Intrarenal pressure detection during flexible ureteroscopy with fiber optic pressure sensor system in porcine model
Open Access
- 24 April 2024
- journal article
- research article
- Published by Springer Nature in Scientific Reports
- Vol. 14 (1) , 1-11
Abstract
To validate the feasibility of a fiber-optic pressure sensor-based pressure measurement device for monitoring intrarenal pressure and to analyze the effects of ureteral acess sheath (UAS) type, surgical location, perfusion flow rate, and measurement location on intrarenal pressure (IRP). The measurement deviations and response times to transient pressure changes were compared between a fiber-optic pressure sensing device and a urodynamic device IRP in an in vitro porcine kidney and in a water tank. Finally, pressure measurements were performed in anesthetized female pigs using fiber-optic pressure sensing device with different UAS, different perfusion flow rates, and different surgical positions at different renal calyces and ureteropelvic junctions (UPJ). According to our operation, the result is fiber optic pressure sensing devices are highly accurate and sensitive. Under the same conditions, IRP varied among different renal calyces and UPJ (P < 0.05). IRP was lowest at 50 ml/min and highest at 150 ml/min (P < 0.05). Surgical position had a significant effect on IRP (P < 0.05). 12/14 Fr UAS had a lower IRP than 11/13 Fr UAS. Therefore fiber optic pressure sensing devices are more advantageous for IRP measurements. In ureteroscopy, the type of ureteral sheath, the surgical position, the perfusion flow rate, and the location of the measurement all affect the intrarenal pressure value.Keywords
This publication has 31 references indexed in Scilit:
- Comparison of intrarenal pressure between convention and vacuum-assisted ureteral access sheath using an ex vivo porcine kidney modelWorld Journal of Urology, 2022
- An irrigation system for noninvasively estimating intrarenal pressure during flexible ureteroscopyThe International Journal of Medical Robotics and Computer Assisted Surgery, 2021
- A novel robotic system for flexible ureteroscopyThe International Journal of Medical Robotics and Computer Assisted Surgery, 2020
- Characterization of intracalyceal pressure during ureteroscopyWorld Journal of Urology, 2020
- Continuous monitoring of intrapelvic pressure during flexible ureteroscopy using a sensor wire: a pilot studyWorld Journal of Urology, 2020
- Elevated Renal Pelvic Pressures during Percutaneous Nephrolithotomy Risk Higher Postoperative Pain and Longer Hospital StayJournal of Urology, 2018
- Magnetic resonance elastography can monitor changes in medullary stiffness in response to treatment in the swine ischemic kidneyMagnetic Resonance Materials in Physics, Biology and Medicine, 2017
- Animal Experimental Study to Test Application of Intelligent Pressure Control Device in Monitoring and Control of Renal Pelvic Pressure During Flexible UreteroscopyUrology, 2016
- From conventional sensors to fibre optic sensors for strain and force measurements in biomechanics applications: A reviewJournal of Biomechanics, 2014
- Fiber Optics Sensors2013