Novel C‐arm based planning spine surgery robot proved in a porcine model and quantitative accuracy assessment methodology
- 13 October 2020
- journal article
- research article
- Published by Wiley in The International Journal of Medical Robotics and Computer Assisted Surgery
- Vol. 17 (2) , e2182
Abstract
Background: We assessed pedicle screw accuracy utilizing a novel navigation‐based spine surgery robotic system by comparing planned pathways with placed pathways in a porcine model.Methods: We placed three mini screws per vertebra for accuracy evaluation and used a reference frame for registration in four pigs (46 screws in 23 vertebrae). We planned screw paths and performed screw insertion under robot guidance. Using C‐arm and CT images, we evaluated accuracy by comparing the 3D distance of the placed screw head/tip from the planned screw head/tip and 3D angular offset.Results: Mean registration deviation between the preoperative 3D space (C‐arm) and postoperative CT scans was 0.475 ± 0.119 mm. The average offset from preoperative plan to final placement was 4.8 ± 2.0 mm from the head (tail), 5.3 ± 2.3 mm from the tip and 3.9 ± 2.4 degrees of angulation.Conclusions: Our spine surgery robot showed good accuracy in executing an intended planned trajectory and screw path. This faster and more accurate robotic system will be applied in future studies, first in cadavers and subsequently in the clinical field.Keywords
This publication has 34 references indexed in Scilit:
- Spinal RoboticsNeurosurgery, 2013
- Perioperative course and accuracy of screw positioning in conventional, open robotic-guided and percutaneous robotic-guided, pedicle screw placementEuropean Spine Journal, 2011
- Clinical Accuracy of Computer-Assisted Two-Dimensional Fluoroscopy for the Percutaneous Placement of Lumbosacral Pedicle ScrewsSpine, 2011
- Clinical Acceptance and Accuracy Assessment of Spinal Implants Guided With SpineAssist Surgical RobotSpine, 2010
- Complications of Pedicle Screw Fixation in Scoliosis SurgerySpine, 2010
- Image-guided pedicle screw insertion accuracy: a meta-analysisInternational Orthopaedics, 2009
- Percutaneous Placement of Pedicle Screws in the Lumbar Spine Using a Bone Mounted Miniature Robotic SystemSpine, 2009
- Pedicle Screw Placement AccuracySpine, 2007
- Accuracy of pedicle screw insertion with and without computer assistance: a randomised controlled clinical study in 100 consecutive patientsEuropean Spine Journal, 2000
- Closed-form solution of absolute orientation using unit quaternionsJournal of the Optical Society of America A, 1987