Stenosis Caused by Suture-Mediated Vascular Closure Device in an Angiographic Normal Common Femoral Artery: Its Mechanism and Management
- 9 August 2018
- journal article
- case report
- Published by SAGE Publications in Vascular and Endovascular Surgery
- Vol. 53 (1) , 58-61
Abstract
Vascular closing devices (VCDs) are widely used to replace manual compression at the femoral puncture site and to reduce the discomfort of patients undergoing percutaneous coronary procedure by shortening bed rest. Among the vascular complications related to these devices, the femoral artery stenosis or occlusion is rarely reported, and its standard management is not well established. We report a case of symptomatic femoral artery stenosis caused by suture-mediated VCD and managed using rotational atherectomy device and balloon angioplasty. In addition, we propose the possible mechanisms for this complication.Keywords
This publication has 10 references indexed in Scilit:
- Manual Compression versus Vascular Closing Device for Closing Access Puncture Site in Femoral Left-Heart Catheterization and Percutaneous Coronary Interventions: A Retrospective Cross-Sectional Comparison of Costs and Effects in Inpatient CareValue in Health, 2016
- Vascular Closure Device Failure: Frequency and ImplicationsCirculation: Cardiovascular Interventions, 2009
- Femoral artery occlusion after using a suture‐mediated arterial puncture closure deviceCatheterization and Cardiovascular Interventions, 2008
- Iatrogenic Limb Ischemia Caused by Angiography Closure DevicesVascular and Endovascular Surgery, 2007
- Occlusion of the femoral artery after using the percutaneous suture-mediated closure system percloseClinical Research in Cardiology, 2006
- Claudication secondary to Perclose use after percutaneous proceduresCatheterization and Cardiovascular Interventions, 2006
- Vascular complications associated with arteriotomy closure devices in patients undergoing percutaneous coronary proceduresJACC, 2004
- Arterial Puncture Closing Devices Compared With Standard Manual Compression After Cardiac CatheterizationJAMA, 2004
- Successful Angioplasty of a Superficial Femoral Artery Stenosis Caused by a Suture-Mediated Closure DeviceCardioVascular and Interventional Radiology, 2003
- Iatrogenic vascular injuries from percutaneous vascular suturing devicesJournal of Vascular Surgery, 2001