Effect of scatter and an antiscatter grid on the performance of a slot‐scanning digital mammography system
Authors:
- Aili K. BloomquistABAili K. Bloomquist
- Gord E. MawdsleyGMGord E. Mawdsley
- Martin J. YaffeMYMartin J. Yaffe
- Idris ElbakriIEIdris Elbakri
- 29 March 2006
- journal article
- Published by Wiley in Medical Physics
- Vol. 33 (4) , 1108-1115
Abstract
The use of a grid increases perceptibility of low contrast objects in mammography. Slot‐scan mammography provides a more dose efficient reduction of the scattered radiation reaching the detector than obtained with an antiscatter grid in screen‐film or flat‐panel digital mammography. In this paper, the potential of using a grid in a slot‐scan system to provide a further reduction of scattered radiation is investigated. The components of the digital signal: primary radiation, off‐focus radiation, scattered radiation, and optical fluorescence glare in a detector were quantified. Based on these measurements, the primary and scatter transmission factors , scatter‐to‐primary ratio (SPR), signal‐difference‐to‐noise ratio (SDNR), and the SDNR improvement factor were obtained. Our results showed that the SPR ranged from 0.05 to 0.19 for breast thicknesses between 2 and , respectively. The values of ranged from 0.85 to 0.94. Because the slot‐scanning system has an inherently low SPR, the increase in dose required when the grid is used outweighs the benefit of the small increase in SDNR. It is possible that greater benefit could be achieved by using a grid with a higher , such as obtained using air‐core technology.Keywords
This publication has 15 references indexed in Scilit:
- Quality control for digital mammography in the ACRIN DMIST trial: Part IMedical Physics, 2006
- Scatter/primary in mammography: Comprehensive resultsMedical Physics, 2000
- Scatter/primary in mammography: Monte Carlo validationMedical Physics, 2000
- Scattered radiation in scanning slot mammographyMedical Physics, 1998
- A method for measuring the presampled MTF of digital radiographic systems using an edge test deviceMedical Physics, 1998
- Performance of glass fiber antiscatter devices at mammographic energiesMedical Physics, 1994
- Grids or air gaps for scatter reduction in digital radiography: A model calculationMedical Physics, 1992
- A simple method for determining the modulation transfer function in digital radiographyIEEE Transactions on Medical Imaging, 1992
- Studies of performance of antiscatter grids in digital radiography: Effect on signal‐to‐noise ratioMedical Physics, 1990
- Effect of reduced scatter on radiographic information content and patient exposure: A quantitative demonstrationMedical Physics, 1980