The Effects of Temperature and Relative Humidity on Point-of-Care Glucose Measurements in Hospital Practice in a Tropical Clinical Setting
Open Access
- 9 July 2016
- journal article
- research article
- Published by SAGE Publications in Journal of Diabetes Science and Technology
- Vol. 10 (5) , 1094-1100
Abstract
Hospitals in tropical countries experience conditions that exceed manufacturer temperature and humidity limits for point-of-care (POC) glucose reagents. Our goal was to assess the effects of out-of-limits storage temperature, operating temperature, and operating humidity on POC glucose measurement reliability. Quality control measurements were performed monthly using glucose test strips stored under controlled conditions and in inpatient wards under ambient conditions. Glucose test strips were evaluated in groups organized by operating temperatures of 24-25 (group 1), 28-29 (group 2), and 33-34°C (group 3), and relative humidity (RH) of ≤70 (group A), ~80 (group B), and ~90% (group C). Glucose results for different storage conditions were inconsistent. Measurements at higher operating temperatures had lower values with mean differences of −2.4 (P < .001) and −36.5 (P < .001) mg/dL (28-29 vs 24-25°C), and −3.6 (P < .001) and −37.4 (P < .001) mg/dL (33-34 vs 24-25°C) for low and high control levels, respectively. Measurements at higher RH had lower values with mean differences of −4.0 (P < .001) and −13.2 (P < .001) mg/dL (~80 vs ≤70% RH), and −5.8 (P < .001) and −16.6 (P < .001) mg/dL (~90 vs ≤70% RH) for low and high levels, respectively. High temperature and high RH decreased glucose concentrations for the POC oxidase-based system we evaluated. We recommend that individual hospitals perform stress testing, then determine if maximum absolute differences, which represent highest risk for patients, are clinically significant for decision making by using error grid analysis.Keywords
This publication has 19 references indexed in Scilit:
- Short-Term Thermal-Humidity Shock Affects Point-of-Care Glucose TestingJournal of Diabetes Science and Technology, 2014
- Vulnerability of point-of-care test reagents and instruments to environmental stresses: implications for health professionals and developersClinical Chemistry and Laboratory Medicine (CCLM), 2013
- Effects of Dynamic Temperature and Humidity Stresses on Point-of-Care Glucose Testing for Disaster CareDisaster Medicine and Public Health Preparedness, 2012
- The Impact of Inpatient Point-Of-Care Blood Glucose Quality Control TestingJournal for Healthcare Quality, 2012
- Effect of Ambient Temperature on Analytical Performance of Self-Monitoring Blood Glucose SystemsDiabetes Technology & Therapeutics, 2011
- Thermal Stress and Point-of-Care Testing Performance: Suitability of Glucose Test Strips and Blood Gas Cartridges for Disaster ResponseDisaster Medicine and Public Health Preparedness, 2009
- Adverse Impact of Temperature and Humidity on Blood Glucose Monitoring Reliability: A Pilot StudyDiabetes Technology & Therapeutics, 2007
- Irreversible Thermal Denaturation of Glucose Oxidase from Aspergillus niger Is the Transition to the Denatured State with Residual StructureJournal of Biological Chemistry, 2004
- Thermal Inactivation of Glucose OxidaseJournal of Biological Chemistry, 2003
- Thermal stability of glucose oxidase and its admixtures with synthetic polymersBiotechnology and Bioengineering, 1973