Yaqiong Fu
Current Organizations / Institutes
The organizations/institutes from the scholar's latest article
Yaqiong Fu has a total of 18 co-authors. Here are the top co-authors by number of shared publications:
- Lushuai Qian
College of Mechanical and Electrical Engineering, China Jiliang University, Hangzhou, China
3 shared publications - Suan Xu
College of Mechanical and Electrical Engineering, China Jiliang University, Hangzhou, China
3 shared publications - Changwei Zhai
School of Instrument Science and Engineering, Harbin Institute of Technology, Harbin 150080, China
1 shared publication - Hao Wang
National Institute of Metrology, Beijing, China
1 shared publication - Hong-Wei Xu
China Jiliang University
1 shared publication - Jianting Zhao
National Institute of Metrology, Beijing, China
1 shared publication
Latest Publications
- 16 April 2026
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Instrumentation and Measurement
- Vol. 75 (00189456) , 1-10
Measurement and Correction of the Resistor Load Effect Under Low-Pressure Conditions
The measurement of the load effect of the shunt resistor under low-pressure conditions is necessary for accurate current measurement, especially in aerospace applications such as power system fault diagnosis and monitoring, where shunt resistors are used for high-current sensing. In this paper, a measurement system to evaluate the resistor load effect under low-pressure conditions is designed and experimentally investigated, featuring a resistance measurement module, a pressure control module, and a temperature control module. In the pressure control module, a vacuum pump, an air reservoir, solenoid valves, and pressure sensors are applied to set and maintain various pressure test conditions in a vacuum chamber. The air reservoir is connected in series between the vacuum pump and the vacuum chamber to enhance the stability of pressure control. Experiments are performed to quantify the resistor load effect under low-pressure conditions by measuring its temperature, load, and pressure coefficients. Afterward, the resistance variation with temperature, load power, and pressure is characterized by polynomial models, which is then used to establish a resistance error correction model. Finally, a current compensation scheme is designed to verify the improvement in current measurement accuracy achieved by correction of the shunt resistor load effect under low-pressure conditions. The accuracy of the corrected current measurement is improved by one order of magnitude.
- 16 April 2026
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Instrumentation and Measurement
- Vol. 75 (00189456) , 1-10
Measurement and Correction of the Resistor Load Effect Under Low-Pressure Conditions
The measurement of the load effect of the shunt resistor under low-pressure conditions is necessary for accurate current measurement, especially in aerospace applications such as power system fault diagnosis and monitoring, where shunt resistors are used for high-current sensing. In this paper, a measurement system to evaluate the resistor load effect under low-pressure conditions is designed and experimentally investigated, featuring a resistance measurement module, a pressure control module, and a temperature control module. In the pressure control module, a vacuum pump, an air reservoir, solenoid valves, and pressure sensors are applied to set and maintain various pressure test conditions in a vacuum chamber. The air reservoir is connected in series between the vacuum pump and the vacuum chamber to enhance the stability of pressure control. Experiments are performed to quantify the resistor load effect under low-pressure conditions by measuring its temperature, load, and pressure coefficients. Afterward, the resistance variation with temperature, load power, and pressure is characterized by polynomial models, which is then used to establish a resistance error correction model. Finally, a current compensation scheme is designed to verify the improvement in current measurement accuracy achieved by correction of the shunt resistor load effect under low-pressure conditions. The accuracy of the corrected current measurement is improved by one order of magnitude.
- 29 May 2025
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Instrumentation and Measurement
- Vol. 74 (00189456) , 1-11
Development of DC Power Standard Source Based on Programmable Josephson Voltage Standard
To meet the growing demand for high-precision DC power measurement and calibration, this paper presents a quantum-based DC power standard developed with a “standard source” architecture. This system integrates a compact heliumfree dual-channel programmable Josephson voltage standard, a three-stage cascaded high-voltage amplifier with low common-mode rejection ratio error, a large-current transconductance amplifier based on DC current transformer, voltage/current ratio self-calibration device, and other essential components. The design principles and key performance metrics of these core components are thoroughly described. Test results indicate that the system achieves a combined uncertainty of 1.5 × 10−6 (k=1) at DC 1000V and 120 A. When used with a time pulse counter, the system can also serves as a direct calibration reference for DC energy meters. Finally, this paper discusses potential future advancements in quantum-based standards for DC power calibration technology.
- 29 May 2025
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Instrumentation and Measurement
- Vol. 74 (00189456) , 1-11
Development of DC Power Standard Source Based on Programmable Josephson Voltage Standard
To meet the growing demand for high-precision DC power measurement and calibration, this paper presents a quantum-based DC power standard developed with a “standard source” architecture. This system integrates a compact heliumfree dual-channel programmable Josephson voltage standard, a three-stage cascaded high-voltage amplifier with low common-mode rejection ratio error, a large-current transconductance amplifier based on DC current transformer, voltage/current ratio self-calibration device, and other essential components. The design principles and key performance metrics of these core components are thoroughly described. Test results indicate that the system achieves a combined uncertainty of 1.5 × 10−6 (k=1) at DC 1000V and 120 A. When used with a time pulse counter, the system can also serves as a direct calibration reference for DC energy meters. Finally, this paper discusses potential future advancements in quantum-based standards for DC power calibration technology.
- Open Access
- 10 September 2024
- Published by MDPI AG in Electronics
- Vol. 13 (18) , 3596
A Wireless Ad Hoc Network Communication Platform and Data Transmission Strategies for Multi-Bus Instruments
As automatic test technology advances, the number of programmable instruments in a single test system increases. Traditional wired communication methods have a limited range and involve complex cable layouts. Single-function wireless converters provide a viable alternative, but they have limitations. These include complicated configuration, issues with multi-system collaboration, and data blocking. This paper proposes a wireless ad hoc network platform for multi-bus instruments based on a low-cost ESP-12H WiFi module. The platform supports GPIB, RS232, RS485, and CAN bus interface instrument access. It features easy configuration, ad hoc networking, and self-repairing capabilities. A relay multi-hop network with a tree topology expands capacity and coverage. Additionally, a dynamic window-receiving mode and an improved multi-priority queue ensure data transmission integrity. The experimental results show that the platform’s networking time is less than 10 s, and the coverage range reaches 50 m in complex indoor environments. It also shows good stability when running for a long time. However, due to hardware and software design limitations, the actual upload speeds fall short of the theoretical values. For example, RS232 and RS485 are about 10% slower than the theoretical values, and GPIB is about 80% slower. Further optimization is required in the future.
- 10 September 2024
- Published by MDPI AG in Electronics
- Vol. 13 (18) , 3596
Open AccessA Wireless Ad Hoc Network Communication Platform and Data Transmission Strategies for Multi-Bus Instruments
As automatic test technology advances, the number of programmable instruments in a single test system increases. Traditional wired communication methods have a limited range and involve complex cable layouts. Single-function wireless converters provide a viable alternative, but they have limitations. These include complicated configuration, issues with multi-system collaboration, and data blocking. This paper proposes a wireless ad hoc network platform for multi-bus instruments based on a low-cost ESP-12H WiFi module. The platform supports GPIB, RS232, RS485, and CAN bus interface instrument access. It features easy configuration, ad hoc networking, and self-repairing capabilities. A relay multi-hop network with a tree topology expands capacity and coverage. Additionally, a dynamic window-receiving mode and an improved multi-priority queue ensure data transmission integrity. The experimental results show that the platform’s networking time is less than 10 s, and the coverage range reaches 50 m in complex indoor environments. It also shows good stability when running for a long time. However, due to hardware and software design limitations, the actual upload speeds fall short of the theoretical values. For example, RS232 and RS485 are about 10% slower than the theoretical values, and GPIB is about 80% slower. Further optimization is required in the future.
- Open Access
Research on Weigh-in-Motion Algorithm of Vehicles Based on BSO-BP
Weigh-in-motion (WIM) systems are used to measure the weight of moving vehicles. Aiming at the problem of low accuracy of the WIM system, this paper proposes a WIM model based on the beetle swarm optimization (BSO) algorithm and the error back propagation (BP) neural network. Firstly, the structure and principle of the WIM system used in this paper are analyzed. Secondly, the WIM signal is denoised and reconstructed by wavelet transform. Then, a BP neural network model optimized by BSO algorithm is established to process the WIM signal. Finally, the predictive ability of BP neural network models optimized by different algorithms are compared and conclusions are drawn. The experimental results show that the BSO-BP WIM model has fast convergence speed, high accuracy, the relative error of the maximum gross weight is 1.41%, and the relative error of the maximum axle weight is 6.69%.
Open AccessResearch on Weigh-in-Motion Algorithm of Vehicles Based on BSO-BP
Weigh-in-motion (WIM) systems are used to measure the weight of moving vehicles. Aiming at the problem of low accuracy of the WIM system, this paper proposes a WIM model based on the beetle swarm optimization (BSO) algorithm and the error back propagation (BP) neural network. Firstly, the structure and principle of the WIM system used in this paper are analyzed. Secondly, the WIM signal is denoised and reconstructed by wavelet transform. Then, a BP neural network model optimized by BSO algorithm is established to process the WIM signal. Finally, the predictive ability of BP neural network models optimized by different algorithms are compared and conclusions are drawn. The experimental results show that the BSO-BP WIM model has fast convergence speed, high accuracy, the relative error of the maximum gross weight is 1.41%, and the relative error of the maximum axle weight is 6.69%.