Effect of Applying Positive Pressure With or Without Endotracheal Suctioning During Extubation: A Laboratory Study
Authors:
Open Access
- 1 December 2014
- journal article
- Published by SAGE Publications in Respiratory Care
- Vol. 59 (12) , 1905-1911
Abstract
BACKGROUND: During invasive mechanical ventilation, secretions accumulate in the subglottic space; consequently, there is a risk of aspiration of these secretions into the airway during cuff deflation and extubation. To minimize this risk, 2 extubation methods are used. The first consists of introducing a suction catheter into the endotracheal tube (ETT) and the trachea. After initiating suctioning, the cuff is deflated and the ETT is removed together with the suction catheter. The second technique involves applying positive pressure to the ETT using a resuscitation bag. Once the manual breath is delivered, the ETT cuff is deflated and the ETT is removed without suction. The aim of this laboratory study is to determine the existence and magnitude of differences in leak volume during cuff deflation and extubation using various combinations of positive pressure with or without endotracheal suctioning. METHODS: An ETT connected to a ventilator was placed in a model trachea. Colored water was instilled in the space above the cuff. To measure the leak volume, a collection chamber was attached to the distal end of the model. Nine procedures were defined, based on the delivery of different positive pressure levels with or without endotracheal suctioning during extubation. The volume of leakage, in milliliters, was the unit of analysis. Procedures yielding values lower than 1 mL were assessed by the Friedman test, and a P value of less than .05 was considered significant. Post hoc comparisons were performed with a Wilcoxon test, followed by a Bonferroni correction. RESULTS: The application of CPAP 15 , pressure support ventilation (PSV) 15/10 , and PSV 20/5 produced 0.4, 0.2, and 0.1 mL of leak volume, respectively. Statistically significant differences were found between CPAP 15 and PSV 15/10 ( P = .003) and between CPAP 15 and PSV 20/5 ( P = .01), but not between PSV 15/10 and PSV 20/5 ( P = .30). The addition of suctioning increased leak volume, with statistically significant differences between CPAP 15 and CPAP 15 + endotracheal suctioning ( P = .001) and between PSV 15/10 and PSV 15/10 + endotracheal suctioning ( P = .001). CONCLUSIONS: Endotracheal suctioning during cuff deflation and extubation produced greater leakage. Application of CPAP 15 , PSV 15/10 , and PSV 20/5 resulted in the lowest leak values, with the best results being obtained with the use of PSV 15/10 and PSV 20/5 .Keywords
This publication has 14 references indexed in Scilit:
- Incidencia, características y evolución del barotrauma durante la ventilación mecánica con apertura pulmonarMedicina Intensiva (English Edition), 2012
- Massive Aspiration Past the Tracheal Tube Cuff Caused by Closed Tracheal Suction SystemJournal of Intensive Care Medicine, 2011
- Extubation in intensive care units in the UK: an online surveyNursing in Critical Care, 2010
- Increasing positive end expiratory pressure at extubation reduces subglottic secretion aspiration in a bench-top modelNursing in Critical Care, 2010
- Endotracheal suctioning of the adult intubated patient—What is the evidence?Intensive and Critical Care Nursing, 2009
- Extubation failure: an outcome to be avoidedCritical Care, 2004
- Evaluation of routine tracheal extubation in children: inflating or suctioning technique?British Journal of Anaesthesia, 1998
- Leakage of fluid past the tracheal tube cuff in a benchtop modelBritish Journal of Anaesthesia, 1997
- The pathogenesis of ventilator-associated pneumonia: I. Mechanisms of bacterial transcolonization and airway inoculationIntensive Care Medicine, 1995
- Laryngeal Competence after Tracheal ExtubationAnesthesiology, 1979