Yuri Simeonov
Previous Organizations / Institutes
The organizations / institutes from the scholar's other articles in the past ten years
Research Subjects
Yuri Simeonov has a total of 39 co-authors. Here are the top co-authors by number of shared publications:
- Klemens Zink
University Medical Center Giessen-Marburg, Department of Radiotherapy and Radiooncology, Baldingerstrasse, Marburg, 35043, GERMANY
4 shared publications Petar Penchev
University of Applied Sciences, Institute of Medical Physics and Radiation Protection (IMPS), Giessen, Germany
4 shared publicationsRita Engenhart-Cabillic
University Medical Center Giessen-Marburg, Department of Radiotherapy and Radiooncology, Marburg, GERMANY
4 shared publicationsUli Weber
GSI Helmholtzzentrum für Schwerionenforschung GmbH, Biophysics division, Darmstadt, Germany
4 shared publicationsChristoph Schuy
GSI Helmholtzzentrum für Schwerionenforschung GmbH, Biophysics division, Darmstadt, Germany
3 shared publications- Christoph Schuy
Biophysics Department, GSI Helmholtzzentrum für Schwerionenforschung, Planckstrasse 1, 64291, Darmstadt, Germany
3 shared publications
Latest Publications
- Open Access
- 5 November 2026
- Published by Elsevier in Zeitschrift für Medizinische Physik
A prototype system for fast clinical treatment with carbon ions using a 3D range modulator
- 5 November 2026
- Published by Elsevier in Zeitschrift für Medizinische Physik
Open AccessA prototype system for fast clinical treatment with carbon ions using a 3D range modulator
- Open Access
- 20 April 2025
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 70 (8) , 085017
Monte Carlo simulation of beam quality correction factor kQ for carbon-ion beams using FLUKA and GATE for selected cylindrical and plane-parallel ionization chambers
Objective. To benchmark Monte Carlo codes FLUKA and GATE/Geant4 regarding the beam quality correction factors of ionization chambers for monoenergetic carbon-ion beams against experimental results by Holm et al (2022 Phys. Med. Biol. 67 49401). Approach. Monte Carlo codes FLUKA and GATE/Geant4 were used to simulate the beam quality correction factor k Q for one plane-parallel (PTW 34001) and two cylindrical ionization chambers (PTW 30013 and IBA FC65-G) using two monoenergetic carbon-ion beams and an energy modulated beam in accordance with Holm et al (2022 Phys. Med. Biol. 67 49401). Additionally, chamber-specific factor f Q and perturbation factor p Q were calculated. Differences between Geant4 reference physics lists were investigated by comparing simulated depth dose distributions, f Q , and p Q values for 429 MeV u−1. Main results. Simulated k Q factors were found to differ from experimentally determined k Q factors of Holm et al (2022 Phys. Med. Biol. 67 49401) by 2.5% for cylindrical chambers, whereas the plane-parallel chamber showed larger deviations of 3.1/2.6% (GATE/FLUKA), exceeding the simulation uncertainty of 1.7%. f Q and p Q factors simulated using different Geant4 physics lists were comparable within the Type-A uncertainty of 0.2%. Nevertheless, the depth dose curves for physics lists using the INCL++ model showed an increase in dose at all depths except for the fragmentation tail. Differences in f Q factors between Monte Carlo codes FLUKA and GATE of up to 1.8% have been observed. Significance. More investigations are needed to understand the cause of the observed deviations between experimental results and Monte Carlo calculations of beam quality correction factor k Q . No statistically significant differences are observed between investigated Geant4 physics lists for f Q , k Q , and p Q simulations. Notably, differences between Monte Carlo codes FLUKA and GATE are one of the main sources that limit the current simulation uncertainty.
- 20 April 2025
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 70 (8) , 085017
Open AccessMonte Carlo simulation of beam quality correction factor kQ for carbon-ion beams using FLUKA and GATE for selected cylindrical and plane-parallel ionization chambers
Objective. To benchmark Monte Carlo codes FLUKA and GATE/Geant4 regarding the beam quality correction factors of ionization chambers for monoenergetic carbon-ion beams against experimental results by Holm et al (2022 Phys. Med. Biol. 67 49401). Approach. Monte Carlo codes FLUKA and GATE/Geant4 were used to simulate the beam quality correction factor k Q for one plane-parallel (PTW 34001) and two cylindrical ionization chambers (PTW 30013 and IBA FC65-G) using two monoenergetic carbon-ion beams and an energy modulated beam in accordance with Holm et al (2022 Phys. Med. Biol. 67 49401). Additionally, chamber-specific factor f Q and perturbation factor p Q were calculated. Differences between Geant4 reference physics lists were investigated by comparing simulated depth dose distributions, f Q , and p Q values for 429 MeV u−1. Main results. Simulated k Q factors were found to differ from experimentally determined k Q factors of Holm et al (2022 Phys. Med. Biol. 67 49401) by 2.5% for cylindrical chambers, whereas the plane-parallel chamber showed larger deviations of 3.1/2.6% (GATE/FLUKA), exceeding the simulation uncertainty of 1.7%. f Q and p Q factors simulated using different Geant4 physics lists were comparable within the Type-A uncertainty of 0.2%. Nevertheless, the depth dose curves for physics lists using the INCL++ model showed an increase in dose at all depths except for the fragmentation tail. Differences in f Q factors between Monte Carlo codes FLUKA and GATE of up to 1.8% have been observed. Significance. More investigations are needed to understand the cause of the observed deviations between experimental results and Monte Carlo calculations of beam quality correction factor k Q . No statistically significant differences are observed between investigated Geant4 physics lists for f Q , k Q , and p Q simulations. Notably, differences between Monte Carlo codes FLUKA and GATE are one of the main sources that limit the current simulation uncertainty.
- 7 January 2025
- Published by Elsevier in Physica Medica
- Vol. 130, 104883
Characterisation of the HollandPTC R&D proton beamline for physics and radiobiology studies
- 7 January 2025
- Published by Elsevier in Physica Medica
- Vol. 130, 104883
Characterisation of the HollandPTC R&D proton beamline for physics and radiobiology studies
- Open Access
- 11 March 2022
- Published by IOP Publishing in Biomedical Physics & Engineering Express
- Vol. 8 (3) , 035006
Development, Monte Carlo simulations and experimental evaluation of a 3D range-modulator for a complex target in scanned proton therapy
The purpose of this work was to develop and manufacture a 3D range-modulator (3D RM) for a complex target contour for scanned proton therapy. The 3D RM is considered to be a viable technique for the very fast dose application in patient-specific tumors with only one fixed energy. The RM was developed based on a tumor from a patient CT and manufactured with high-quality 3D printing techniques with both polymer resin and aluminum. Monte Carlo simulations were utilized to investigate its modulating properties and the resulting dose distribution. Additionally, the simulation results were validated with measurements at the Marburg Ion-Beam Therapy Centre. For this purpose, a previously developed water phantom was used to conduct fast, automated high-resolution dose measurements. The results show a very good agreement between simulations and measurements and indicate that highly homogeneous dose distributions are possible. The delivered dose is conformed to the distal as well as to the proximal edge of the target. The 3D range-modulator concept combines a high degree of dose homogeneity and conformity, comparable to standard IMPT with very short irradiation times, promising clinically applicable dose distributions for lung and/or FLASH treatment, comparable and competitive to those from conventional irradiation techniques.
- 11 March 2022
- Published by IOP Publishing in Biomedical Physics & Engineering Express
- Vol. 8 (3) , 035006
Open AccessDevelopment, Monte Carlo simulations and experimental evaluation of a 3D range-modulator for a complex target in scanned proton therapy
The purpose of this work was to develop and manufacture a 3D range-modulator (3D RM) for a complex target contour for scanned proton therapy. The 3D RM is considered to be a viable technique for the very fast dose application in patient-specific tumors with only one fixed energy. The RM was developed based on a tumor from a patient CT and manufactured with high-quality 3D printing techniques with both polymer resin and aluminum. Monte Carlo simulations were utilized to investigate its modulating properties and the resulting dose distribution. Additionally, the simulation results were validated with measurements at the Marburg Ion-Beam Therapy Centre. For this purpose, a previously developed water phantom was used to conduct fast, automated high-resolution dose measurements. The results show a very good agreement between simulations and measurements and indicate that highly homogeneous dose distributions are possible. The delivered dose is conformed to the distal as well as to the proximal edge of the target. The 3D range-modulator concept combines a high degree of dose homogeneity and conformity, comparable to standard IMPT with very short irradiation times, promising clinically applicable dose distributions for lung and/or FLASH treatment, comparable and competitive to those from conventional irradiation techniques.
- 25 February 2022
- Published by Elsevier in Physica Medica
- Vol. 94, S112
3D RANGE-MODULATORS: DOSE SIMULATIONS UNDER THE ASPECT OF POTENTIAL FLASH IRRADIATION WITH PROTONS
- 25 February 2022
- Published by Elsevier in Physica Medica
- Vol. 94, S112
3D RANGE-MODULATORS: DOSE SIMULATIONS UNDER THE ASPECT OF POTENTIAL FLASH IRRADIATION WITH PROTONS
- 1 May 2021
- Published by Elsevier in Zeitschrift für Medizinische Physik
- Vol. 31 (2) , 203-214
Monte Carlo simulations and dose measurements of 2D range-modulators for scanned particle therapy
- 1 May 2021
- Published by Elsevier in Zeitschrift für Medizinische Physik
- Vol. 31 (2) , 203-214
Monte Carlo simulations and dose measurements of 2D range-modulators for scanned particle therapy
- 20 September 2017
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 62 (19) , 7923
Corrigendum: 3D range-modulator for scanned particle therapy: development, Monte Carlo simulations and experimental evaluation (2017 Phys. Med. Biol . 62 7075)
- 20 September 2017
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 62 (19) , 7923
Corrigendum: 3D range-modulator for scanned particle therapy: development, Monte Carlo simulations and experimental evaluation (2017 Phys. Med. Biol . 62 7075)
- 11 August 2017
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 62 (17) , 7075-7096
3D range-modulator for scanned particle therapy: development, Monte Carlo simulations and experimental evaluation
The purpose of this work was to design and manufacture a 3D range-modulator for scanned particle therapy. The modulator is intended to create a highly conformal dose distribution with only one fixed energy, simultaneously reducing considerably the treatment time. As a proof of concept, a 3D range-modulator was developed for a spherical target volume with a diameter of 5 cm, placed at a depth of 25 cm in a water phantom. It consists of a large number of thin pins with a well-defined shape and different lengths to modulate the necessary shift of the Bragg peak. The 3D range-modulator was manufactured with a rapid prototyping technique. The FLUKA Monte Carlo package was used to simulate the modulating effect of the 3D range-modulator and the resulting dose distribution. For that purpose, a special user routine was implemented to handle its complex geometrical contour. Additionally, FLUKA was extended with the capability of intensity modulated scanning. To validate the simulation results, dose measurements were carried out at the Heidelberg Ion Beam Therapy Center with a 400.41 MeV/u 12 C beam. The high resolution dosimetric measurements show a good agreement between simulated and measured dose distributions. Irradiation of the monoenergetic raster plan took 3 s, which is approximately 20 times shorter than a comparable plan with 16 different energies. The combination of only one energy and a 3D range-modulator leads to a tremendous decrease in irradiation time. ‘Interplay effects’, typical for moving targets and pencil beam scanning, can be immensely reduced or disappear completely, making the delivery of a homogeneous dose to moving targets more reliable. Combining high dose conformity, very good homogeneity and extremely short irradiation times, the 3D range-modulator is considered to become a clinically applicable method for very fast treatment of lung tumours.
- 11 August 2017
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 62 (17) , 7075-7096
3D range-modulator for scanned particle therapy: development, Monte Carlo simulations and experimental evaluation
The purpose of this work was to design and manufacture a 3D range-modulator for scanned particle therapy. The modulator is intended to create a highly conformal dose distribution with only one fixed energy, simultaneously reducing considerably the treatment time. As a proof of concept, a 3D range-modulator was developed for a spherical target volume with a diameter of 5 cm, placed at a depth of 25 cm in a water phantom. It consists of a large number of thin pins with a well-defined shape and different lengths to modulate the necessary shift of the Bragg peak. The 3D range-modulator was manufactured with a rapid prototyping technique. The FLUKA Monte Carlo package was used to simulate the modulating effect of the 3D range-modulator and the resulting dose distribution. For that purpose, a special user routine was implemented to handle its complex geometrical contour. Additionally, FLUKA was extended with the capability of intensity modulated scanning. To validate the simulation results, dose measurements were carried out at the Heidelberg Ion Beam Therapy Center with a 400.41 MeV/u 12 C beam. The high resolution dosimetric measurements show a good agreement between simulated and measured dose distributions. Irradiation of the monoenergetic raster plan took 3 s, which is approximately 20 times shorter than a comparable plan with 16 different energies. The combination of only one energy and a 3D range-modulator leads to a tremendous decrease in irradiation time. ‘Interplay effects’, typical for moving targets and pencil beam scanning, can be immensely reduced or disappear completely, making the delivery of a homogeneous dose to moving targets more reliable. Combining high dose conformity, very good homogeneity and extremely short irradiation times, the 3D range-modulator is considered to become a clinically applicable method for very fast treatment of lung tumours.