A method for determination of the absorption and scattering properties interstitially in turbid media
Authors:
- Jarod C FinlayJFJarod C Finlay
- Timothy C ZhuTZTimothy C Zhu
- 5 May 2005
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 50 (10) , 2291-2311
Abstract
We have developed a method to quickly determine tissue optical properties (absorption coefficient mu(a) and transport scattering coefficient mu'(s)) by measuring the ratio of light fluence rate to source power along a linear channel at a fixed distance (5 mm) from an isotropic point source. Diffuse light is collected by an isotropic detector whose position is determined by a computer-controlled step motor, with a positioning accuracy of better than 0.1 mm. The system automatically records and plots the light fluence rate per unit source power as a function of position. The result is fitted with a diffusion equation to determine mu(a) and mu'(s). We use an integrating sphere to calibrate each source-detector pair, thus reducing uncertainty of individual calibrations. To test the ability of this algorithm to accurately recover the optical properties of the tissue, we made measurements in tissue simulating phantoms consisting of Liposyn at concentrations of 0.23, 0.53 and 1.14% (mu'(s) = 1.7-9.1 cm(-1)) in the presence of Higgins black India ink at concentrations of 0.002, 0.012 and 0.023% (mu(a) = 0.1-1 cm(-1)). For comparison, the optical properties of each phantom are determined independently using broad-beam illumination. We find that mu(a) and mu'(s) can be determined by this method with a standard (maximum) deviation of 8% (15%) and 18% (32%) for mu(a) and mu'(s), respectively. The current method is effective for samples whose optical properties satisfy the requirement of the diffusion approximation. The error caused by the air cavity introduced by the catheter is small, except when mu(a) is large (mu(a) > 1 cm(-1)). We presented in vivo data measured in human prostate using this method.Keywords
This publication has 38 references indexed in Scilit:
- Optical Properties of Human Prostate at 732 nm Measured In Vivo during Motexafin Lutetium-Mediated Photodynamic TherapyPhotochemistry and Photobiology, 2004
- In Vivo Near-Infrared SpectroscopyAnnual Review of Biomedical Engineering, 2000
- Predictions and measurements of scattering and absorption over broad wavelength ranges in tissue phantomsApplied Optics, 1997
- Design and testing of a white-light, steady-state diffuse reflectance spectrometer for determination of optical properties of highly scattering systemsApplied Optics, 1997
- Improved solutions of the steady-state and the time-resolved diffusion equations for reflectance from a semi-infinite turbid mediumJournal of the Optical Society of America A, 1997
- Determination of optical properties and blood oxygenation in tissue using continuous NIR lightPhysics in Medicine & Biology, 1995
- Optical properties of intralipid: A phantom medium for light propagation studiesLasers in Surgery and Medicine, 1992
- The optical properties of aqueous suspensions of Intralipid, a fat emulsionPhysics in Medicine & Biology, 1989
- Time resolved reflectance and transmittance for the noninvasive measurement of tissue optical propertiesApplied Optics, 1989
- Measurements and calculations of the energy fluence rate in a scattering and absorbing phantom at 633 nmApplied Optics, 1989