A multichannel continuously selectable multifrequency electrical impedance spectroscopy measurement system
Authors:
- R.A. MazzareseRMR.A. Mazzarese
- F.R. ReissFRF.R. Reiss
- T.E. KernerTKT.E. Kerner
- K.S. OstermanKOK.S. Osterman
- D.B. WilliamsDWD.B. Williams
- K.D. PaulsenKPK.D. Paulsen
- 1 January 2000
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. 47 (1) , 49-58
Abstract
There is increasing evidence that alterations in the electrical property spectrum of tissues below 10 MHz is diagnostic for tissue pathology and/or pathophysiology. Yet, the complexity associated with constructing a high-fidelity multichannel, multifrequency data acquisition instrument has limited widespread development of spectroscopic electrical impedance imaging concepts. To contribute to the relatively sparse experience with multichannel spectroscopy systems this paper reports on the design, realization and evaluation of a prototype 32-channel instrument. The salient features of the system include a continuously selectable driving frequency up to 1 MHz, either voltage or current source modes of operation and simultaneous measurement of both voltage and current on each channel in either of these driving configurations. Comparisons of performance with recently reported fixed-frequency systems is favorable. Volts dc (VDC) signal-to-noise ratios of 75-80 dB are achieved and the noise floor for ac signals is near 100 dB below the signal strength of interest at 10 kHz and 60 dB down at 1 MHz. The added benefit of being able to record multispectral information on source and sense signal amplitudes and phases has also been realized. Phase-sensitive detection schemes and multiperiod undersampling techniques have been deployed to ensure measurement fidelity over the full bandwidth of system operation.Keywords
This publication has 29 references indexed in Scilit:
- The dielectric properties of biological tissues: III. Parametric models for the dielectric spectrum of tissuesPhysics in Medicine & Biology, 1996
- The dielectric properties of biological tissues: II. Measurements in the frequency range 10 Hz to 20 GHzPhysics in Medicine & Biology, 1996
- The dielectric properties of biological tissues: I. Literature surveyPhysics in Medicine & Biology, 1996
- Deconvolved electrical impedance spectra track distinct cell morphology changesIEEE Transactions on Biomedical Engineering, 1996
- A 32-electrode data collection system for electrical impedance tomographyIEEE Transactions on Biomedical Engineering, 1995
- Tissue spectroscopy with electrical impedance tomography: computer simulationsIEEE Transactions on Biomedical Engineering, 1995
- A real-time electrical impedance tomography system for clinical use-design and preliminary resultsIEEE Transactions on Biomedical Engineering, 1995
- Bioelectrical spectroscopy from multi-frequency EITPhysiological Measurement, 1994
- Changes in electrical impedance of skeletal muscle measured during hyperthermiaInternational Journal of Hyperthermia, 1993
- A constant-perturbation saline phantom for electrical impedance tomographyPhysics in Medicine & Biology, 1989