Latest Publications
- Open Access
- 13 January 2023
- Published by MDPI AG in Micromachines
- Vol. 14 (1) , 202
Ultrasonic Enhancement of Chondrogenesis in Mesenchymal Stem Cells by Bolt-Clamped Langevin Transducers
We recently investigated the design and fabrication of Langevin-type transducers for therapeutic ultrasound. Effect of ultrasonic energy arising from the transducer on biological tissue was examined. In this study, the transducer was set to radiate acoustic energy to mesenchymal stem cells (MSCs) for inducing differentiation into cartilage tissue. The average chondrogenic ratio in area was 20.82% in the control group, for which no external stimulation was given. Shear stress was applied to MSCs as the contrast group, which resulted in 42.66% on average with a 25.92% minimum rate; acoustic pressure from the flat tip causing transient cavitation enhanced chondrogenesis up to 52.96%. For the round tip excited by 20 Vpp, the maximum differentiation value of 69.43% was found, since it delivered relatively high acoustic pressure to MSCs. Hence, the results from this study indicate that ultrasound pressure at the kPa level can enhance MSC chondrogenesis compared to the tens of kHz range by Langevin transducers.
- 13 January 2023
- Published by MDPI AG in Micromachines
- Vol. 14 (1) , 202
Open AccessUltrasonic Enhancement of Chondrogenesis in Mesenchymal Stem Cells by Bolt-Clamped Langevin Transducers
We recently investigated the design and fabrication of Langevin-type transducers for therapeutic ultrasound. Effect of ultrasonic energy arising from the transducer on biological tissue was examined. In this study, the transducer was set to radiate acoustic energy to mesenchymal stem cells (MSCs) for inducing differentiation into cartilage tissue. The average chondrogenic ratio in area was 20.82% in the control group, for which no external stimulation was given. Shear stress was applied to MSCs as the contrast group, which resulted in 42.66% on average with a 25.92% minimum rate; acoustic pressure from the flat tip causing transient cavitation enhanced chondrogenesis up to 52.96%. For the round tip excited by 20 Vpp, the maximum differentiation value of 69.43% was found, since it delivered relatively high acoustic pressure to MSCs. Hence, the results from this study indicate that ultrasound pressure at the kPa level can enhance MSC chondrogenesis compared to the tens of kHz range by Langevin transducers.
- Open Access
Acoustic and Thermal Characterization of Therapeutic Ultrasonic Langevin Transducers under Continuous- and Pulsed Wave Excitations
We previously conducted an empirical study on Langevin type transducers in medical use by examining the heat effect on porcine tissue. For maximum acoustic output, the transducer was activated by a continuous sinusoidal wave. In this work, pulsed waves with various duty factors were applied to our transducer model in order to examine their effect on functionality. Acoustic power, electro-acoustic conversion efficiency, acoustic pressure, thermal effect on porcine tissue and bovine muscle, and heat generation in the transducer were investigated under various input conditions. For example, the results of applying a continuous wave of 200 VPP and a pulse wave of 70% duty factor with the same amplitude to the transducer were compared. It was found that continuous waves generated 9.79 W of acoustic power, 6.40% energy efficiency, and 24.84 kPa acoustic pressure. In pulsed excitation, the corresponding values were 9.04 W, 8.44%, and 24.7 kPa, respectively. The maximum temperature increases in bovine muscle are reported to be 83.0 °C and 89.5 °C for each waveform, whereas these values were 102.5 °C and 84.5 °C in fatty porcine tissue. Moreover, the heat generation around the transducer was monitored under continuous and pulsed modes and was found to be 51.3 °C and 50.4 °C. This shows that pulsed excitation gives rise to less thermal influence on the transducer. As a result, it is demonstrated that a transducer triggered by pulsed waves improves the energy efficiency and provides sufficient thermal impact on biological tissues by selecting proper electrical excitation types.
Open AccessAcoustic and Thermal Characterization of Therapeutic Ultrasonic Langevin Transducers under Continuous- and Pulsed Wave Excitations
We previously conducted an empirical study on Langevin type transducers in medical use by examining the heat effect on porcine tissue. For maximum acoustic output, the transducer was activated by a continuous sinusoidal wave. In this work, pulsed waves with various duty factors were applied to our transducer model in order to examine their effect on functionality. Acoustic power, electro-acoustic conversion efficiency, acoustic pressure, thermal effect on porcine tissue and bovine muscle, and heat generation in the transducer were investigated under various input conditions. For example, the results of applying a continuous wave of 200 VPP and a pulse wave of 70% duty factor with the same amplitude to the transducer were compared. It was found that continuous waves generated 9.79 W of acoustic power, 6.40% energy efficiency, and 24.84 kPa acoustic pressure. In pulsed excitation, the corresponding values were 9.04 W, 8.44%, and 24.7 kPa, respectively. The maximum temperature increases in bovine muscle are reported to be 83.0 °C and 89.5 °C for each waveform, whereas these values were 102.5 °C and 84.5 °C in fatty porcine tissue. Moreover, the heat generation around the transducer was monitored under continuous and pulsed modes and was found to be 51.3 °C and 50.4 °C. This shows that pulsed excitation gives rise to less thermal influence on the transducer. As a result, it is demonstrated that a transducer triggered by pulsed waves improves the energy efficiency and provides sufficient thermal impact on biological tissues by selecting proper electrical excitation types.
- Open Access
Acoustic Power Measurement and Thermal Bioeffect Evaluation of Therapeutic Langevin Transducers
We recently proposed an analytical design method of Langevin transducers for therapeutic ultrasound treatment by conducting parametric study to estimate the effect of compression force on resonance characteristics. In this study, experimental investigations were further performed under various electrical conditions to observe the acoustic power of the fully equipped transducer and to assess its heat-related bioeffect. Thermal index (TI) tests were carried out to examine temperature rise and thermal damage induced by the acoustic energy in fatty porcine tissue. Acoustic power emission, TI values, temperature characteristics, and depth/size of thermal ablation were measured as a function of transducer’s driving voltage. By exciting the transducer with 300 Vpp sinusoidal continuous waveform, for instance, the average power was 23.1 W and its corresponding TI was 4.1, less than the 6 specified by the Food and Drug Administration (FDA) guideline. The maximum temperature and the depth of the affected site were 74.5 °C and 19 mm, respectively. It is shown that thermal ablation is likely to be more affected by steep heat surge for a short duration rather than by slow temperature rise over time. Hence, the results demonstrate the capability of our ultrasonic transducer intended for therapeutic procedures by safely interrogating soft tissue and yet delivering enough energy to thermally stimulate the tissue in depth.
Open AccessAcoustic Power Measurement and Thermal Bioeffect Evaluation of Therapeutic Langevin Transducers
We recently proposed an analytical design method of Langevin transducers for therapeutic ultrasound treatment by conducting parametric study to estimate the effect of compression force on resonance characteristics. In this study, experimental investigations were further performed under various electrical conditions to observe the acoustic power of the fully equipped transducer and to assess its heat-related bioeffect. Thermal index (TI) tests were carried out to examine temperature rise and thermal damage induced by the acoustic energy in fatty porcine tissue. Acoustic power emission, TI values, temperature characteristics, and depth/size of thermal ablation were measured as a function of transducer’s driving voltage. By exciting the transducer with 300 Vpp sinusoidal continuous waveform, for instance, the average power was 23.1 W and its corresponding TI was 4.1, less than the 6 specified by the Food and Drug Administration (FDA) guideline. The maximum temperature and the depth of the affected site were 74.5 °C and 19 mm, respectively. It is shown that thermal ablation is likely to be more affected by steep heat surge for a short duration rather than by slow temperature rise over time. Hence, the results demonstrate the capability of our ultrasonic transducer intended for therapeutic procedures by safely interrogating soft tissue and yet delivering enough energy to thermally stimulate the tissue in depth.
- Open Access
- 9 January 2022
- Published by MDPI AG in Electronics
- Vol. 11 (2) , 194
Theoretical and Empirical Verification of Electrical Impedance Matching Method for High-Power Transducers
In our prior study, a systematic approach was used to devise Langevin transducers for high-power applications where the energy efficiency was not considered in the design criteria. In this paper, the impedance matching methods are thus proposed to evaluate what matching topology is appropriate for their use. Both the series inductor scheme and low pass filter composed of a series inductor and shunt capacitor are examined as matching circuits. According to MATLAB simulation, the resonance frequency is seen at 36.79 kHz due to a series L circuit, and its associated impedance is reduced by 70.45% from that of its non-matching condition. The measured resonance frequency is 36.77 kHz and the corresponding impedance is decreased by 59.52%. Furthermore, the acoustic pressure is measured to determine the effect of the matching circuit on the transducer’s actual behavior. The transducer with a series L circuit shows more efficient matching results, 2.28 kPa of positive acoustic pressure is emitted without matching and 3.35 kPa is emitted with a series L element, respectively. As a result, this study demonstrates how to evaluate the influence of matching circuits by using our customized approach rather than commercial SPICE programs, as well as how to experimentally verify the acoustic behavior of high-power Langevin transducers.
- 9 January 2022
- Published by MDPI AG in Electronics
- Vol. 11 (2) , 194
Open AccessTheoretical and Empirical Verification of Electrical Impedance Matching Method for High-Power Transducers
In our prior study, a systematic approach was used to devise Langevin transducers for high-power applications where the energy efficiency was not considered in the design criteria. In this paper, the impedance matching methods are thus proposed to evaluate what matching topology is appropriate for their use. Both the series inductor scheme and low pass filter composed of a series inductor and shunt capacitor are examined as matching circuits. According to MATLAB simulation, the resonance frequency is seen at 36.79 kHz due to a series L circuit, and its associated impedance is reduced by 70.45% from that of its non-matching condition. The measured resonance frequency is 36.77 kHz and the corresponding impedance is decreased by 59.52%. Furthermore, the acoustic pressure is measured to determine the effect of the matching circuit on the transducer’s actual behavior. The transducer with a series L circuit shows more efficient matching results, 2.28 kPa of positive acoustic pressure is emitted without matching and 3.35 kPa is emitted with a series L element, respectively. As a result, this study demonstrates how to evaluate the influence of matching circuits by using our customized approach rather than commercial SPICE programs, as well as how to experimentally verify the acoustic behavior of high-power Langevin transducers.
- Open Access
Parametric Study of Bolt Clamping Effect on Resonance Characteristics of Langevin Transducers with Lumped Circuit Models
We recently proposed a numerical model using equivalent circuit models to analyze the resonance characteristics of Langevin transducers and design them in a systematic manner. However, no pre-load torque biased by a metal bolt was considered in the model. Here, a parametric study is, therefore, carried out to reveal how model parameters are adapted to incorporate the pre-compression effect into our existing model. Analytical results are compared with corresponding experimental data, particularly regarding the input electrical impedance and effective electromechanical coupling coefficient for the transducer at resonance modes. The frequency response of input impedance is presented as a function of torque, both theoretically and experimentally. For 10.0 N·m bias, for instance, both resonance and anti-resonance frequencies are calculated as 38.64 kHz and 39.78 kHz, while these are measured as 38.62 kHz and 39.77 kHz by the impedance analyzer. The impedance difference between these cases is 14 Ω at resonance and 9 kΩ at anti-resonance, while the coupling coefficients in both cases become 0.238 and 0.239, respectively. Hence, these test results are closely matched with their theoretical values. Consequently, this study provides a quantitative guideline that specifies the pre-loading condition of bolt clamps with proper parameter settings to predict the intended resonance characteristics of Langevin transducers.
Open AccessParametric Study of Bolt Clamping Effect on Resonance Characteristics of Langevin Transducers with Lumped Circuit Models
We recently proposed a numerical model using equivalent circuit models to analyze the resonance characteristics of Langevin transducers and design them in a systematic manner. However, no pre-load torque biased by a metal bolt was considered in the model. Here, a parametric study is, therefore, carried out to reveal how model parameters are adapted to incorporate the pre-compression effect into our existing model. Analytical results are compared with corresponding experimental data, particularly regarding the input electrical impedance and effective electromechanical coupling coefficient for the transducer at resonance modes. The frequency response of input impedance is presented as a function of torque, both theoretically and experimentally. For 10.0 N·m bias, for instance, both resonance and anti-resonance frequencies are calculated as 38.64 kHz and 39.78 kHz, while these are measured as 38.62 kHz and 39.77 kHz by the impedance analyzer. The impedance difference between these cases is 14 Ω at resonance and 9 kΩ at anti-resonance, while the coupling coefficients in both cases become 0.238 and 0.239, respectively. Hence, these test results are closely matched with their theoretical values. Consequently, this study provides a quantitative guideline that specifies the pre-loading condition of bolt clamps with proper parameter settings to predict the intended resonance characteristics of Langevin transducers.