Sirisha Senthil
Current Organizations / Institutes
The organizations/institutes from the scholar's latest article
Research Subjects
Sirisha Senthil has a total of 19 co-authors. Here are the top co-authors by number of shared publications:
- Harsha Laxmana Rao
Narayana Nethralaya, 63, Bannerghatta Road, Hulimavu, Bangalore 560076, India
6 shared publications - Nikhil S. Choudhari
VST Glaucoma Center, Kallam Anji Reddy Campus
3 shared publications - Aparna Rao
Faculty Glaucoma Services, ICARE Eye Hospital and PG Institute, Noida 201301, UP, India
2 shared publications Chandra S. Garudadri
VST Glaucoma Center L V Prasad Eye Institute Hyderabad India
2 shared publicationsChandra S. Md Garudadri
VST Glaucoma Center, Dr. Kallam Anji Reddy Campus, L.V. Prasad Eye Institute, Hyderabad
1 shared publicationChandra Sekhar Garudadri
1 shared publication
Latest Publications
- 1 January 2018
- Published by Wolters Kluwer Health in RETINAL Cases & Brief Reports
- Vol. 12 (1) , 33-35
25G PARS PLANA VITRECTOMY AND IRIDO-ZONULO-HYALOIDO-VITRECTOMY IN THE MANAGEMENT OF MALIGNANT GLAUCOMA IN PHAKIC EYES FOLLOWING TRABECULECTOMY
To describe a new surgical technique in the management of malignant glaucoma in phakic eyes. We performed 25-gauge pars plana vitrectomy with anterior irido-zonulo-hyaloido-vitrectomy in 2 young phakic patients with malignant glaucoma after trabeculectomy. This technique allowed posteroanterior aqueous communication, relieved aqueous misdirection, helped restoring anterior chamber depth, normalized the intraocular pressure and most importantly helped to preserve the crystalline lens. At 8 months follow-up, both patients have stable visual acuity, normal intraocular pressure, well-functioning blebs, and clear crystalline lens. One can consider 25-gauge pars plana vitrectomy with anterior irido-zonulo-hyaloido-vitrectomy as a safe and effective treatment option for the management of malignant glaucoma in phakic eyes.
- 1 January 2018
- Published by Wolters Kluwer Health in RETINAL Cases & Brief Reports
- Vol. 12 (1) , 33-35
25G PARS PLANA VITRECTOMY AND IRIDO-ZONULO-HYALOIDO-VITRECTOMY IN THE MANAGEMENT OF MALIGNANT GLAUCOMA IN PHAKIC EYES FOLLOWING TRABECULECTOMY
To describe a new surgical technique in the management of malignant glaucoma in phakic eyes. We performed 25-gauge pars plana vitrectomy with anterior irido-zonulo-hyaloido-vitrectomy in 2 young phakic patients with malignant glaucoma after trabeculectomy. This technique allowed posteroanterior aqueous communication, relieved aqueous misdirection, helped restoring anterior chamber depth, normalized the intraocular pressure and most importantly helped to preserve the crystalline lens. At 8 months follow-up, both patients have stable visual acuity, normal intraocular pressure, well-functioning blebs, and clear crystalline lens. One can consider 25-gauge pars plana vitrectomy with anterior irido-zonulo-hyaloido-vitrectomy as a safe and effective treatment option for the management of malignant glaucoma in phakic eyes.
- 1 November 2016
- Published by Wolters Kluwer Health in Journal of Glaucoma
- Vol. 25 (11) , 908-913
How Often the Goldmann Applanation Tonometer Should be Checked for Calibration Error?
Purpose: To evaluate the frequency of Goldmann applanation tonometer (GAT) calibration error (CE) check. Methods: One observer at each of the 3 tertiary care ophthalmic centers was involved. The tonometers were checked at baseline for CE. A tonometer was defined as faulty when CE exceeded 2 mm Hg at any testing level. Faulty GATs were repaired in-house. Subsequent CE check was done once per month for 6 months. GATs were divided into 3 groups: group 1 (G1)≤1 year, group 2 (G2)>1 to 10 years, and group 3 (G3)>10 years of usage. Results: In total, 76 slit-lamp mounted GATs were included. The number of GATs in groups 1 to 3 was 19, 36, and 21, respectively. Seven (9.2%) tonometers were faulty at the baseline. None in G1, 5 in G2, and 16 in G3 demonstrated unacceptable CE over the study course (P<0.01). The survival function of G1 tonometers was 1.0 throughout, whereas that (95% confidence interval) of the G2 and G3 tonometers were 0.97 (0.81-0.99) and 0.76 (0.51-0.89) at 1 month, and 0.86 (0.69-0.93) and 0.23 (0.08-0.43) at 6 months, respectively. The probability of CE development increased with increasing age of the tonometer. The frequency of use of the tonometer was not associated with the development of CE (odds ratio=1.00). Conclusions: CE occurs more frequently in older tonometers. Although newer GATs (<1 y old) can be checked twice yearly, GATs older than a year should be checked at least monthly.
- 1 November 2016
- Published by Wolters Kluwer Health in Journal of Glaucoma
- Vol. 25 (11) , 908-913
How Often the Goldmann Applanation Tonometer Should be Checked for Calibration Error?
Purpose: To evaluate the frequency of Goldmann applanation tonometer (GAT) calibration error (CE) check. Methods: One observer at each of the 3 tertiary care ophthalmic centers was involved. The tonometers were checked at baseline for CE. A tonometer was defined as faulty when CE exceeded 2 mm Hg at any testing level. Faulty GATs were repaired in-house. Subsequent CE check was done once per month for 6 months. GATs were divided into 3 groups: group 1 (G1)≤1 year, group 2 (G2)>1 to 10 years, and group 3 (G3)>10 years of usage. Results: In total, 76 slit-lamp mounted GATs were included. The number of GATs in groups 1 to 3 was 19, 36, and 21, respectively. Seven (9.2%) tonometers were faulty at the baseline. None in G1, 5 in G2, and 16 in G3 demonstrated unacceptable CE over the study course (P<0.01). The survival function of G1 tonometers was 1.0 throughout, whereas that (95% confidence interval) of the G2 and G3 tonometers were 0.97 (0.81-0.99) and 0.76 (0.51-0.89) at 1 month, and 0.86 (0.69-0.93) and 0.23 (0.08-0.43) at 6 months, respectively. The probability of CE development increased with increasing age of the tonometer. The frequency of use of the tonometer was not associated with the development of CE (odds ratio=1.00). Conclusions: CE occurs more frequently in older tonometers. Although newer GATs (<1 y old) can be checked twice yearly, GATs older than a year should be checked at least monthly.
- 1 October 2016
- Published by Wolters Kluwer Health in Journal of Glaucoma
- Vol. 25 (10) , 812-814
A Simplified (Screening) Approach to Check the Calibration Status of the Goldmann Applanation Tonometer
To investigate whether a lower-end calibration error check alone can identify a Goldmann applanation tonometer (GAT) with acceptable calibration error as per Asia-Pacific Glaucoma Society (APGS) and World Glaucoma Association (WGA) guidelines.This was a multicentric prospective study in a tertiary eye-care institutional setting. Seventy-six slit-lamp-mounted Haag-Streit GATs (Model AT 900 C/M) were included in the study. All GATs were checked for calibration error at 0, 20, and 60 mm Hg testing levels as recommended by the manufacturer. The ability of 2 screening criteria (calibration error within ±1 or ±2 mm Hg, respectively, at the 0 mm Hg testing level without using the calibration error check weight bar) to identify a tonometer within acceptable calibration limits defined by APGS (calibration error within ±2, ±3, and ±4 mm Hg at 0, 20, and 60 mm Hg testing levels, respectively) and WGA (calibration error within ±1 mm Hg at all testing levels) was determined.The negative likelihood ratio of the screening criteria RESULTS:: Five (6.6%) and 42 (55%) GATs were outside of the APGS and the WGA calibration limits, respectively. Negative likelihood ratios of the first and the second screening criteria at the APGS definition of calibration error tolerance were 0.11 and 0.25, respectively, and at that of the WGA definition of calibration error tolerance were 0.38 and 0.90, respectively.The screening criteria of calibration error within ±1 mm Hg at the 0 mm Hg testing level seem to be clinically useful to detect GAT with an acceptable calibration error as per the APGS guideline.
- 1 October 2016
- Published by Wolters Kluwer Health in Journal of Glaucoma
- Vol. 25 (10) , 812-814
A Simplified (Screening) Approach to Check the Calibration Status of the Goldmann Applanation Tonometer
To investigate whether a lower-end calibration error check alone can identify a Goldmann applanation tonometer (GAT) with acceptable calibration error as per Asia-Pacific Glaucoma Society (APGS) and World Glaucoma Association (WGA) guidelines.This was a multicentric prospective study in a tertiary eye-care institutional setting. Seventy-six slit-lamp-mounted Haag-Streit GATs (Model AT 900 C/M) were included in the study. All GATs were checked for calibration error at 0, 20, and 60 mm Hg testing levels as recommended by the manufacturer. The ability of 2 screening criteria (calibration error within ±1 or ±2 mm Hg, respectively, at the 0 mm Hg testing level without using the calibration error check weight bar) to identify a tonometer within acceptable calibration limits defined by APGS (calibration error within ±2, ±3, and ±4 mm Hg at 0, 20, and 60 mm Hg testing levels, respectively) and WGA (calibration error within ±1 mm Hg at all testing levels) was determined.The negative likelihood ratio of the screening criteria RESULTS:: Five (6.6%) and 42 (55%) GATs were outside of the APGS and the WGA calibration limits, respectively. Negative likelihood ratios of the first and the second screening criteria at the APGS definition of calibration error tolerance were 0.11 and 0.25, respectively, and at that of the WGA definition of calibration error tolerance were 0.38 and 0.90, respectively.The screening criteria of calibration error within ±1 mm Hg at the 0 mm Hg testing level seem to be clinically useful to detect GAT with an acceptable calibration error as per the APGS guideline.
- 1 October 2014
- Published by Wiley in Acta Ophthalmologica
- Vol. 93 (2) , e105-10
Atypical birefringence pattern and the diagnostic ability of scanning laser polarimetry with enhanced corneal compensation in glaucoma
To evaluate the effect of typical scan score (TSS), when within the acceptable limits, on the diagnostic performance of retinal nerve fibre layer (RNFL) parameters with the enhanced corneal compensation (ECC) protocol of scanning laser polarimetry (SLP) in glaucoma. In a cross-sectional study, 203 eyes of 160 glaucoma patients and 140 eyes of 104 control subjects underwent RNFL imaging with the ECC protocol of SLP. TSS was used to quantify atypical birefringence pattern (ABP) images. Influence of TSS on the diagnostic ability of SLP parameters was evaluated by receiver operating characteristic (ROC) regression models after adjusting for the effect of disease severity [based on mean deviation (MD)] on standard automated perimetry). Diagnostic abilities of all RNFL parameters of SLP increased when the TSS values were higher. This effect was statistically significant for TSNIT (coefficient: 0.08, p<0.001) and inferior average parameters (coefficient: 0.06, p=0.002) but not for nerve fibre indicator (NFI, coefficient: 0.03, p=0.21). In early glaucoma (MD of -5 dB), predicted area under ROC curve (AUC) for TSNIT average parameter improved from 0.642 at a TSS of 90 to 0.845 at a TSS of 100. In advanced glaucoma (MD of -15 dB), AUC for TSNIT average improved from 0.832 at a TSS of 90 to 0.947 at 100. Diagnostic performances of TSNIT and inferior average RNFL parameters with ECC protocol of SLP were significantly influenced by TSS even when the TSS values were within the acceptable limits. Diagnostic ability of NFI was unaffected by TSS values.
- 1 October 2014
- Published by Wiley in Acta Ophthalmologica
- Vol. 93 (2) , e105-10
Atypical birefringence pattern and the diagnostic ability of scanning laser polarimetry with enhanced corneal compensation in glaucoma
To evaluate the effect of typical scan score (TSS), when within the acceptable limits, on the diagnostic performance of retinal nerve fibre layer (RNFL) parameters with the enhanced corneal compensation (ECC) protocol of scanning laser polarimetry (SLP) in glaucoma. In a cross-sectional study, 203 eyes of 160 glaucoma patients and 140 eyes of 104 control subjects underwent RNFL imaging with the ECC protocol of SLP. TSS was used to quantify atypical birefringence pattern (ABP) images. Influence of TSS on the diagnostic ability of SLP parameters was evaluated by receiver operating characteristic (ROC) regression models after adjusting for the effect of disease severity [based on mean deviation (MD)] on standard automated perimetry). Diagnostic abilities of all RNFL parameters of SLP increased when the TSS values were higher. This effect was statistically significant for TSNIT (coefficient: 0.08, p<0.001) and inferior average parameters (coefficient: 0.06, p=0.002) but not for nerve fibre indicator (NFI, coefficient: 0.03, p=0.21). In early glaucoma (MD of -5 dB), predicted area under ROC curve (AUC) for TSNIT average parameter improved from 0.642 at a TSS of 90 to 0.845 at a TSS of 100. In advanced glaucoma (MD of -15 dB), AUC for TSNIT average improved from 0.832 at a TSS of 90 to 0.947 at 100. Diagnostic performances of TSNIT and inferior average RNFL parameters with ECC protocol of SLP were significantly influenced by TSS even when the TSS values were within the acceptable limits. Diagnostic ability of NFI was unaffected by TSS values.
- 1 September 2012
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 38 (9) , 1583-1589
Posterior chamber toric phakic intraocular lenses for myopic astigmatism: First experience in India
To evaluate the efficacy and safety of the toric Implantable Collamer Lens phakic intraocular lens (pIOL) in the management of myopic astigmatism in Indian eyes. Tertiary eyecare center, South India. Retrospective interventional case series. The study included consecutive patients with myopic astigmatism with a minimum follow-up of 1 year. Safety, efficacy, predictability, and adverse events were assessed preoperatively and 12 months postoperatively. The study enrolled 110 eyes (65 patients). The mean corrected distance visual acuity was 0.23 logMAR ± 0.23 (SD) preoperatively and 0.18 ± 0.18 logMAR 12 months postoperatively (P<.001). The mean refractive cylinder decreased from 2.3 ± 1.1 diopters (D) to 0.4 ± 0.6 D, respectively (P<.001). Postoperatively, the uncorrected distance visual acuity improved to 20/20 in 15 eyes (13.6%) and 20/40 or better in 72 eyes (65.4%). Fourteen eyes (12.7%) gained 2 or more lines of corrected distance visual acuity, 21 eyes (19.1%) gained 1 line, 61 eyes (55.5%) had no change, and 1 eye (0.9%) lost 2 lines. Predictability within ±0.50 D of cylinder was achieved in 77 eyes (70 %) and within ±1.00 D in 97 eyes (88.2%). Twelve months postoperatively, the safety index was 0.75 and the efficacy index was 1.04. Endothelial cell loss at the end of 12 months was 10.3%. Clinically significant cataract developed in 1 eye (0.9%). Implantation of posterior chamber toric pIOLs to correct myopic astigmatism was effective, safe, and predictable.
- 1 September 2012
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 38 (9) , 1583-1589
Posterior chamber toric phakic intraocular lenses for myopic astigmatism: First experience in India
To evaluate the efficacy and safety of the toric Implantable Collamer Lens phakic intraocular lens (pIOL) in the management of myopic astigmatism in Indian eyes. Tertiary eyecare center, South India. Retrospective interventional case series. The study included consecutive patients with myopic astigmatism with a minimum follow-up of 1 year. Safety, efficacy, predictability, and adverse events were assessed preoperatively and 12 months postoperatively. The study enrolled 110 eyes (65 patients). The mean corrected distance visual acuity was 0.23 logMAR ± 0.23 (SD) preoperatively and 0.18 ± 0.18 logMAR 12 months postoperatively (P<.001). The mean refractive cylinder decreased from 2.3 ± 1.1 diopters (D) to 0.4 ± 0.6 D, respectively (P<.001). Postoperatively, the uncorrected distance visual acuity improved to 20/20 in 15 eyes (13.6%) and 20/40 or better in 72 eyes (65.4%). Fourteen eyes (12.7%) gained 2 or more lines of corrected distance visual acuity, 21 eyes (19.1%) gained 1 line, 61 eyes (55.5%) had no change, and 1 eye (0.9%) lost 2 lines. Predictability within ±0.50 D of cylinder was achieved in 77 eyes (70 %) and within ±1.00 D in 97 eyes (88.2%). Twelve months postoperatively, the safety index was 0.75 and the efficacy index was 1.04. Endothelial cell loss at the end of 12 months was 10.3%. Clinically significant cataract developed in 1 eye (0.9%). Implantation of posterior chamber toric pIOLs to correct myopic astigmatism was effective, safe, and predictable.
- 26 June 2012
- Published by Springer Nature in International Ophthalmology
- Vol. 32 (6) , 603-605
Refractive surprise following implantation of correct powered intraocular lens—a real surprise!
We report a case of a 60-year-old male with unsatisfactory visual outcome following an uneventful phacoemulsification surgery for the left eye at our hospital. The evaluation of his anterior segment revealed a clear cornea, a well-positioned in-the-bag intraocular lens (IOL) and normal posterior segment. His vision in the left eye was 20/160 improving to 20/20 with +6.5 D Sphere. Repeat biometry of the left eye revealed IOL power of 22.0 D, which was same as the power of the IOL implanted during the cataract surgery. The patient underwent IOL explantation and implantation of another foldable IOL of 22.0 D. On follow-up he had an uncorrected visual acuity of 20/20. On evaluation with the Scheiner optical system, the explanted lens was found to have power of +13.92 D. This case highlights the rare possibility of postoperative refractory surprise due to incorrect labelling of the IOL by the manufacturer.
- 26 June 2012
- Published by Springer Nature in International Ophthalmology
- Vol. 32 (6) , 603-605
Refractive surprise following implantation of correct powered intraocular lens—a real surprise!
We report a case of a 60-year-old male with unsatisfactory visual outcome following an uneventful phacoemulsification surgery for the left eye at our hospital. The evaluation of his anterior segment revealed a clear cornea, a well-positioned in-the-bag intraocular lens (IOL) and normal posterior segment. His vision in the left eye was 20/160 improving to 20/20 with +6.5 D Sphere. Repeat biometry of the left eye revealed IOL power of 22.0 D, which was same as the power of the IOL implanted during the cataract surgery. The patient underwent IOL explantation and implantation of another foldable IOL of 22.0 D. On follow-up he had an uncorrected visual acuity of 20/20. On evaluation with the Scheiner optical system, the explanted lens was found to have power of +13.92 D. This case highlights the rare possibility of postoperative refractory surprise due to incorrect labelling of the IOL by the manufacturer.
- 1 January 2012
- Published by Elsevier in Ophthalmology
- Vol. 119 (1) , 204
Author reply