Latest Publications
- 1 June 2001
- Published by Georg Thieme Verlag KG in Endoscopy
- Vol. 33 (6) , 523-528
Freely Accessible Endoscope Channels Improve Efficacy of Cleaning
Background and Study Aims: Inadequate cleaning and disinfection of medical devices, including flexible endoscopes, can result in the transmission of micro-organisms to patients. The aim of this study was to investigate the influence of the design of medical devices on the efficacy of manual cleaning of endoscope channels. Materials and Methods: The investigation was carried out using four endoscopes (two duodenoscopes and two gastroscopes). The air/water channels of one duodenoscope and one gastroscope were freely accessible and could be brushed. The instrumentation and the air/water channels were contaminated with blood containing Enterococcus faecium as a test organism. After manual cleaning of the channels by flushing and, where possible, brushing, the recovery rates for the test organism were studied. Results: The comparable rates for recovery of the test organism after cleaning of the instrumentation channels proved that the method used was reproducible. With regard to the air/water channels, the rate of micro-organisms in the cleaning solution recovered after flushing alone was a maximum of 3 % relative to the rate detected after brushing and flushing. Conclusions: The data collected in the study show that only flushing channels that are not freely accessible resulted in significantly lower (P < 0.001) recovery rates for the test organism. In practice, this means that contamination may remain in the channels, and it shows that the design of a medical device has an important influence on the reprocessing of reusable instruments such as flexible endoscopes.
- 1 June 2001
- Published by Georg Thieme Verlag KG in Endoscopy
- Vol. 33 (6) , 523-528
Freely Accessible Endoscope Channels Improve Efficacy of Cleaning
Background and Study Aims: Inadequate cleaning and disinfection of medical devices, including flexible endoscopes, can result in the transmission of micro-organisms to patients. The aim of this study was to investigate the influence of the design of medical devices on the efficacy of manual cleaning of endoscope channels. Materials and Methods: The investigation was carried out using four endoscopes (two duodenoscopes and two gastroscopes). The air/water channels of one duodenoscope and one gastroscope were freely accessible and could be brushed. The instrumentation and the air/water channels were contaminated with blood containing Enterococcus faecium as a test organism. After manual cleaning of the channels by flushing and, where possible, brushing, the recovery rates for the test organism were studied. Results: The comparable rates for recovery of the test organism after cleaning of the instrumentation channels proved that the method used was reproducible. With regard to the air/water channels, the rate of micro-organisms in the cleaning solution recovered after flushing alone was a maximum of 3 % relative to the rate detected after brushing and flushing. Conclusions: The data collected in the study show that only flushing channels that are not freely accessible resulted in significantly lower (P < 0.001) recovery rates for the test organism. In practice, this means that contamination may remain in the channels, and it shows that the design of a medical device has an important influence on the reprocessing of reusable instruments such as flexible endoscopes.
- 1 June 2001
- Published by Georg Thieme Verlag KG in Endoscopy
- Vol. 33 (6) , 529-532
Determination of Glutaraldehyde Residues on Flexible Endoscopes after Chemothermal Treatment in an Endoscope Washer-Disinfector
Background and Study Aims: Although there are several cases of glutaraldehyde-induced colitis following colonoscopy there is only one study on residues after disinfecting. Our report describes the determination of water-soluble glutaraldehyde residues after processing endoscopes with a glutaraldehyde-containing disinfectant. Materials and Methods: A gastroscope and a colonoscope were processed in a washer-disinfector and then immersed in 20 l of NaCl solution. Glutaraldehyde was determined by high performance liquid chromatography in 84 samples. Samples were taken after immersing the endoscopes for 5 h and 24 h with and without rinsing the channels. Results: After 5 h and with the channels having been rinsed, the median for colonoscopes was only 0.409 mg/l. The highest level of glutaraldehyde was 0.846 mg/l. Conclusions: After endoscopes have been processed in the washer-disinfector there is no risk of a glutaraldehyde-induced colitis, proctitis or diarrhoea.
- 1 June 2001
- Published by Georg Thieme Verlag KG in Endoscopy
- Vol. 33 (6) , 529-532
Determination of Glutaraldehyde Residues on Flexible Endoscopes after Chemothermal Treatment in an Endoscope Washer-Disinfector
Background and Study Aims: Although there are several cases of glutaraldehyde-induced colitis following colonoscopy there is only one study on residues after disinfecting. Our report describes the determination of water-soluble glutaraldehyde residues after processing endoscopes with a glutaraldehyde-containing disinfectant. Materials and Methods: A gastroscope and a colonoscope were processed in a washer-disinfector and then immersed in 20 l of NaCl solution. Glutaraldehyde was determined by high performance liquid chromatography in 84 samples. Samples were taken after immersing the endoscopes for 5 h and 24 h with and without rinsing the channels. Results: After 5 h and with the channels having been rinsed, the median for colonoscopes was only 0.409 mg/l. The highest level of glutaraldehyde was 0.846 mg/l. Conclusions: After endoscopes have been processed in the washer-disinfector there is no risk of a glutaraldehyde-induced colitis, proctitis or diarrhoea.