Aims: This prospective, two-centre study derived and validated predictive algorithms for the Siemens Atellica IM high-sensitivity cardiac troponin I (hs-cTnI) assay in the emergency department (ED). Methods and results: Algorithms for predicting 30-day myocardial infarction (MI) Types 1 and 2 and death or non-ST-elevation MI (NSTEMI, Types 1 and 2) at index admission were developed from a derivation cohort of 1896 patients and validated using a synthetic data set with nearly 1 million patient cases. Performance was compared with the European Society of Cardiology algorithms for hs-cTnT (Roche Diagnostics) and hs-cTnI (Abbott Diagnostics). An admission hs-cTnI concentration < 5 ng/L had a negative predictive value (NPV) and sensitivity for 30-day MI or death of 99.5–99.7% and 98.1–98.8%, respectively, in the derivation cohort and validation data set. The NPV and sensitivity were ≥99.7% and ≥98.8% for ruling out index NSTEMI. A 0- to 1-h algorithm with baseline hs-cTnI concentration < 10 ng/L and Δ change < 3 ng/L had NPV of ≥99.5% and sensitivity ≥ 97.3% for predicting 30-day MI or death and a ≥99.5% sensitivity and NPV for index NSTEMI. Rule-in algorithms of either 0-h hs-cTnI ≥ 120 ng/L or 0- to 1-h Δ change ≥ 12 ng/L had positive predictive value ≥ 73% and specificity > 96% for 30-day MI or death and index NSTEMI. The results were comparable with established hs-cTn algorithms. Conclusion: This study presents Siemens Atellica hs-cTnI algorithms for diagnosis and risk prediction in the ED with performance comparable with established hs-cTnT (Roche) and hs-cTnI (Abbott) algorithms.Keywords
Funding Information
- Western Norway Regional Health Authority (912265, ID: F-12501)
This publication has 28 references indexed in Scilit:
- New Creatinine- and Cystatin C–Based Equations to Estimate GFR without RaceThe New England journal of medicine, 2021
- Performance of Novel High-Sensitivity Cardiac Troponin I Assays for 0/1-Hour and 0/2- to 3-Hour Evaluations for Acute Myocardial Infarction: Results From the HIGH-US StudyAnnals of Emergency Medicine, 2020
- Myocardial Infarction Risk Stratification With a Single Measurement of High-Sensitivity Troponin IJACC, 2019
- Aiming toWards Evidence baSed inTerpretation of Cardiac biOmarkers in patients pResenting with chest pain-the WESTCOR study: study designScandinavian Cardiovascular Journal, 2019
- Application of High-Sensitivity Troponin in Suspected Myocardial InfarctionNew England Journal of Medicine, 2019
- High-Sensitivity Cardiac Troponin Assays Potentially Differentiate Acute From Chronic Myocardial InjuryJACC, 2019
- Description of chest pain patients in a Norwegian emergency departmentScandinavian Cardiovascular Journal, 2019
- Validation in prediction research: the waste by data splittingJournal of Clinical Epidemiology, 2018
- The small number problem in diagnostic algorithms and why we need to bootstrapClinical Biochemistry, 2017
- What is an acceptable risk of major adverse cardiac event in chest pain patients soon after discharge from the Emergency Department?International Journal of Cardiology, 2013