In vivo determination of nitrogen using Pu-Be sources
Authors:
- J R MernaghJMJ R Mernagh
- J E HarrisonJHJ E Harrison
- K G McNeilKMK G McNeil
- 1 September 1977
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 22 (5) , 831-835
Abstract
A mehtod of measuring body nitrogen by neutron capture gamma analysis is described. The neutron flux is provided by four collimated 5 Ci Pu-Be sources, placed to give a bilateral irradiation of the subject. The 10-83 MeV thermal neutron capture gamma rays from 14N are detected by two heavily shielded 5 in X 4 in NaI(Tl) detectors. Results indicate a reproducibility +/-3% (1 SD) for a 10 min irradiation of the chest area, while giving a radiation dose equivalent of 50 mREM.Keywords
This publication has 4 references indexed in Scilit:
- Effects of body thickness on in vivo neutron activation analysisThe International Journal of Applied Radiation and Isotopes, 1974
- MEASUREMENT OF WHOLE-BODY NITROGEN BY NEUTRON-ACTIVATION ANALYSISThe Lancet, 1973
- Neutron Dosimetry with Nuclear Track Detectors Applied to in Vivo Neutron Activation AnalysisHealth Physics, 1971
- NEUTRON-ACTIVATION ANALYSIS IN MAN IN VIVOThe Lancet, 1964