Sulafa Ahmed
Current Organizations / Institutes
The organizations/institutes from the scholar's latest article
Previous Organizations / Institutes
The organizations / institutes from the scholar's other articles in the past ten years
Research Subjects
Sulafa Ahmed has a total of 4 co-authors. Here are the top co-authors by number of shared publications:
Faihaa Eltigani
U of G: University of Gezira
3 shared publicationsMaged Yahya
Electronics Engineering Department, Faculty of Engineering and Technology, University of Gezira, Wad Madani, Sudan
2 shared publicationsMawahib Ahmed
Electronics Engineering Department, Faculty of Engineering and Technology, University of Gezira, Wad Madani, Sudan
2 shared publications- Noha S Hassan
Biomedical Engineering Department, Faculty of Engineering, Cairo University, Giza, 12613, Egypt
1 shared publication
Latest Publications
- Open Access
- 14 December 2023
- Published by Springer Nature in Research Square
Utilizing a Floating Sleeve Antenna for Interstitial Microwave Hyperthermia Treatment of Superficial and Deep-seated Breast Tumors
Purpose: Breast cancer is a leading cause of mortality among women worldwide. While chemotherapy and radiotherapy are commonly used treatment modalities, they have several side effects that can negatively impact the patient. In this study, we explore the application of a floating sleeve antenna for interstitial microwave hyperthermia for breast tumors. Method: Two distinct 3D breast models were employed: a simplified semi-spherical and a highly detailed anatomical model. Both models featured superficial and deep-seated tumors. The floating sleeve antenna was inserted into the breast models, and simulations were con-ducted to assess the specific absorption rate (SAR) and temperature distribution. Results: In the simple model, the superficial tumor reached a maximum temperature of 50.7 ºC, while the deep-seated tumor maintained a uniform temperature of 48.9 ºC. In contrast, the realistic model achieved a maximum temperature of 46.2 ºC for the superficial tumor and 41.47 ºC for the deep-seated tumor within the same 15-minute treatment period. These results demonstrate the ability of the antenna to attain the desired hyperthermia temperatures (above 39 °C to 45 °C). Conclusion: microwave hyperthermia using a floating sleeve antenna has the potential to be an effective and safe alternative treatment for breast cancer with minimal side effects
- 14 December 2023
- Published by Springer Nature in Research Square
Open AccessUtilizing a Floating Sleeve Antenna for Interstitial Microwave Hyperthermia Treatment of Superficial and Deep-seated Breast Tumors
Purpose: Breast cancer is a leading cause of mortality among women worldwide. While chemotherapy and radiotherapy are commonly used treatment modalities, they have several side effects that can negatively impact the patient. In this study, we explore the application of a floating sleeve antenna for interstitial microwave hyperthermia for breast tumors. Method: Two distinct 3D breast models were employed: a simplified semi-spherical and a highly detailed anatomical model. Both models featured superficial and deep-seated tumors. The floating sleeve antenna was inserted into the breast models, and simulations were con-ducted to assess the specific absorption rate (SAR) and temperature distribution. Results: In the simple model, the superficial tumor reached a maximum temperature of 50.7 ºC, while the deep-seated tumor maintained a uniform temperature of 48.9 ºC. In contrast, the realistic model achieved a maximum temperature of 46.2 ºC for the superficial tumor and 41.47 ºC for the deep-seated tumor within the same 15-minute treatment period. These results demonstrate the ability of the antenna to attain the desired hyperthermia temperatures (above 39 °C to 45 °C). Conclusion: microwave hyperthermia using a floating sleeve antenna has the potential to be an effective and safe alternative treatment for breast cancer with minimal side effects
- 15 April 2022
- Published by Springer Nature in Physical and Engineering Sciences in Medicine
- Vol. 45 (2) , 569-575
Modeling of interstitial microwave hyperthermia for hepatic tumors using floating sleeve antenna
Microwave hyperthermia is a treatment modality that uses microwaves to destroy cancer cells by increasing their temperature to 41–45 °C. This study aims to design, model, and simulate a microwave sleeve antenna for hepatic (liver) hyperthermia. A floating sleeve antenna with 0.5 w input power was designed to resonate at 2.45 GHz. The antenna was tested in six different 3D liver models. The models were varied from a very simple model without a tumor and blood vessels to complex models that contain realistic tumors and blood vessels. To test the capability of the proposed antenna for heating the interstitial tumors, the size, shape, and location of the tumor were changed. The specific absorption rate (SAR) and temperature were calculated for each model. The tumors’ temperature was elevated between 43 and 45 °C, while the temperature of the surrounding tissues was below 41 °C. The Specific Absorption Rate (SAR) was between 29 and 30 W/kg in the tumors and below 24 W/Kg in the surrounding tissues. The return loss of the antenna varied from − 45 to − 25 dB for the six models. The antenna could heat hepatic tumors with different sizes and locations. The heating process was performed in a short time by using a very low input power compared to all previous studies.
- 15 April 2022
- Published by Springer Nature in Physical and Engineering Sciences in Medicine
- Vol. 45 (2) , 569-575
Modeling of interstitial microwave hyperthermia for hepatic tumors using floating sleeve antenna
Microwave hyperthermia is a treatment modality that uses microwaves to destroy cancer cells by increasing their temperature to 41–45 °C. This study aims to design, model, and simulate a microwave sleeve antenna for hepatic (liver) hyperthermia. A floating sleeve antenna with 0.5 w input power was designed to resonate at 2.45 GHz. The antenna was tested in six different 3D liver models. The models were varied from a very simple model without a tumor and blood vessels to complex models that contain realistic tumors and blood vessels. To test the capability of the proposed antenna for heating the interstitial tumors, the size, shape, and location of the tumor were changed. The specific absorption rate (SAR) and temperature were calculated for each model. The tumors’ temperature was elevated between 43 and 45 °C, while the temperature of the surrounding tissues was below 41 °C. The Specific Absorption Rate (SAR) was between 29 and 30 W/kg in the tumors and below 24 W/Kg in the surrounding tissues. The return loss of the antenna varied from − 45 to − 25 dB for the six models. The antenna could heat hepatic tumors with different sizes and locations. The heating process was performed in a short time by using a very low input power compared to all previous studies.
- Open Access
- 13 January 2022
- Published by Springer Nature in Research Square
Modeling of Interstitial Microwave Hyperthermia for Hepatic Tumors Using Floating Sleeve Antenna
PurposeMicrowave hyperthermia is a treatment modality that uses microwaves to destroy cancer cells by increasing their temperature to 41- 45°C. This study aims to design, modeling, and simulation of a microwave sleeve antenna for hepatic (liver) hyperthermia. MethodThe designed antenna resonated at 2.45 GHz. The antenna was tested in six different 3D liver models: Model A: without a tumor and blood vessels; Model B: with a realistic tumor (2x3 cm) and without blood vessels; Model C: created by adding blood vessels to model B; Model D: created by adding a small tumor (1.5x1.5 cm) to model C and changed its location; Model E: same as model C with a different tumor size; Model F: model with a simple spherical tumor (1.5x1.5 cm).ResultsThe return loss of the antenna varied from -45 dB to -25 dB for the 6 models. The Specific Absorption Rate (SAR) was between 29 W/kg to 30W/kg in the tumors and below 24 W/Kg in the surrounding tissues. The tumors’ temperature elevated to 43- 45°C, while the temperature of the surrounding tissues was below 41°C.ConclusionsThe results showed the capability of the designed antenna to raise the temperature of hepatic tumors to the therapeutic ranges of hyperthermia.
- 13 January 2022
- Published by Springer Nature in Research Square
Open AccessModeling of Interstitial Microwave Hyperthermia for Hepatic Tumors Using Floating Sleeve Antenna
PurposeMicrowave hyperthermia is a treatment modality that uses microwaves to destroy cancer cells by increasing their temperature to 41- 45°C. This study aims to design, modeling, and simulation of a microwave sleeve antenna for hepatic (liver) hyperthermia. MethodThe designed antenna resonated at 2.45 GHz. The antenna was tested in six different 3D liver models: Model A: without a tumor and blood vessels; Model B: with a realistic tumor (2x3 cm) and without blood vessels; Model C: created by adding blood vessels to model B; Model D: created by adding a small tumor (1.5x1.5 cm) to model C and changed its location; Model E: same as model C with a different tumor size; Model F: model with a simple spherical tumor (1.5x1.5 cm).ResultsThe return loss of the antenna varied from -45 dB to -25 dB for the 6 models. The Specific Absorption Rate (SAR) was between 29 W/kg to 30W/kg in the tumors and below 24 W/Kg in the surrounding tissues. The tumors’ temperature elevated to 43- 45°C, while the temperature of the surrounding tissues was below 41°C.ConclusionsThe results showed the capability of the designed antenna to raise the temperature of hepatic tumors to the therapeutic ranges of hyperthermia.
- Open Access
- 1 January 2012
- Published by Springer Nature in BioMedical Engineering OnLine
- Vol. 11 (1) , 47
FDTD analysis of a noninvasive hyperthermia system for brain tumors
Hyperthermia is considered one of the new therapeutic modalities for cancer treatment and is based on the difference in thermal sensitivity between healthy tissues and tumors. During hyperthermia treatment, the temperature of the tumor is raised to 40–45°C for a definite period resulting in the destruction of cancer cells. This paper investigates design, modeling and simulation of a new non-invasive hyperthermia applicator system capable of effectively heating deep seated as well as superficial brain tumors using inexpensive, simple, and easy to fabricate components without harming surrounding healthy brain tissues. The proposed hyperthermia applicator system is composed of an air filled partial half ellipsoidal chamber, a patch antenna, and a head model with an embedded tumor at an arbitrary location. The irradiating antenna is placed at one of the foci of the hyperthermia chamber while the center of the brain tumor is placed at the other focus. The finite difference time domain (FDTD) method is used to compute both the SAR patterns and the temperature distribution in three different head models due to two different patch antennas at a frequency of 915 MHz. The obtained results suggest that by using the proposed noninvasive hyperthermia system it is feasible to achieve sufficient and focused energy deposition and temperature rise to therapeutic values in deep seated as well as superficial brain tumors without harming surrounding healthy tissue. The proposed noninvasive hyperthermia system proved suitable for raising the temperature in tumors embedded in the brain to therapeutic values by carefully selecting the systems components. The operator of the system only needs to place the center of the brain tumor at a pre-specified location and excite the antenna at a single frequency of 915 MHz. Our study may provide a basis for a clinical applicator prototype capable of heating brain tumors.
- 1 January 2012
- Published by Springer Nature in BioMedical Engineering OnLine
- Vol. 11 (1) , 47
Open AccessFDTD analysis of a noninvasive hyperthermia system for brain tumors
Hyperthermia is considered one of the new therapeutic modalities for cancer treatment and is based on the difference in thermal sensitivity between healthy tissues and tumors. During hyperthermia treatment, the temperature of the tumor is raised to 40–45°C for a definite period resulting in the destruction of cancer cells. This paper investigates design, modeling and simulation of a new non-invasive hyperthermia applicator system capable of effectively heating deep seated as well as superficial brain tumors using inexpensive, simple, and easy to fabricate components without harming surrounding healthy brain tissues. The proposed hyperthermia applicator system is composed of an air filled partial half ellipsoidal chamber, a patch antenna, and a head model with an embedded tumor at an arbitrary location. The irradiating antenna is placed at one of the foci of the hyperthermia chamber while the center of the brain tumor is placed at the other focus. The finite difference time domain (FDTD) method is used to compute both the SAR patterns and the temperature distribution in three different head models due to two different patch antennas at a frequency of 915 MHz. The obtained results suggest that by using the proposed noninvasive hyperthermia system it is feasible to achieve sufficient and focused energy deposition and temperature rise to therapeutic values in deep seated as well as superficial brain tumors without harming surrounding healthy tissue. The proposed noninvasive hyperthermia system proved suitable for raising the temperature in tumors embedded in the brain to therapeutic values by carefully selecting the systems components. The operator of the system only needs to place the center of the brain tumor at a pre-specified location and excite the antenna at a single frequency of 915 MHz. Our study may provide a basis for a clinical applicator prototype capable of heating brain tumors.