A Compact Microwave Device for Fracture Diagnosis of the Human Tibia
Authors:
- Vimal Samsingh RamalingamVRVimal Samsingh Ramalingam
- Malathi KanagasabaiMKMalathi Kanagasabai
- Esther Florence SundarsinghESEsther Florence Sundarsingh
- 21 January 2019
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Components, Packaging and Manufacturing Technology
- Vol. 9 (4) , 661-668
Abstract
In this paper, a portable system for non-invasive detection of fractures in human bones, is presented. A planar microwave ring resonator is used as the sensor. The proposed planar microwave sensor is designed using the Finite Integration (FIT) based CST Microwave Studio and tested on a human bone covered by porcine tissues. Fracture induced in the bone was detected by monitoring the transmission characteristics of the sensor as it scans the affected region. The resonant frequency of the ring resonator has been observed at every scan point and the values are plotted using Iterative Curvature-Based Interpolation (ICBI), Improved New Edge Directed Interpolation (iNEDI) and Lanczos imaging technique. The extent of fracture is then derived from the image. The average accuracy of the detected fracture for Lanczos method of resampling was98.86 %.Specific Absorption Rate (SAR) was below the minimum level of 1.6 W/Kg at every point of scanning. Portable, non-invasive monitoring of fractures in the tibial bone will help in immobilizing the affected part at an earlier stage, thereby reducing the recovery time for patients.Keywords
This publication has 52 references indexed in Scilit:
- Accurate Human Tissue Characterization for Energy-Efficient Wireless On-Body CommunicationsSensors, 2013
- Stress Fractures About the Tibia, Foot, and AnkleJournal of the American Academy of Orthopaedic Surgeons, 2012
- Cole–Cole parameters for the dielectric properties of porcine tissues as a function of age at microwave frequenciesPhysics in Medicine & Biology, 2010
- Effects of delayed stabilization on fracture healingJournal of Orthopaedic Research, 2007
- The ultrasonic wave propagation in composite material and its characteristic evaluationComposite Structures, 2006
- New edge-directed interpolationIEEE Transactions on Image Processing, 2001
- Delayed union stress fractures of the anterior tibia: conservative managementBritish Journal of Sports Medicine, 2001
- Measurement of dielectric constant using a microstrip ring resonatorIEEE Transactions on Microwave Theory and Techniques, 1991
- Negative bone scans in impending tibial stress fracturesThe American Journal of Sports Medicine, 1984
- Analysis of bone as a composite materialInternational Journal of Engineering Science, 1973