KLKristin Lamvik

Kristin Lamvik

*This is an algorithmically generated author record based on publicly available information.

Current Organizations / Institutes

The organizations/institutes from the scholar's latest article

Previous Organizations / Institutes

The organizations / institutes from the scholar's other articles in the past ten years


Kristin Lamvik has a total of 6 co-authors. Here are the top co-authors by number of shared publications:

  • Richard D. Jones

    New Zealand Brain Research Institute, Christchurch, New Zealand

    2 shared publications
  • E. Guiu Hernandez

    The University of Canterbury Rose Centre for Stroke Recovery and Research; Christchurch New Zealand

    1 shared publication
  • Kerstin Erfmann

    The University of Canterbury Rose Centre for Stroke Recovery and Research, Christchurch, New Zealand; Department of Communication Disorders, University of Canterbury, Christchurch, New Zealand; New Zealand Brain Research Institute, Christchurch, New Zealand

    1 shared publication
  • M.-L. Huckabee

    The University of Canterbury Rose Centre for Stroke Recovery and Research; Christchurch New Zealand

    1 shared publication
  • Maggie-Lee Huckabee

    The University of Canterbury Rose Centre for Stroke Recovery and Research, Christchurch, New Zealand; Department of Communication Disorders, University of Canterbury, Christchurch, New Zealand; New Zealand Brain Research Institute, Christchurch, New Zealand

    1 shared publication
  • Sonja Sauer

    Department of Communication Disorders, University of Canterbury, Christchurch, New Zealand; New Zealand Brain Research Institute, Christchurch, New Zealand

    1 shared publication

Latest Publications

  • Characterization and correction of pressure drift in the ManoScan™high-resolution manometry system:In vitroandin vivo

    by K. Lamvik, E. Guiu Hernandez, R. Jones, M.-L. Huckabee

    A substantial pressure drift in high-resolution manometry (HRM) has been reported; however, fundamental questions remain regarding the origin and management of this drift. The aim of this study was to provide critical in-depth analyses of ManoScan(™) HRM drift in vitro and in vivo. A total of sixteen 15-min studies and twelve 5-h studies were performed in a water bath at 37 °C at 4.0 cm depth (2.9 mmHg) with ESO and ESO Z catheters. Six 5-h in vitro studies were performed similarly at a depth of 9.0 cm (6.6 mmHg). Eight 15-min studies and nine 8-h in vivo studies were performed with healthy participants. Two correction methods - thermal compensation (TC) and interpolated thermal compensation (ITC) - were tested. Overall pressure drift varied both between studies (p < 0.01) and within sensors (p < 0.01). Drift resulted from thermal shock, an initial pressure change at intubation, and baseline drift, a linear drift over time (R(2) > 0.96). Contrary to previous reports, there was no correlation between drift and average (r = -0.02) or maximum pressure exposure (r = -0.05). Following data correction, ITC had the lowest median error but persisted with a maximum error of 2.5 mmHg (IQR = 3.0). The substantial drift in the ManoScan(™) HRM system is highly variable and not corrected via the standard operating instructions. ITC has superior performance but requires communication with the manufacturer to enable this option. This has a substantial impact on clinical diagnosis, utility of existing normative data, and future research of HRM.


  • The capacity for volitional control of pharyngeal swallowing in healthy adults

    by Kristin Lamvik, Richard Jones, Sonja Sauer, Kerstin Erfmann, Maggie-Lee Huckabee