Willie Gerhardt
Current Organizations / Institutes
The organizations/institutes from the scholar's latest article
Research Subjects
Willie Gerhardt has a total of 26 co-authors. Here are the top co-authors by number of shared publications:
H A Katus
2 shared publicationsM Müller-Bardorff
2 shared publicationsP O Collinson
2 shared publicationsR Leinberger
2 shared publicationsR Zerback
2 shared publicationsA N Laggner
1 shared publication
Latest Publications
- Open Access
- 1 June 2000
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 46 (6) , 817-821
Troponin T and I Assays Show Decreased Concentrations in Heparin Plasma Compared with Serum: Lower Recoveries in Early than in Late Phases of Myocardial Injury
Background: Heparinized plasma samples allow more rapid analysis than serum samples, but preliminary studies showed lower cardiac troponin T (cTnT) results in plasma. We undertook a multicenter study to characterize this effect for cTnT and cardiac troponin I (cTnI). Methods: Blood samples were collected with and without heparin at five hospitals. cTnT was measured by a “third generation” assay (Elecsys®), and cTnI was measured by a commercial immunoassay (IMMULITE®). Results: Mean cTnT was 15% lower in heparin sampling tubes than in serum. Measured concentrations of cardiac troponins also decreased with increasing heparin concentrations added to sera. Heparin-induced losses were greater in early than in late phases after onset of chest pain. Addition of heparin (∼100 IU/mL) to serial samples from nine acute myocardial infarction patients produced mean cTnT losses of 33% at 1–12 h after onset of chest pain, 17% at 13–48 h, and 7% after 48 h. The changing heparin effects were seen for both cTnT and cTnI during time courses of individual patients with myocardial infarction. Conclusion: We suggest that binding of heparin to troponins decreases immunoreactivity, especially in early phases of myocardial injury. The resulting losses may depend on the antibodies used in each troponin assay.
- 1 June 2000
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 46 (6) , 817-821
Open AccessTroponin T and I Assays Show Decreased Concentrations in Heparin Plasma Compared with Serum: Lower Recoveries in Early than in Late Phases of Myocardial Injury
Background: Heparinized plasma samples allow more rapid analysis than serum samples, but preliminary studies showed lower cardiac troponin T (cTnT) results in plasma. We undertook a multicenter study to characterize this effect for cTnT and cardiac troponin I (cTnI). Methods: Blood samples were collected with and without heparin at five hospitals. cTnT was measured by a “third generation” assay (Elecsys®), and cTnI was measured by a commercial immunoassay (IMMULITE®). Results: Mean cTnT was 15% lower in heparin sampling tubes than in serum. Measured concentrations of cardiac troponins also decreased with increasing heparin concentrations added to sera. Heparin-induced losses were greater in early than in late phases after onset of chest pain. Addition of heparin (∼100 IU/mL) to serial samples from nine acute myocardial infarction patients produced mean cTnT losses of 33% at 1–12 h after onset of chest pain, 17% at 13–48 h, and 7% after 48 h. The changing heparin effects were seen for both cTnT and cTnI during time courses of individual patients with myocardial infarction. Conclusion: We suggest that binding of heparin to troponins decreases immunoreactivity, especially in early phases of myocardial injury. The resulting losses may depend on the antibodies used in each troponin assay.
- 1 April 2000
- Published by Wiley
- Vol. 124 (4) , 583-7
Analytical and clinical performance of an improved qualitative troponin T rapid test in laboratories and critical care units.
To evaluate the performance of a visual troponin T rapid test in the hands of nontraditionally trained personnel of 2 critical care units in comparison to 3 laboratories. Method comparisons of the troponin T rapid test versus cardiac troponin T enzyme-linked immunosorbent assay were performed with 804 samples from 510 patients with suspected acute coronary syndromes. Cross-reactivity with skeletal troponin T was studied up to 5000 microg/L. Laboratories and critical care units obtained comparable results in the analytical cutoff of the test (0.11 and 0. 10 microg/L) and in the diagnostic sensitivities in the detection of acute myocardial infarction (96% and 93% after 8 hours) and of high-risk patients with unstable angina pectoris (100% and 100%). Different percentages of false-positive results (0.2% and 3%) were found, which may reflect different objectives and strategies in these hospital units. The cross-reactivity with skeletal troponin T was less than 0.01%. The troponin T rapid test gives reliable results not only when used by laboratory personnel experienced in the execution of analytical methods, but also in the hands of nurses and physicians working in clinical units outside the laboratory.
- 1 April 2000
- Published by Wiley
- Vol. 124 (4) , 583-7
Analytical and clinical performance of an improved qualitative troponin T rapid test in laboratories and critical care units.
To evaluate the performance of a visual troponin T rapid test in the hands of nontraditionally trained personnel of 2 critical care units in comparison to 3 laboratories. Method comparisons of the troponin T rapid test versus cardiac troponin T enzyme-linked immunosorbent assay were performed with 804 samples from 510 patients with suspected acute coronary syndromes. Cross-reactivity with skeletal troponin T was studied up to 5000 microg/L. Laboratories and critical care units obtained comparable results in the analytical cutoff of the test (0.11 and 0. 10 microg/L) and in the diagnostic sensitivities in the detection of acute myocardial infarction (96% and 93% after 8 hours) and of high-risk patients with unstable angina pectoris (100% and 100%). Different percentages of false-positive results (0.2% and 3%) were found, which may reflect different objectives and strategies in these hospital units. The cross-reactivity with skeletal troponin T was less than 0.01%. The troponin T rapid test gives reliable results not only when used by laboratory personnel experienced in the execution of analytical methods, but also in the hands of nurses and physicians working in clinical units outside the laboratory.
- 1 January 1999
- Published by Taylor & Francis in Scandinavian Journal of Clinical and Laboratory Investigation
- Vol. 59, 83-89
Can Troponin T replace CK MBmass as “gold standard” for Acute Myocardial Infarction (“AMI”)?
- 1 January 1999
- Published by Taylor & Francis in Scandinavian Journal of Clinical and Laboratory Investigation
- Vol. 59, 83-89
Can Troponin T replace CK MBmass as “gold standard” for Acute Myocardial Infarction (“AMI”)?
- 1 July 1996
- Vol. 34 (7) , 591-8
Multicentre evaluation of an immunological rapid test for the detection of troponin T in whole blood samples.
In a multicentre study we assessed the analytical and diagnostic performance of a rapid test (TROPT rapid test, Boehringer Mannheim; in the USA: CARDIACT) for cardiac troponin T compared to quantitative troponin T ELISA and creatine kinase-MB mass determinations. The rapid test requires 150 microliters of heparinized or EDTA whole blood; serum is not suitable. Interference testing with biotin, haemoglobin and 27 standard drugs yielded no significant influence in the physiological range. Skeletal muscle troponin T concentrations > or = 40 micrograms/l gave positive results with the rapid test. We used the rapid test for 369 samples from 203 patients with suspected acute coronary syndromes and compared the results to troponin T ELISA and creatine kinase-MB mass. 90 patients (44%) were primarily classified as having myocardial infarction by the WHO criteria. Twenty-two (20%) of the 113 non-myocardial infarction patients were unstable angina pectoris cases showing increased troponin T ELISA but not increased creatine kinase-MB mass values. Consequently, these were classified as minor myocardial damage cases. The rapid test was positive in 99% of all samples with a troponin T ELISA value > or = 0.30 micrograms/l and negative in 95 to 96% of all samples below this value. Diagnostic sensitivities for the detection of acute myocardial infarction within the first 12 hours after onset of pain were the same, 90%, for the rapid test, troponin T ELISA and creatine kinase-MB mass. After 48 hours, diagnostic sensitivity of creatine kinase-MB mass sharply decreased whereas that of the troponin T assays remained close to 100% beyond 72 hours after onset of symptoms. Diagnostic specificities for acute myocardial infarction (WHO) of all markers remained between 80 and 100% over this time. The diagnostic sensitivity of the rapid test for the detection of high risk unstable angina pectoris patients with minor myocardial damage was nearly the same as for troponin T ELISA. A major advantage of the rapid test is the ease of use and 20 minute turn around time. This facilitates the detection of increased troponin T at alternate sites.
- 1 July 1996
- Vol. 34 (7) , 591-8
Multicentre evaluation of an immunological rapid test for the detection of troponin T in whole blood samples.
In a multicentre study we assessed the analytical and diagnostic performance of a rapid test (TROPT rapid test, Boehringer Mannheim; in the USA: CARDIACT) for cardiac troponin T compared to quantitative troponin T ELISA and creatine kinase-MB mass determinations. The rapid test requires 150 microliters of heparinized or EDTA whole blood; serum is not suitable. Interference testing with biotin, haemoglobin and 27 standard drugs yielded no significant influence in the physiological range. Skeletal muscle troponin T concentrations > or = 40 micrograms/l gave positive results with the rapid test. We used the rapid test for 369 samples from 203 patients with suspected acute coronary syndromes and compared the results to troponin T ELISA and creatine kinase-MB mass. 90 patients (44%) were primarily classified as having myocardial infarction by the WHO criteria. Twenty-two (20%) of the 113 non-myocardial infarction patients were unstable angina pectoris cases showing increased troponin T ELISA but not increased creatine kinase-MB mass values. Consequently, these were classified as minor myocardial damage cases. The rapid test was positive in 99% of all samples with a troponin T ELISA value > or = 0.30 micrograms/l and negative in 95 to 96% of all samples below this value. Diagnostic sensitivities for the detection of acute myocardial infarction within the first 12 hours after onset of pain were the same, 90%, for the rapid test, troponin T ELISA and creatine kinase-MB mass. After 48 hours, diagnostic sensitivity of creatine kinase-MB mass sharply decreased whereas that of the troponin T assays remained close to 100% beyond 72 hours after onset of symptoms. Diagnostic specificities for acute myocardial infarction (WHO) of all markers remained between 80 and 100% over this time. The diagnostic sensitivity of the rapid test for the detection of high risk unstable angina pectoris patients with minor myocardial damage was nearly the same as for troponin T ELISA. A major advantage of the rapid test is the ease of use and 20 minute turn around time. This facilitates the detection of increased troponin T at alternate sites.