    A multicentre prospective observational study comparing arterial blood gas values to those obtained by pulse oximeters used in adult patients attending Australian and New Zealand hospitals

    by Janine Pilcher, Laura Ploen, Steve McKinstry, George Bardsley, Jimmy Chien, Lesley Howard, Sharon Lee, Lutz Beckert, Maureen Swanney, Mark Weatherall, et al.

    Background: Pulse oximetry is widely used in the clinical setting. The purpose of this validation study was to investigate the level of agreement between oxygen saturations measured by pulse oximeter (SpO2) and arterial blood gas (SaO2) in a range of oximeters in clinical use in Australia and New Zealand. Methods: Paired SpO2 and SaO2 measurements were collected from 400 patients in one Australian and two New Zealand hospitals. The ages of the patients ranged from 18 to 95 years. Bias and limits of agreement were estimated. Sensitivity and specificity for detecting hypoxaemia, defined as SaO2 < 90%, were also estimated. Results: The majority of participants were recruited from the Outpatient, Ward or High Dependency Unit setting. Bias, oximeter-measured minus arterial blood gas-measured oxygen saturation, was − 1.2%, with limits of agreement − 4.4 to 2.0%. SpO2 was at least 4% lower than SaO2 for 10 (2.5%) of the participants and SpO2 was at least 4% higher than the SaO2 in 3 (0.8%) of the participants. None of the participants with a SpO2 ≥ 92% were hypoxaemic, defined as SaO2 < 90%. There were no clinically significant differences in oximetry accuracy in relation to clinical characteristics or oximeter brand. Conclusions: In the majority of the participants, pulse oximetry was an accurate method to assess SaO2 and had good performance in detecting hypoxaemia. However, in a small proportion of participants, differences between SaO2 and SpO2 could have clinical relevance in terms of patient monitoring and management. A SpO2 ≥ 92% indicates that hypoxaemia, defined as a SaO2 < 90%, is not present. Trial registration: Australian and New Zealand Clinical Trials Registry (ACTRN12614001257651). Date of registration: 2/12/2014.


  • Open Access

    Oxygen versus air-driven nebulisers for exacerbations of chronic obstructive pulmonary disease: a randomised controlled trial

    by George Bardsley, Janine Pilcher, Steven McKinstry, Philippa Shirtcliffe, James Berry, James Fingleton, Mark Weatherall, Richard Beasley

    In exacerbations of chronic obstructive pulmonary disease, administration of high concentrations of oxygen may cause hypercapnia and increase mortality compared with oxygen titrated, if required, to achieve an oxygen saturation of 88–92%. Optimally titrated oxygen regimens require two components: titrated supplemental oxygen to achieve the target oxygen saturation and, if required, bronchodilators delivered by air-driven nebulisation. The effect of repeated air vs oxygen-driven bronchodilator nebulisation in acute exacerbations of chronic obstructive pulmonary disease is unknown. We aimed to compare the effects of air versus oxygen-driven bronchodilator nebulisation on arterial carbon dioxide tension in exacerbations of chronic obstructive pulmonary disease. A parallel group double-blind randomised controlled trial in 90 hospital in-patients with an acute exacerbation of COPD. Participants were randomised to receive two 2.5 mg salbutamol nebulisers, both driven by air or oxygen at 8 L/min, each delivered over 15 min with a 5 min interval in-between. The primary outcome measure was the transcutaneous partial pressure of carbon dioxide at the end of the second nebulisation (35 min). The primary analysis used a mixed linear model with fixed effects of the baseline PtCO2, time, the randomised intervention, and a time by intervention interaction term; to estimate the difference between randomised treatments at 35 min. Analysis was by intention-to-treat. Oxygen-driven nebulisation was terminated in one participant after 27 min when the PtCO2 rose by > 10 mmHg, a predefined safety criterion. The mean (standard deviation) change in PtCO2 at 35 min was 3.4 (1.9) mmHg and 0.1 (1.4) mmHg in the oxygen and air groups respectively, difference (95% confidence interval) 3.3 mmHg (2.7 to 3.9), p < 0.001. The proportion of patients with a PtCO2 change ≥4 mmHg during the intervention was 18/45 (40%) and 0/44 (0%) for oxygen and air groups respectively. Oxygen-driven nebulisation leads to an increase in PtCO2 in exacerbations of COPD. We propose that air-driven bronchodilator nebulisation is preferable to oxygen-driven nebulisation in exacerbations of COPD. Australian New Zealand Clinical Trials Registry number ACTRN12615000389505 . Registration confirmed on 28/4/15.