RF Exposure Assessment for Various Poses of Patient Assistant in Open MRI Environment
Open Access
- 28 May 2021
- journal article
- research article
- Published by MDPI AG in Applied Sciences
- Vol. 11 (11) , 4967
Abstract
In this study, the radio-frequency (RF) energy exposure of patient assistants was assessed for an open magnetic resonance imaging (MRI) system based on numerical computations of the head and body RF coil. Various poses of the patient assistants were defined to see how poorly they affected the RF energy exposure. For the assessments, the peak spatial-averaged specific absorption rate (SAR) levels were carefully compared with each patient assistant pose based on the finite-difference time domain calculations of RF coil models when the patient was placed in such coils in a 0.3 Tesla open MRI system. Overall, the SAR levels of the patient assistant were much lower than those of the patient. However, significantly increased SAR levels were observed under specific conditions, including a larger loop size of the patient assistants’ arms and a closer distance to the RF coils. A comparably high level of SAR to the patient’s body was also found. More careful investigations are needed to prevent the increase of SAR in patient assistants for open MRI systems at higher field strengths.Keywords
Funding Information
- ICT R&D program of MSIT/IITP (2019-0-00102)
This publication has 28 references indexed in Scilit:
- Specific absorption rate in neonates undergoing magnetic resonance procedures at 1.5 T and 3 TNMR in Biomedicine, 2015
- Whole‐body and local RF absorption in human models as a function of anatomy and position within 1.5T MR body coilMagnetic Resonance in Medicine, 2013
- Calculation of radiofrequency electromagnetic fields and their effects in MRI of human subjectsMagnetic Resonance in Medicine, 2011
- Parallel transmit and receive technology in high‐field magnetic resonance neuroimagingInternational Journal of Imaging Systems and Technology, 2010
- The Virtual Family—development of surface-based anatomical models of two adults and two children for dosimetric simulationsPhysics in Medicine & Biology, 2009
- AMENDMENT TO THE ICNIRP “STATEMENT ON MEDICAL MAGNETIC RESONANCE (MR) PROCEDURES: PROTECTION OF PATIENTS”Health Physics, 2009
- Whole‐body imaging at 7T: Preliminary resultsMagnetic Resonance in Medicine, 2008
- Factors influencing flip angle mapping in MRI: RF pulse shape, slice‐select gradients, off‐resonance excitation, and B0 inhomogeneitiesMagnetic Resonance in Medicine, 2006
- Temperature and SAR calculations for a human head within volume and surface coils at 64 and 300 MHzJournal of Magnetic Resonance Imaging, 2004
- The dielectric properties of biological tissues: III. Parametric models for the dielectric spectrum of tissuesPhysics in Medicine & Biology, 1996