Breast MR Imaging at 3 T with Dual-Source Radiofrequency Transmission Offers Superior B1Homogeneity: An Intraindividual Comparison with Breast MR Imaging at 1.5 T
Authors:
- Isabelle TropITIsabelle Trop
- Guillaume GilbertGGGuillaume Gilbert
- Marko K. IvancevicMIMarko K. Ivancevic
- Gilles BeaudoinGBGilles Beaudoin
- 1 May 2013
- journal article
- research article
- Published by Radiological Society of North America (RSNA) in Radiology
- Vol. 267 (2) , 602-608
Abstract
To evaluate and compare B1 homogeneity for breast magnetic resonance (MR) imaging performed at 3 T with dual-source radiofrequency (RF) transmission to 1.5-T MR imaging and 3-T MR imaging with quadrature transmission. This prospective study received institutional review board approval and patients provided informed consent. Women (n = 25; mean age, 53 years; range, 30-68 years) suspected of having breast lesions underwent breast MR imaging examinations on comparable 1.5-T and 3-T clinical systems between February and May 2012. B1 maps were obtained at 1.5 T and at 3 T with quadrature and dual-source RF transmission. Intrabreast differences and differences in mean B1 values between right and left breasts were investigated by using two-sided multivariate analysis of variance with interaction; t tests were used to compare the differences between measured whole-breast mean B1 values and requested B1 values. With quadrature transmission at 1.5 T and 3 T, the mean B1 values showed a statistically significant difference: left-breast measured B1 was -8.9% of requested B1 value at 1.5 T and -13.7% at 3 T (P < .001), whereas right-breast measured B1 was +5.4% of requested B1 value at 1.5 T (P < .001) and +2.7% at 3 T (P = .01). With dual-source RF transmission at 3 T, mean B1 values across the breasts were not statistically different, nor were the measured B1 values compared with requested B1 values (left breast, -0.6%; right breast, -0.7%). At 3 T with dual-source transmission, slight intrabreast local variations in B1 were recorded. MR imaging at 3 T with dual-source RF transmission offered an overall B1 homogeneity for breast imaging that was better than that obtained at 1.5 T and with quadrature transmission.Keywords
This publication has 16 references indexed in Scilit:
- Dual-Source Parallel Radiofrequency Excitation Body MR Imaging Compared with Standard MR Imaging at 3.0 T: Initial Clinical ExperienceRadiology, 2010
- B1 transmission‐field inhomogeneity and enhancement ratio errors in dynamic contrast‐enhanced MRI (DCE‐MRI) of the breast at 3TJournal of Magnetic Resonance Imaging, 2009
- The accuracy of pharmacokinetic parameter measurement in DCE-MRI of the breast at 3 TPhysics in Medicine & Biology, 2009
- Whole-Body High-Field-Strength (3.0-T) MR Imaging in Clinical Practice Part I. Technical Considerations and Clinical ApplicationsRadiology, 2008
- Artifacts in 3-T MRI: Physical background and reduction strategiesEuropean Journal of Radiology, 2008
- Effect of B1Inhomogeneity on Breast MR Imaging at 3.0 TRadiology, 2007
- Body and Cardiovascular MR Imaging at 3.0 TRadiology, 2007
- A Review of MR Physics: 3T versus 1.5TMagnetic Resonance Imaging Clinics of North America, 2007
- Parallel RF transmission in MRINMR in Biomedicine, 2006
- Central brightening due to constructive interference with, without, and despite dielectric resonanceJournal of Magnetic Resonance Imaging, 2005