Computer-Simulation Technique for Low Dose Computed Tomographic Screening

Authors:
  • THTetsuya Horiuchi
    Tetsuya Horiuchi
  • JSJunko Sekiguchi
    Junko Sekiguchi
  • YMYoshihisa Muramatsu
    Yoshihisa Muramatsu
  • RKRyutaro Kakinuma
    Ryutaro Kakinuma
  • NMNoriyuki Moriyama
    Noriyuki Moriyama
  • RTRyosuke Tuchiiya
    Ryosuke Tuchiiya
  • NNNoboru Niki
    Noboru Niki

Abstract
The purpose of this study is to assess the relative influence of noise and artifact in detecting lung nodules on low dose computed tomographic (CT) screening. We develop the computer-simulation technique that allows tube current simulation and virtual nodule insertion in any CT images. The tube current simulation uses a reduction model that adds random Gaussian noise distribution to existing projection data. The virtual nodules are generated using a dedicated CT simulation tool with same scanner geometry. The coefficient of the correlations between the contrast-to-noise ratio of the actual scan and simulated tube current images was 0.98. There was no difference in CT number between virtual nodules and actual nodules [t test results = 0.60, t50(0.01) = 2.70 at 10 mA] and the coefficient of the correlations of the image noise was 0.99. Our technique is useful for systematic evaluation of radiation dose reduction and structure visibility in low-dose CT screening.