Automatic Biopsy Instruments Used through a Coaxial Bone Biopsy System with an Eccentric Drill Tip

Authors:
  • KÅK. G. O. Åström
    K. G. O. Åström
  • JSJ. C. Sundström
    J. C. Sundström
  • PLP. G. Lindgren
    P. G. Lindgren
  • KAK. H. Ahlström
    K. H. Ahlström

Abstract
Twenty-eight consecutive CT (n=23) or ultrasonographically (n=5) guided biopsy procedures were performed on musculoskeletal lytic lesions covered (n=13) or not covered (n=15) with intact bone. Specimens were obtained by means of Biopty techniques (n=27), i.e. Biopty and Monopty instruments, through different cannulas with normal or shortened needle-throws. Four out of 5 bone penetrations were successful with an Ostycut needle, and all 8 bone penetrations by a coaxial bone biopsy system with an eccentric drill. The eccentric drill makes a hole in the cortical bone larger than the diameter of the outer cannula of this system, making it easy to anchor the cannula and then coaxially insert a Biopty-Gun needle for example. The overall histopathological diagnostic accuracy of the Biopty techniques was 25/27 (92.6%).

This publication has 13 references indexed in Scilit: