Note: A versatile, stable, high-resolution readout system for RTD and thermistor sensors
Authors:
- 1 September 2012
- journal article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 83 (9) , 096101
Abstract
A readout system for resistance temperature detectors and thermistors is described featuring temperature resolution better than 1 mK and capability to fit sensors having different resistance or requiring different excitation current. For instance, with a sensor equivalent to an ideal 100 Ω Pt, an excitation current of 0.7 mA, and reading @ 1Hz, the system resolution corresponds to 0.38 mK and its temperature coefficient (TC) to 0.26 mK/K. The system, however, can control its own temperature accurately enough to make its TC negligible. When thermostated, the overall stability of the system was better than 10 ppm for 230 h.Keywords
This publication has 5 references indexed in Scilit:
- A versatile thermoelectric temperature controller with 10 mK reproducibility and 100 mK absolute accuracyReview of Scientific Instruments, 2009
- Thermal diagnostics front-end electronics for LISA PathfinderReview of Scientific Instruments, 2007
- ±30 μK temperature controller from 25 to 103 °C: Study and analysisReview of Scientific Instruments, 2003
- Low noise temperature control: Application to an active cavity radiometerReview of Scientific Instruments, 2003
- Review of temperature measurementReview of Scientific Instruments, 2000