surface channelsSensors and Actuators A: Physical, 2008
- VENTRICULOPERITONEAL SHUNT COMPLICATIONS IN CALIFORNIANeurosurgery, 2007
- Microfluidic flow rate detection based on integrated optical fiber cantileverLab on a Chip, 2007
- A double heater integrated gas flow sensor with thermal feedbackSensors and Actuators A: Physical, 2005
- PCB-integrated metallic thermal micro-actuatorsSensors and Actuators A: Physical, 2003
- Predicting shunt failure on the basis of clinical symptoms and signs in childrenJournal of Neurosurgery, 2001
- Air temperature fluctuations in roomsBuilding and Environment, 1997
- Giant Magnetoresistance of (001)Fe/(001)Cr Magnetic SuperlatticesPhysical Review Letters, 1988
- Radio Telemetry for the Measurement of Intracranial PressureJournal of Neurosurgery, 1967