Performance evaluation of the automated nucleated red blood cell count of five commercial hematological analyzers
- 8 October 2017
- journal article
- research article
- Published by Wiley in International Journal of Laboratory Hematology
- Vol. 39 (6) , 663-670
Abstract
Introduction: Recent automated hematology analyzers (HAs) can identify and report nucleated red blood cells (NRBC) count as a separate population out of white blood cells (WBC). The aim of this study was to investigate the analytical performances of NRBC enumeration on five top of the range HAs.Methods: We evaluated the within‐run and between‐day precision, limit of blank (LoB), limit of detection (LoD), and limit of quantitation (LoQ) of XE‐2100 and XN‐module (Sysmex), ADVIA 2120i (Siemens), BC‐6800 (Mindray), and UniCel DxH 800 (Beckman Coulter). Automated NRBC counts were also compared with optical microscopy (OM).Results: The limits of detection for NRBC of the BC‐6800, XN‐module, XE‐2100, UniCel DxH 800, and ADVIA 2120i are 0.035×109/L, 0.019×109/L, 0.067×109/L, 0.038×109/L, and 0.167×109/L, respectively. Our data indicated excellent performance in terms of precision. The agreement with OM was excellent for BC‐6800, XN‐module, and XE‐2100 (Bias 0.023, 0.019, and 0.033×109/L, respectively). ADVIA 2120i displayed a significant constant error and UniCel DxH 800 both proportional and small constant error.Conclusion: Regards to NRBC counting, the performances shown by BC‐6800, XN‐module, and XE‐2100 are excellent also a low count, ADVIA 2120i and UniCel DxH 800 need to be improved.Keywords
This publication has 29 references indexed in Scilit:
- Preliminary evaluation of complete blood cell count on Mindray BC-6800Clinical Chemistry and Laboratory Medicine (CCLM), 2012
- Pathology harmony moves on: progress on implementation in haematologyBritish Journal of Haematology, 2012
- Getting back to zero with nucleated red blood cells: following trends is not necessarily a bad thingThe American Journal of Surgery, 2012
- Evaluation of the Beckman Coulter UniCel DxH 800 and Abbott Diagnostics Cell-Dyn Sapphire Hematology Analyzers on Pediatric and Neonatal Specimens in a Tertiary Care HospitalAmerican Journal of Clinical Pathology, 2011
- Evaluation of the Beckman Coulter UniCel DxH 800, Beckman Coulter LH 780, and Abbott Diagnostics Cell-Dyn Sapphire Hematology Analyzers on Adult Specimens in a Tertiary Care HospitalAmerican Journal of Clinical Pathology, 2011
- The new haematology analyzer DxH 800: an evaluation of the analytical performances and leucocyte flags, comparison with the LH 755International Journal of Laboratory Hematology, 2010
- Automated Blood Cell CountsAmerican Journal of Clinical Pathology, 2008
- Nucleated red blood cell count in term and preterm newborns: reference values at birthArchives of Disease in Childhood - Fetal and Neonatal Edition, 2005
- Nucleated red blood cells in the fetus and newbornArchives of Disease in Childhood - Fetal and Neonatal Edition, 2001
- Nucleated RBCs—Significance in the Peripheral Blood FilmLaboratory Medicine, 2000