Turki Almubrad
Current Organizations / Institutes
The organizations/institutes from the scholar's latest article
Research Subjects
Turki Almubrad has a total of 9 co-authors. Here are the top co-authors by number of shared publications:
- Ali Abusharha
Department of Optometry, College of Applied Medical Sciences, King Saud University, Riyadh, Saudi Arabia
1 shared publication - Ali Mohammed Masmali
Department of Optometry, College of Applied Medical Sciences, King Saud University, Riyadh 11433, Saudi Arabia
1 shared publication - Gamal Abdel-Rahman El-Hiti
King Saud University
1 shared publication I Al Abbadi
1 shared publication- Kelechi Ogbuehi
Department of Medicine, Dunedin School of Medicine, University of Otago, Dunedin, New Zealand
1 shared publication - Mana A. Alanazi
Department of Optometry, College of Applied Medical Sciences, King Saud University, Riyadh 11433, Saudi Arabia
1 shared publication
Latest Publications
- Open Access
- 21 February 2022
- Published by MDPI AG in Healthcare
- Vol. 10 (2) , 405
Measurements of Tear Evaporation Rate in Subjects with Refractive Errors Using a Portable Evaporimeter
Dry eye symptoms are associated with refractive errors. We aimed to measure the tear evaporation rate (TER) in subjects with refractive errors (RE) using a portable evaporimeter. This nonrandomized, case–control, and observational study included 75 subjects, including 25 subjects aged 18–38 years (28.8 ± 6.8 years) with myopia (−0.75 to −3.75 D) and 25 subjects aged 18–39 years (27.7 ± 5.5 years) with hyperopia (+0.50 to +3.75 D). In addition, a control group of 25 subjects with emmetropic eyes aged 20–30 years (23.6 ± 2.6 years) was recruited. The ocular surface disease index (OSDI) was completed, followed by the TER measurements using a portable evaporimeter. The OSDI score and TER measurements showed dry eye symptoms in 36% and 48% of myopic subjects, respectively. For hyperopic participants, dry eye was observed in 24% and 56% of the subjects based on the OSDI and TER scores, respectively. Significant differences (Mann-Whitney U test; p < 0.001) were found among the OSDI and TER scores recorded within the study (myopic and hyperopic subjects) and control groups. Significant strong positive correlations were detected between the OSDI and TER scores in both myopic (p = 0.004; r = 0.559) and hyperopic (p = 0.001; r = 0.619) subjects. The TER scores were significantly higher in subjects with RE (myopic and hyperopic) as compared with individuals with normal eyes.
- 21 February 2022
- Published by MDPI AG in Healthcare
- Vol. 10 (2) , 405
Open AccessMeasurements of Tear Evaporation Rate in Subjects with Refractive Errors Using a Portable Evaporimeter
Dry eye symptoms are associated with refractive errors. We aimed to measure the tear evaporation rate (TER) in subjects with refractive errors (RE) using a portable evaporimeter. This nonrandomized, case–control, and observational study included 75 subjects, including 25 subjects aged 18–38 years (28.8 ± 6.8 years) with myopia (−0.75 to −3.75 D) and 25 subjects aged 18–39 years (27.7 ± 5.5 years) with hyperopia (+0.50 to +3.75 D). In addition, a control group of 25 subjects with emmetropic eyes aged 20–30 years (23.6 ± 2.6 years) was recruited. The ocular surface disease index (OSDI) was completed, followed by the TER measurements using a portable evaporimeter. The OSDI score and TER measurements showed dry eye symptoms in 36% and 48% of myopic subjects, respectively. For hyperopic participants, dry eye was observed in 24% and 56% of the subjects based on the OSDI and TER scores, respectively. Significant differences (Mann-Whitney U test; p < 0.001) were found among the OSDI and TER scores recorded within the study (myopic and hyperopic subjects) and control groups. Significant strong positive correlations were detected between the OSDI and TER scores in both myopic (p = 0.004; r = 0.559) and hyperopic (p = 0.001; r = 0.619) subjects. The TER scores were significantly higher in subjects with RE (myopic and hyperopic) as compared with individuals with normal eyes.
- Open Access
- 1 June 2011
- Published by Taylor & Francis in Clinical Ophthalmology
- Vol. 5, 871-876
Comparison of the precision of the Topcon SP-3000P specular microscope and an ultrasound pachymeter
Aim: To compare the precision of the Topcon SP-3000P noncontact specular microscope (NCSM) and the DGH 500 ultrasound pachymeter (USP). Methods: Triplicate measurements of central corneal thickness (CCT) for 100 eyes were taken with an NCSM and a USP in 2 visits separated by 1 week. Repeatability was assessed by computing the differences between all 3 readings from each subject. Coefficients of repeatability and reproducibility were computed. Results: Mean CCT as measured by each instrument were: 518.53 ± 34.96 µm (range 417.33–592.67) and 516.94 ± 33.60 µm (range 431.67–582.67) for sessions 1 and 2 respectively, with the NCSM; 546.69 ± 36.62 µm (range 457.33–617.00) and 549.78 ± 35.26 µm (range 454.00–618.67) for sessions 1 and 2 respectively, with the USP. The ultrasound CCT measurements were consistently higher than those obtained with the NCSM in both sessions 28.17 ± 19.20 µm (mean ± SD, session 1) and 32.81 ± 14.04 (mean ± SD, session 2). The repeatability coefficient for the NCSM was better in both sessions than those for USP (±10 µm vs ± 12 µm in session 1 and ±8 µm vs ±10 µm in session 2). The reproducibility coefficient with the NCSM was half that with the USP (±21 µm vs ±41 µm). Conclusion: The SP-3000P NCSM is a more precise and reproducible instrument for measurement of CCT than the USP, but both instruments are reliable, useful instruments for measuring CCT.
- 1 June 2011
- Published by Taylor & Francis in Clinical Ophthalmology
- Vol. 5, 871-876
Open AccessComparison of the precision of the Topcon SP-3000P specular microscope and an ultrasound pachymeter
Aim: To compare the precision of the Topcon SP-3000P noncontact specular microscope (NCSM) and the DGH 500 ultrasound pachymeter (USP). Methods: Triplicate measurements of central corneal thickness (CCT) for 100 eyes were taken with an NCSM and a USP in 2 visits separated by 1 week. Repeatability was assessed by computing the differences between all 3 readings from each subject. Coefficients of repeatability and reproducibility were computed. Results: Mean CCT as measured by each instrument were: 518.53 ± 34.96 µm (range 417.33–592.67) and 516.94 ± 33.60 µm (range 431.67–582.67) for sessions 1 and 2 respectively, with the NCSM; 546.69 ± 36.62 µm (range 457.33–617.00) and 549.78 ± 35.26 µm (range 454.00–618.67) for sessions 1 and 2 respectively, with the USP. The ultrasound CCT measurements were consistently higher than those obtained with the NCSM in both sessions 28.17 ± 19.20 µm (mean ± SD, session 1) and 32.81 ± 14.04 (mean ± SD, session 2). The repeatability coefficient for the NCSM was better in both sessions than those for USP (±10 µm vs ± 12 µm in session 1 and ±8 µm vs ±10 µm in session 2). The reproducibility coefficient with the NCSM was half that with the USP (±21 µm vs ±41 µm). Conclusion: The SP-3000P NCSM is a more precise and reproducible instrument for measurement of CCT than the USP, but both instruments are reliable, useful instruments for measuring CCT.
- Open Access
- 1 May 2011
- Published by Taylor & Francis in Clinical Ophthalmology
- Vol. 5, 661-666
Performance of the PT100 noncontact tonometer in healthy eyes
Background: The purpose of this study was to assess the repeatability and reproducibility of the PT100 noncontact tonometer and to compare its consistency with the Goldmann applanation tonometer (GAT) in measuring intraocular pressure (IOP). Methods: Triplicate IOP measurements were obtained on two separate occasions using the PT100 and GAT from randomly selected eyes in 66 healthy volunteers aged 22 ± 1 years. The repeatability and reproducibility of each techniques was assessed. Agreement between the techniques was statistically quantified using intrasession repeatability for each technique as the basis for comparison. Results: Both techniques returned equal IOP values in the first measurement session (15 ± 3 mmHg). The second session showed a mean difference in average IOP (1 ± 0.71). The 95% limits of agreement between the techniques were –5.2 to 5.5 mmHg and –4.0 to 4.7 mmHg (sessions 1 and 2, respectively). These mean differences were not statistically significant (P > 0.05, paired t-test), with the PT100 underestimating IOP measurement by 1.00 mmHg. The mean intrasession IOP for GAT sessions 1 and 2 was 0 ± 0.90 mmHg and 0.04 ± 1.06 mmHg, respectively, and the corresponding mean IOP measurement difference for the PT100 was –0.06 ± 0.96 and –0.39 ± 0.94 mmHg (sessions 1 and 2, respectively; P > 0.05, paired t-test). Repeatability coefficients for the GAT IOP measurements were 1.8 mmHg and 2.1 mmHg for sessions 1 and 2, while the PT100 repeatability coefficient was 1.9 mmHg and 1.8 mmHg for sessions 1 and 2, respectively. The intrasession repeatability coefficient of both techniques for test–retest differences were within ±5 mmHg. Conclusion: The PT100 noncontact tonometer produced greater repeatability than the GAT in assessment of IOP, whereas GAT resulted in more reproducible results. Both techniques showed a close level of agreement on comparison, with the PT100 underestimating IOP measurement by 1.0 mmHg only, although this was not clinically or statistically significant. Of importance is that the IOP measurements using these techniques could be interchangeable in the IOP range studied here.
- 1 May 2011
- Published by Taylor & Francis in Clinical Ophthalmology
- Vol. 5, 661-666
Open AccessPerformance of the PT100 noncontact tonometer in healthy eyes
Background: The purpose of this study was to assess the repeatability and reproducibility of the PT100 noncontact tonometer and to compare its consistency with the Goldmann applanation tonometer (GAT) in measuring intraocular pressure (IOP). Methods: Triplicate IOP measurements were obtained on two separate occasions using the PT100 and GAT from randomly selected eyes in 66 healthy volunteers aged 22 ± 1 years. The repeatability and reproducibility of each techniques was assessed. Agreement between the techniques was statistically quantified using intrasession repeatability for each technique as the basis for comparison. Results: Both techniques returned equal IOP values in the first measurement session (15 ± 3 mmHg). The second session showed a mean difference in average IOP (1 ± 0.71). The 95% limits of agreement between the techniques were –5.2 to 5.5 mmHg and –4.0 to 4.7 mmHg (sessions 1 and 2, respectively). These mean differences were not statistically significant (P > 0.05, paired t-test), with the PT100 underestimating IOP measurement by 1.00 mmHg. The mean intrasession IOP for GAT sessions 1 and 2 was 0 ± 0.90 mmHg and 0.04 ± 1.06 mmHg, respectively, and the corresponding mean IOP measurement difference for the PT100 was –0.06 ± 0.96 and –0.39 ± 0.94 mmHg (sessions 1 and 2, respectively; P > 0.05, paired t-test). Repeatability coefficients for the GAT IOP measurements were 1.8 mmHg and 2.1 mmHg for sessions 1 and 2, while the PT100 repeatability coefficient was 1.9 mmHg and 1.8 mmHg for sessions 1 and 2, respectively. The intrasession repeatability coefficient of both techniques for test–retest differences were within ±5 mmHg. Conclusion: The PT100 noncontact tonometer produced greater repeatability than the GAT in assessment of IOP, whereas GAT resulted in more reproducible results. Both techniques showed a close level of agreement on comparison, with the PT100 underestimating IOP measurement by 1.0 mmHg only, although this was not clinically or statistically significant. Of importance is that the IOP measurements using these techniques could be interchangeable in the IOP range studied here.