Stefan Weick
Current Organizations / Institutes
The organizations/institutes from the scholar's latest article
Previous Organizations / Institutes
The organizations / institutes from the scholar's other articles in the past ten years
Research Subjects
Stefan Weick has a total of 82 co-authors. Here are the top co-authors by number of shared publications:
- Bülent Polat
Department of Radiation Oncology, University Hospital Wuerzburg, Wuerzburg, Germany
12 shared publications - Anne Richter
Department of Radiation Oncology, University Hospital Wuerzburg, Josef-Schneider-Straße 11, Würzburg, 97080, Germany
11 shared publications - Michael Flentje
Department of Radiation Oncology, University of Wuerzburg, 97080 Wuerzburg, Germany
8 shared publications - Florian Exner
Department of Radiation Oncology, University Hospital Wurzburg, Josef-Schneider-Str. 11, Wurzburg, Bavaria, 97080, Germany
6 shared publications Florian Exner
Department of Radiation Oncology, University Hospital Würzburg, Würzburg, Germany
5 shared publications- Gary Razinskas
Department of Radiation Oncology University Hospital Würzburg Würzburg Germany
5 shared publications
Latest Publications
- Open Access
- 19 November 2026
- Published by Elsevier in Radiography
- Vol. 32 (8)
Comparison of ocular radiation doses between ultra-low-dose photon-counting computed tomography and digital radiography for orbital foreign body screening: A phantom study
- 19 November 2026
- Published by Elsevier in Radiography
- Vol. 32 (8)
Open AccessComparison of ocular radiation doses between ultra-low-dose photon-counting computed tomography and digital radiography for orbital foreign body screening: A phantom study
- Open Access
- 20 August 2026
- Published by Springer Nature in Pediatric Radiology
- Vol. 56 (12) , 2764-2774
Radiographic imaging in pediatric suspected foreign body ingestion: potential for dose reduction through collimation and exposure strategies
Background: Foreign body ingestion constitutes a common emergency in children. Radiography is essential for confirming presence, type, and location of ingested objects. International guidelines recommend broad anatomical coverage from the neck to the pelvis; however, the impact of radiographic exposure parameters and collimation on radiation dose remains underexplored. Objective: To evaluate the effects of different scan protocols on radiation dose in pediatric foreign body ingestion. Materials and methods: Two-hundred-twenty-one radiographs from 183 pediatric patients (median age 2.8 years [interquartile range 1.5–5.5 years]; 112 males, 71 females) performed between September 2019 and April 2025 were retrospectively analyzed. Radiographs were categorized as neck–upper abdomen, extended collimation including pelvis, or abdomen only. Notably, the first two protocols used chest radiograph exposure parameters. Data included patient demographics, foreign body type and location, radiographic protocol, dose-area product, and estimated effective dose. Results: A foreign body was detected in 55.7% of radiographs, most of them in the stomach (59.4%). Coins were most common (31.7%), followed by button batteries (13.8%) and magnets (12.2%). Dose-area products were comparable between standard neck-to-upper-abdomen and extended neck-to-pelvis collimation but were markedly higher for abdomen-only imaging ( P P P <0.001). No differences were observed in patient age or repeat imaging rates. Conclusion: In suspected pediatric foreign body ingestion, collimation and exposure parameters substantially influence radiation dose. Low-dose chest parameters combined with restrictive collimation minimize radiation exposure, whereas abdominal radiographs carry higher dose, attributable predominantly to the abdominal exposure parameters employed.
- 20 August 2026
- Published by Springer Nature in Pediatric Radiology
- Vol. 56 (12) , 2764-2774
Open AccessRadiographic imaging in pediatric suspected foreign body ingestion: potential for dose reduction through collimation and exposure strategies
Background: Foreign body ingestion constitutes a common emergency in children. Radiography is essential for confirming presence, type, and location of ingested objects. International guidelines recommend broad anatomical coverage from the neck to the pelvis; however, the impact of radiographic exposure parameters and collimation on radiation dose remains underexplored. Objective: To evaluate the effects of different scan protocols on radiation dose in pediatric foreign body ingestion. Materials and methods: Two-hundred-twenty-one radiographs from 183 pediatric patients (median age 2.8 years [interquartile range 1.5–5.5 years]; 112 males, 71 females) performed between September 2019 and April 2025 were retrospectively analyzed. Radiographs were categorized as neck–upper abdomen, extended collimation including pelvis, or abdomen only. Notably, the first two protocols used chest radiograph exposure parameters. Data included patient demographics, foreign body type and location, radiographic protocol, dose-area product, and estimated effective dose. Results: A foreign body was detected in 55.7% of radiographs, most of them in the stomach (59.4%). Coins were most common (31.7%), followed by button batteries (13.8%) and magnets (12.2%). Dose-area products were comparable between standard neck-to-upper-abdomen and extended neck-to-pelvis collimation but were markedly higher for abdomen-only imaging ( P P P <0.001). No differences were observed in patient age or repeat imaging rates. Conclusion: In suspected pediatric foreign body ingestion, collimation and exposure parameters substantially influence radiation dose. Low-dose chest parameters combined with restrictive collimation minimize radiation exposure, whereas abdominal radiographs carry higher dose, attributable predominantly to the abdominal exposure parameters employed.
- Open Access
- 26 March 2026
- Published by Elsevier in Radiography
- Vol. 32 (4) , 103400
Radiation dose comparison of conventional radiography and photon-counting CT for ventriculoperitoneal shunt imaging: A phantom study
- 26 March 2026
- Published by Elsevier in Radiography
- Vol. 32 (4) , 103400
Open AccessRadiation dose comparison of conventional radiography and photon-counting CT for ventriculoperitoneal shunt imaging: A phantom study
- Open Access
- 28 February 2025
- Published by MDPI AG in Diagnostics
- Vol. 15 (5) , 596
Patient Shielding in Ultra-High-Resolution Cone-Beam CT of the Upper Extremity with a Twin Robotic X-Ray System
Background/Objectives: Gantry-free cone-beam CT (CBCT) allows for ultra-high-resolution (UHR) upper extremity imaging in a comfortable tableside position. The aim of this study was to assess the organ-specific radiation burden and the effect of dedicated lead shielding in the UHR-CBCT of the wrist and elbow. Methods: A modified Alderson-Rando phantom was scanned with the tableside UHR-CBCT mode of a twin robotic X-ray system employing identical scan parameters for wrist and elbow imaging. An ion chamber was used in conjunction with an electrometer to obtain representative organ dose measurements for the eye lens, thyroid gland, breast tissue, and abdomen. All measurements were performed with and without lead shielding. Results: Irrespective of the examined upper extremity joint, the highest absorbed dose among the assessed organs was determined for the eye lens (wrist imaging: 0.10 ± 0.01 mGy, elbow imaging: 0.12 ± 0.01 mGy). The most effective organ dose reduction by means of shielding in wrist CBCT was achieved for the thyroid gland (−17%). In elbow CBCT, the abdomen (−48%) and the ipsilateral breast (−39%) benefited particularly from shield protection. Conclusions: Although shielding was more effective in elbow than wrist scans, the overall impact in terms of absolute dose reduction was marginal.
- 28 February 2025
- Published by MDPI AG in Diagnostics
- Vol. 15 (5) , 596
Open AccessPatient Shielding in Ultra-High-Resolution Cone-Beam CT of the Upper Extremity with a Twin Robotic X-Ray System
Background/Objectives: Gantry-free cone-beam CT (CBCT) allows for ultra-high-resolution (UHR) upper extremity imaging in a comfortable tableside position. The aim of this study was to assess the organ-specific radiation burden and the effect of dedicated lead shielding in the UHR-CBCT of the wrist and elbow. Methods: A modified Alderson-Rando phantom was scanned with the tableside UHR-CBCT mode of a twin robotic X-ray system employing identical scan parameters for wrist and elbow imaging. An ion chamber was used in conjunction with an electrometer to obtain representative organ dose measurements for the eye lens, thyroid gland, breast tissue, and abdomen. All measurements were performed with and without lead shielding. Results: Irrespective of the examined upper extremity joint, the highest absorbed dose among the assessed organs was determined for the eye lens (wrist imaging: 0.10 ± 0.01 mGy, elbow imaging: 0.12 ± 0.01 mGy). The most effective organ dose reduction by means of shielding in wrist CBCT was achieved for the thyroid gland (−17%). In elbow CBCT, the abdomen (−48%) and the ipsilateral breast (−39%) benefited particularly from shield protection. Conclusions: Although shielding was more effective in elbow than wrist scans, the overall impact in terms of absolute dose reduction was marginal.
- Open Access
Impact of Radiotherapy on Malfunctions and Battery Life of Cardiac Implantable Electronic Devices in Cancer Patients
Purpose: This study analyses a large number of cancer patients with CIEDs for device malfunction and premature battery depletion by device interrogation after each radiotherapy fraction and compares different guidelines in regard to patient safety. Methods: From 2007 to 2022, a cohort of 255 patients was analyzed for CIED malfunctions via immediate device interrogation after every RT fraction. Results: Out of 324 series of radiotherapy treatments, with a total number of 5742 CIED interrogations, nine device malfunctions (2.8%) occurred. Switching into back-up/safety mode and software errors occurred four times each. Once, automatic read-out could not be performed. The median prescribed cumulative dose at planning target volume (PTV) associated with CIED malfunction was 45.0 Gy (IQR 36.0–64.0 Gy), with a median dose per fraction of 2.31 Gy (IQR 2.0–3.0 Gy). The median maximum dose at the CIED at time of malfunction was 0.3 Gy (IQR 0.0–1.3 Gy). No correlation between CIED malfunction and maximum photon energy (p = 0.07), maximum dose at the CIED (p = 0.59) nor treatment localization (p = 0.41) could be detected. After excluding the nine malfunctions, premature battery depletion was only observed three times (1.2%). Depending on the national guidelines, 1–9 CIED malfunctions in this study would have been detected on the day of occurrence and in none of the cases would patient safety have been compromised. Conclusion: Radiation-induced malfunctions of CIEDs and premature battery depletion are rare. If recommendations of national safety guidelines are followed, only a portion of the malfunctions would be detected directly after occurrence. Nevertheless, patient safety would not be compromised.
Open AccessImpact of Radiotherapy on Malfunctions and Battery Life of Cardiac Implantable Electronic Devices in Cancer Patients
Purpose: This study analyses a large number of cancer patients with CIEDs for device malfunction and premature battery depletion by device interrogation after each radiotherapy fraction and compares different guidelines in regard to patient safety. Methods: From 2007 to 2022, a cohort of 255 patients was analyzed for CIED malfunctions via immediate device interrogation after every RT fraction. Results: Out of 324 series of radiotherapy treatments, with a total number of 5742 CIED interrogations, nine device malfunctions (2.8%) occurred. Switching into back-up/safety mode and software errors occurred four times each. Once, automatic read-out could not be performed. The median prescribed cumulative dose at planning target volume (PTV) associated with CIED malfunction was 45.0 Gy (IQR 36.0–64.0 Gy), with a median dose per fraction of 2.31 Gy (IQR 2.0–3.0 Gy). The median maximum dose at the CIED at time of malfunction was 0.3 Gy (IQR 0.0–1.3 Gy). No correlation between CIED malfunction and maximum photon energy (p = 0.07), maximum dose at the CIED (p = 0.59) nor treatment localization (p = 0.41) could be detected. After excluding the nine malfunctions, premature battery depletion was only observed three times (1.2%). Depending on the national guidelines, 1–9 CIED malfunctions in this study would have been detected on the day of occurrence and in none of the cases would patient safety have been compromised. Conclusion: Radiation-induced malfunctions of CIEDs and premature battery depletion are rare. If recommendations of national safety guidelines are followed, only a portion of the malfunctions would be detected directly after occurrence. Nevertheless, patient safety would not be compromised.
- Open Access
- 26 April 2023
- Published by MDPI AG in Diagnostics
- Vol. 13 (9) , 1558
Obesity-Related Pitfalls of Virtual versus True Non-Contrast Imaging—An Intraindividual Comparison in 253 Oncologic Patients
Objectives: Dual-source dual-energy CT (DECT) facilitates reconstruction of virtual non-contrast images from contrast-enhanced scans within a limited field of view. This study evaluates the replacement of true non-contrast acquisition with virtual non-contrast reconstructions and investigates the limitations of dual-source DECT in obese patients. Materials and Methods: A total of 253 oncologic patients (153 women; age 64.5 ± 16.2 years; BMI 26.6 ± 5.1 kg/m2) received both multi-phase single-energy CT (SECT) and DECT in sequential staging examinations with a third-generation dual-source scanner. Patients were allocated to one of three BMI clusters: non-obese: 30 kg/m2 (n = 70). Radiation dose and image quality were compared for each scan. DECT examinations were evaluated regarding liver coverage within the dual-energy field of view. Results: While arterial contrast phases in DECT were associated with a higher CTDIvol than in SECT (11.1 vs. 8.1 mGy; p < 0.001), replacement of true with virtual non-contrast imaging resulted in a considerably lower overall dose-length product (312.6 vs. 475.3 mGy·cm; p < 0.001). The proportion of DLP variance predictable from patient BMI was substantial in DECT (R2 = 0.738) and SECT (R2 = 0.620); however, DLP of SECT showed a stronger increase in obese patients (p < 0.001). Incomplete coverage of the liver within the dual-energy field of view was most common in the obese subgroup (17.1%) compared with non-obese (0%) and pre-obese patients (4.1%). Conclusion: DECT facilitates a 30.8% dose reduction over SECT in abdominal oncologic staging examinations. Employing dual-source scanner architecture, the risk for incomplete liver coverage increases in obese patients.
- 26 April 2023
- Published by MDPI AG in Diagnostics
- Vol. 13 (9) , 1558
Open AccessObesity-Related Pitfalls of Virtual versus True Non-Contrast Imaging—An Intraindividual Comparison in 253 Oncologic Patients
Objectives: Dual-source dual-energy CT (DECT) facilitates reconstruction of virtual non-contrast images from contrast-enhanced scans within a limited field of view. This study evaluates the replacement of true non-contrast acquisition with virtual non-contrast reconstructions and investigates the limitations of dual-source DECT in obese patients. Materials and Methods: A total of 253 oncologic patients (153 women; age 64.5 ± 16.2 years; BMI 26.6 ± 5.1 kg/m2) received both multi-phase single-energy CT (SECT) and DECT in sequential staging examinations with a third-generation dual-source scanner. Patients were allocated to one of three BMI clusters: non-obese: 30 kg/m2 (n = 70). Radiation dose and image quality were compared for each scan. DECT examinations were evaluated regarding liver coverage within the dual-energy field of view. Results: While arterial contrast phases in DECT were associated with a higher CTDIvol than in SECT (11.1 vs. 8.1 mGy; p < 0.001), replacement of true with virtual non-contrast imaging resulted in a considerably lower overall dose-length product (312.6 vs. 475.3 mGy·cm; p < 0.001). The proportion of DLP variance predictable from patient BMI was substantial in DECT (R2 = 0.738) and SECT (R2 = 0.620); however, DLP of SECT showed a stronger increase in obese patients (p < 0.001). Incomplete coverage of the liver within the dual-energy field of view was most common in the obese subgroup (17.1%) compared with non-obese (0%) and pre-obese patients (4.1%). Conclusion: DECT facilitates a 30.8% dose reduction over SECT in abdominal oncologic staging examinations. Employing dual-source scanner architecture, the risk for incomplete liver coverage increases in obese patients.
- Open Access
- 9 December 2022
- Published by Elsevier in Zeitschrift für Medizinische Physik
- Vol. 34 (3) , 384-396
Prospective risk analysis of the online-adaptive artificial intelligence-driven workflow using the Ethos treatment system
- 9 December 2022
- Published by Elsevier in Zeitschrift für Medizinische Physik
- Vol. 34 (3) , 384-396
Open AccessProspective risk analysis of the online-adaptive artificial intelligence-driven workflow using the Ethos treatment system
- Open Access
- 8 December 2022
- Published by Springer Nature in Scientific Reports
- Vol. 12 (1) , 1-9
Acute toxicities of patients with locally advanced rectal cancer treated with intensified chemoradiotherapy within the CAO/ARO/AIO-12 trial: comparing conventional versus VMAT planning at a single center
In locally advanced rectal cancer (LARC) neoadjuvant chemoradiotherapy is regarded as standard treatment. We assessed acute toxicities in patients receiving conventional 3D-conformal radiotherapy (3D-RT) and correlated them with dosimetric parameters after re-planning with volumetric modulated arc therapy (VMAT). Patients were randomized within the multicenter CAO/ARO/AIO-12 trial and received 50.4 Gy in 28 fractions and simultaneous chemotherapy with fluorouracil and oxaliplatin. Organs at risk (OAR) were contoured in a standardized approach. Acute toxicities and dose volume histogram parameters of 3D-RT plans were compared to retrospectively calculated VMAT plans. From 08/2015 to 01/2018, 35 patients with LARC were treated at one study center. Thirty-four patients were analyzed of whom 1 (3%) was UICC stage II and 33 (97%) patients were UICC stage III. Grade 3 acute toxicities occurred in 5 patients (15%). Patients with acute grade 1 cystitis (n = 9) had significantly higher Dmean values for bladder (29.4 Gy vs. 25.2 Gy, p < 0.01) compared to patients without bladder toxicities. Acute diarrhea was associated with small bowel volume (grade 2: 870.1 ccm vs. grade 0–1: 647.3 ccm; p < 0.01) and with the irradiated volumes V5 to V50. Using VMAT planning, we could reduce mean doses and irradiated volumes for all OAR: Dmean bladder (21.9 Gy vs. 26.3 Gy, p < 0.01), small bowel volumes V5–V45 (p < 0.01), Dmean anal sphincter (34.6 Gy vs. 35.6 Gy, p < 0.01) and Dmean femoral heads (right 11.4 Gy vs. 25.9 Gy, left 12.5 Gy vs. 26.6 Gy, p < 0.01). Acute small bowel and bladder toxicities were dose and volume dependent. Dose and volume sparing for all OAR could be achieved through VMAT planning and might result in less acute toxicities.
- 8 December 2022
- Published by Springer Nature in Scientific Reports
- Vol. 12 (1) , 1-9
Open AccessAcute toxicities of patients with locally advanced rectal cancer treated with intensified chemoradiotherapy within the CAO/ARO/AIO-12 trial: comparing conventional versus VMAT planning at a single center
In locally advanced rectal cancer (LARC) neoadjuvant chemoradiotherapy is regarded as standard treatment. We assessed acute toxicities in patients receiving conventional 3D-conformal radiotherapy (3D-RT) and correlated them with dosimetric parameters after re-planning with volumetric modulated arc therapy (VMAT). Patients were randomized within the multicenter CAO/ARO/AIO-12 trial and received 50.4 Gy in 28 fractions and simultaneous chemotherapy with fluorouracil and oxaliplatin. Organs at risk (OAR) were contoured in a standardized approach. Acute toxicities and dose volume histogram parameters of 3D-RT plans were compared to retrospectively calculated VMAT plans. From 08/2015 to 01/2018, 35 patients with LARC were treated at one study center. Thirty-four patients were analyzed of whom 1 (3%) was UICC stage II and 33 (97%) patients were UICC stage III. Grade 3 acute toxicities occurred in 5 patients (15%). Patients with acute grade 1 cystitis (n = 9) had significantly higher Dmean values for bladder (29.4 Gy vs. 25.2 Gy, p < 0.01) compared to patients without bladder toxicities. Acute diarrhea was associated with small bowel volume (grade 2: 870.1 ccm vs. grade 0–1: 647.3 ccm; p < 0.01) and with the irradiated volumes V5 to V50. Using VMAT planning, we could reduce mean doses and irradiated volumes for all OAR: Dmean bladder (21.9 Gy vs. 26.3 Gy, p < 0.01), small bowel volumes V5–V45 (p < 0.01), Dmean anal sphincter (34.6 Gy vs. 35.6 Gy, p < 0.01) and Dmean femoral heads (right 11.4 Gy vs. 25.9 Gy, left 12.5 Gy vs. 26.6 Gy, p < 0.01). Acute small bowel and bladder toxicities were dose and volume dependent. Dose and volume sparing for all OAR could be achieved through VMAT planning and might result in less acute toxicities.
- Open Access
- 21 November 2022
- Published by Springer Nature in Radiation Oncology
- Vol. 17 (1) , 1-10
Treatment plan comparison for irradiation of multiple brain metastases with hippocampal avoidance whole brain radiotherapy and simultaneous integrated boost using the Varian Halcyon and the Elekta Synergy platforms
- 21 November 2022
- Published by Springer Nature in Radiation Oncology
- Vol. 17 (1) , 1-10
Open AccessTreatment plan comparison for irradiation of multiple brain metastases with hippocampal avoidance whole brain radiotherapy and simultaneous integrated boost using the Varian Halcyon and the Elekta Synergy platforms
- Open Access
- 12 May 2022
- Published by Springer Nature in Strahlentherapie und Onkologie
- Vol. 198 (8) , 735-743
Propensity score-matched analysis comparing dose-escalated intensity-modulated radiation therapy versus external beam radiation therapy plus high-dose-rate brachytherapy for localized prostate cancer
Purpose: Dose-escalated external beam radiation therapy (EBRT) and EBRT + high-dose-rate brachytherapy (HDR-BT) boost are guideline-recommended treatment options for localized prostate cancer. The purpose of this study was to compare long-term outcome and toxicity of dose-escalated EBRT versus EBRT + HDR-BT boost. Methods: From 2002 to 2019, 744 consecutive patients received either EBRT or EBRT + HDR-BT boost, of whom 516 patients were propensity score matched. Median follow-up was 95.3 months. Cone beam CT image-guided EBRT consisted of 33 fractions of intensity-modulated radiation therapy with simultaneous integrated boost up to 76.23 Gy (DMean). Combined treatment was delivered as 46 Gy (DMean) EBRT, followed by two fractions HDR-BT boost with 9 Gy (D90%). Propensity score matching was applied before analysis of the primary endpoint, estimated 10-year biochemical relapse-free survival (bRFS), and the secondary endpoints metastasis-free survival (MFS) and overall survival (OS). Prognostic parameters were analyzed by Cox proportional hazard modelling. Genitourinary (GU)/gastrointestinal (GI) toxicity evaluation used the Common Toxicity Criteria for Adverse Events (v5.0). Results: The estimated 10-year bRFS was 82.0% vs. 76.4% (p = 0.075) for EBRT alone versus combined treatment, respectively. The estimated 10-year MFS was 82.9% vs. 87.0% (p = 0.195) and the 10-year OS was 65.7% vs. 68.9% (p = 0.303), respectively. Cumulative 5‑year late GU ≥ grade 2 toxicities were seen in 23.6% vs. 19.2% (p = 0.086) and 5‑year late GI ≥ grade 2 toxicities in 11.1% vs. 5.0% of the patients (p = 0.002); cumulative 5‑year late grade 3 GU toxicity occurred in 4.2% vs. 3.6% (p = 0.401) and GI toxicity in 1.0% vs. 0.3% (p = 0.249), respectively. Conclusion: Both treatment groups showed excellent long-term outcomes with low rates of severe toxicity.
- 12 May 2022
- Published by Springer Nature in Strahlentherapie und Onkologie
- Vol. 198 (8) , 735-743
Open AccessPropensity score-matched analysis comparing dose-escalated intensity-modulated radiation therapy versus external beam radiation therapy plus high-dose-rate brachytherapy for localized prostate cancer
Purpose: Dose-escalated external beam radiation therapy (EBRT) and EBRT + high-dose-rate brachytherapy (HDR-BT) boost are guideline-recommended treatment options for localized prostate cancer. The purpose of this study was to compare long-term outcome and toxicity of dose-escalated EBRT versus EBRT + HDR-BT boost. Methods: From 2002 to 2019, 744 consecutive patients received either EBRT or EBRT + HDR-BT boost, of whom 516 patients were propensity score matched. Median follow-up was 95.3 months. Cone beam CT image-guided EBRT consisted of 33 fractions of intensity-modulated radiation therapy with simultaneous integrated boost up to 76.23 Gy (DMean). Combined treatment was delivered as 46 Gy (DMean) EBRT, followed by two fractions HDR-BT boost with 9 Gy (D90%). Propensity score matching was applied before analysis of the primary endpoint, estimated 10-year biochemical relapse-free survival (bRFS), and the secondary endpoints metastasis-free survival (MFS) and overall survival (OS). Prognostic parameters were analyzed by Cox proportional hazard modelling. Genitourinary (GU)/gastrointestinal (GI) toxicity evaluation used the Common Toxicity Criteria for Adverse Events (v5.0). Results: The estimated 10-year bRFS was 82.0% vs. 76.4% (p = 0.075) for EBRT alone versus combined treatment, respectively. The estimated 10-year MFS was 82.9% vs. 87.0% (p = 0.195) and the 10-year OS was 65.7% vs. 68.9% (p = 0.303), respectively. Cumulative 5‑year late GU ≥ grade 2 toxicities were seen in 23.6% vs. 19.2% (p = 0.086) and 5‑year late GI ≥ grade 2 toxicities in 11.1% vs. 5.0% of the patients (p = 0.002); cumulative 5‑year late grade 3 GU toxicity occurred in 4.2% vs. 3.6% (p = 0.401) and GI toxicity in 1.0% vs. 0.3% (p = 0.249), respectively. Conclusion: Both treatment groups showed excellent long-term outcomes with low rates of severe toxicity.