Accessorized DPI: a Shortcut towards Flexibility and Patient Adaptability in Dry Powder Inhalation
- 13 September 2016
- journal article
- research article
- Published by Springer Nature in Pharmaceutical Research
- Vol. 33 (12) , 3012-3020
Abstract
In this work, a novel powder dispersion add-on device, the AOS (Axial Oscillating Sphere), was studied in conjunction with commercially available DPI devices to improve the powder dispersion. An ordered mixture of formoterol fumarate and lactose was selected. We studied in two laboratories located at different altitudes the dispensing and dispersion of the drug at different flow rates, paying particular attention to a number of metrics of Fine Particle Dose (FPD). Two novel findings emerged from the data collected. First, the aerosol quality, measured as fine particle dose, can be increased by adding the accessory promoting the dispersion and de-aggregation of the formulation. The second finding was that, albeit the emitted dose was independent of altitude, the drug/lactose carrier DPI aerosolizing performance changed with the altitude of testing. In particular, fine particle dose depended on both altitude and device configuration. The RS01 inhaler without the AOS accessory used at higher altitude gave the lowest FPD values. By combining the AOS accessory with the DPI, however, the performance dependence on altitude/atmospheric pressure was essentially removed. Increasing inhaler performance can be achieved using an add-on accessory that enhances aerosol dispersion and minimizes flow rate dependency.Keywords
This publication has 16 references indexed in Scilit:
- The Role of Inspiratory Pressures in Determining the Flow Rates Though Dry Powder Inhalers; A ReviewCurrent Pharmaceutical Design, 2015
- NEXThaler, an innovative dry powder inhaler delivering an extrafine fixed combination of beclometasone and formoterol to treat large and small airways in asthmaExpert Opinion on Drug Delivery, 2014
- Differences in physical chemistry and dissolution rate of solid particle aerosols from solution pressurised inhalersInternational Journal of Pharmaceutics, 2014
- Lung penetration and patient adherence considerations in the management of asthma: role of extra fine formulationsJournal of Asthma and Allergy, 2013
- Aerosol drug delivery: developments in device design and clinical useThe Lancet, 2011
- Inspiratory flows through dry powder inhaler in chronic obstructive pulmonary disease: age and gender rather than severity mattersInternational Journal of Chronic Obstructive Pulmonary Disease, 2010
- Lung Deposition of BDP/Formoterol HFA pMDI in Healthy Volunteers, Asthmatic, and COPD PatientsJournal of Aerosol Medicine and Pulmonary Drug Delivery, 2010
- Inhaled therapy in elderly COPD patients; time for re-evaluation?Age and Ageing, 2007
- Validation of a RP-HPLC method for the assay of formoterol and its related substances in formoterol fumarate dihydrate drug substanceJournal of Pharmaceutical and Biomedical Analysis, 2003
- Clinical effect of Diskus dry-powder inhaler at low and high inspiratory flow-rates in asthmatic childrenEuropean Respiratory Journal, 1998