SRSonia Ruiz

Sonia Ruiz

*This is an algorithmically generated author record based on publicly available information.

Current Organizations / Institutes

The organizations/institutes from the scholar's latest article

Previous Organizations / Institutes

The organizations / institutes from the scholar's other articles in the past ten years


Sonia Ruiz has a total of 9 co-authors. Here are the top co-authors by number of shared publications:


Latest Publications

  • Open Access

    ATOMS (Adjustable Trans-Obturator Male System) in Patients with Post-Prostatectomy Incontinence and Previously Treated Urethral Stricture or Bladder Neck Contracture

    by Ane Ullate, Ignacio Arance, Miguel Virseda-Chamorro, Sonia Ruiz, Juliusz Szczesniewski, Carlos Téllez, Fabian Queissert, Juan F. Dorado, Javier C. Angulo

    (1) Background: Male stress incontinence in patients with previously treated urethral or bladder neck stricture is a therapeutic challenge. The efficacy and safety of the adjustable trans-obturator male system (ATOMS) in these patients is unknown. (2) Methods: All patients with primary ATOMS implants in our institution between 2014 and 2021 were included. The outcomes of patients with previously treated urethral or bladder neck stricture (≥6 months before ATOMS implant) and stable 16Ch urethral caliber were compared to those without a history of stricture. The primary endpoint was the dry patient rate, defined as the pad test ≤ 20 mL/day, and complication rate, including device removal. The secondary variable was self-perceived satisfaction using the Patient Global Impression of Improvement (PGI-I) scale. Wilcoxon rank sum test, Fisher’s exact test and logistic regression were performed. (3) Results: One hundred and forty-nine consecutive patients were included, twenty-one (14%) previously treated for urethral or bladder neck stricture (seven urethroplasty, nine internal urethrotomy and five bladder neck incision). After ATOMS adjustment, 38% of the patients with treated stricture were continent compared to 83% of those without (p < 0.0001). After weighted matched observations using propensity score pairing, the proportion of continent patients without a previous stricture was 56% (p = 0.236). Complications occurred in 29% of the patients with stricture and in 20% of those without (p = 0.34). The severity of the complications was distributed evenly among the groups (p = 0.42). Regarding self-perceived satisfaction with the implant, 90% of the patients with stricture perceived the results satisfactorily (PGI-I 1–3) compared to 97% of the rest (p = 0.167). Stricture was associated with radiotherapy (p < 0.0001) and time from prostatectomy to implantation (p = 0.012). There was a moderate correlation between previous stricture and the severity of incontinence, both evaluated according to the 24-h pad test (Rho = 0.378; p < 0.0001) and the ICIQ-SF questionnaire (Rho = 0.351; p < 0.0001). Multivariate analysis for the factors predictive of failure after ATOMS adjustment revealed previous stricture (OR 4.66; 95% CI 1.2–18.87), baseline 24-h pad test (per 100 mL, OR 1.28; 95% CI 1.09–1.52) and final cushion volume (per mL, OR 1.34; 95% CI 1.19–1.55). This model predicted dryness with an AUC of 92%. After the PSMATCH procedure using a propensity score, the model remained unchanged, with the previous stricture (OR 8.05; 95% CI 1.08–110.83), baseline 24-h pad test (per 100 mL, OR 1.53; 95% CI 1.15–2.26) and final cushion volume (per mL, OR 1.45; 95% CI 1.17–2) being independent predictors and an AUC of 93%. (4) Conclusions: ATOMS can be used to treat male stress incontinence in patients with a history of stricture, although the effectiveness of the device is reduced. On the other hand, the security and perceived satisfaction were equivalent for both groups.


  • Open Access

    The Mode of Action of Adjustable Transobturator Male System (ATOMS): Intraoperative Urethral Pressure Measurements

    by Sonia Ruiz, Miguel Virseda-Chamorro, Fabian Queissert, Andrés López, Ignacio Arance, Javier C. Angulo

    (1) Background: The Adjustable Transobturator Male System (ATOMS) device is increasingly used to treat post-prostatectomy incontinence as it enhances residual urinary sphincteric function and allows continence recovery or improvement by dorsal compression of the bulbar urethra through a fixed transobturator mesh passage. The mode of action and the profile of the patients with best results are not totally understood. (2) Methods: Intraoperative urethral pressure measurements at different filling levels of the ATOMS device show increased urethral resistance and enhanced residual sphincteric activity. We evaluated whether the pattern of urethral pressure change secondary to serial progressive intraoperative filling of the cushion can predict postoperative results after ATOMS placement. (3) Results: The regression analysis showed a significant direct relationship between cushion volume and intraurethral pressure (p = 0.000). The median intraurethral pressure at atmospheric pressure was 51 ± 22.7 cm H2O, and at atmospheric pressure plus 4 mL was 80 ± 23.1 cm H2O). Cluster analyses defined a group of patients (n = 6) formed by patients with a distensible urethra and 100% continence after adjustment in contrast to another group (n = 3) with rigid urethras and 33% continence after adjustment. (4) Conclusions: As a part of its continence mechanism, the ATOMS device leads to continence by increasing intraurethral pressure owing to the stretching effect on the urethral wall caused by cushion filling that increases urethral resistance.