Guohua Fu
Current Organizations / Institutes
The organizations/institutes from the scholar's latest article
Previous Organizations / Institutes
The organizations / institutes from the scholar's other articles in the past ten years
Research Subjects
Guohua Fu has a total of 64 co-authors. Here are the top co-authors by number of shared publications:
- Huimin Chu
Arrhythmia Center, Ningbo First Hospital The First Affiliated Hospital of Ningbo University Ningbo China
6 shared publications Mingjun Feng
Arrhythmia Center, the First Affiliated Hospital of Ningbo University, Ningbo, China
4 shared publicationsBinhao Wang
Arrhythmia Center, the First Affiliated Hospital of Ningbo University, Ningbo, China
3 shared publications- Binhao Wang
Arrhythmia Center, Ningbo First Hospital, Ningbo 315000, China
3 shared publications Fang Gao
Arrhythmia Center, the First Affiliated Hospital of Ningbo University, Ningbo, China
3 shared publicationsWeidong Zhuo
Arrhythmia Center, the First Affiliated Hospital of Ningbo University, Ningbo, China
3 shared publications
Latest Publications
- Open Access
- 26 November 2024
- Published by Frontiers Media SA in Frontiers in Cardiovascular Medicine
- Vol. 11, 1449623
Ablation index value for transmural lesions based on unipolar electrograms in patients with paroxysmal atrial fibrillation undergoing pulmonary vein isolation
Background: It remains unclear whether the current recommended ablation index (AI) value is suitable for individualized catheter ablation. Prior research has established that the elimination of the negative component of the unipolar electrogram (UP-EGM) applications reflects the formation of transmural lesion during radiofrequency ablation. The aim of this study was to explore the relationship between AI values when UP-EGM turns positive during pulmonary vein isolation and recommended AI values.Methods: A total of 50 patients with drug-refractory PAF who underwent index RFCA were consecutively included from September 2022 to January 2023. All the patients underwent AI-guided ablation. UP-EGM was also recorded during the procedure. The difference in the AI between the value when the UP-EGM turned completely positive [AIUP-EGM(+)] and the recommended value at the end of ablation (AIEND) was compared.Results: A total of 2 954 lesion points were detected in 50 patients. The average values of AIUP-EGM(+) at the anterior wall and the posterior wall were 420.9 and 267.4, respectively. The average AIEND values were 524.3 and 393.9 at the anterior wall and the posterior wall, respectively. The percentage of increase in the AI between the AIUP-EGM(+) and AIEND groups was 22%, 28% at the anterior wall and 47%, 49% at the posterior wall (P < 0.001). After a mean follow-up duration of 11.30 ± 2.10 months, 44 patients (88%) remained in sinus rhythm without antiarrhythmic drugs.Conclusion: The AIUP-EGM(+) was lower than the recommended value for all the pulmonary vein regions. The recommended AI value seems to be too high for the posterior and inferior walls, but this remains to be proven in future research.
- 26 November 2024
- Published by Frontiers Media SA in Frontiers in Cardiovascular Medicine
- Vol. 11, 1449623
Open AccessAblation index value for transmural lesions based on unipolar electrograms in patients with paroxysmal atrial fibrillation undergoing pulmonary vein isolation
Background: It remains unclear whether the current recommended ablation index (AI) value is suitable for individualized catheter ablation. Prior research has established that the elimination of the negative component of the unipolar electrogram (UP-EGM) applications reflects the formation of transmural lesion during radiofrequency ablation. The aim of this study was to explore the relationship between AI values when UP-EGM turns positive during pulmonary vein isolation and recommended AI values.Methods: A total of 50 patients with drug-refractory PAF who underwent index RFCA were consecutively included from September 2022 to January 2023. All the patients underwent AI-guided ablation. UP-EGM was also recorded during the procedure. The difference in the AI between the value when the UP-EGM turned completely positive [AIUP-EGM(+)] and the recommended value at the end of ablation (AIEND) was compared.Results: A total of 2 954 lesion points were detected in 50 patients. The average values of AIUP-EGM(+) at the anterior wall and the posterior wall were 420.9 and 267.4, respectively. The average AIEND values were 524.3 and 393.9 at the anterior wall and the posterior wall, respectively. The percentage of increase in the AI between the AIUP-EGM(+) and AIEND groups was 22%, 28% at the anterior wall and 47%, 49% at the posterior wall (P < 0.001). After a mean follow-up duration of 11.30 ± 2.10 months, 44 patients (88%) remained in sinus rhythm without antiarrhythmic drugs.Conclusion: The AIUP-EGM(+) was lower than the recommended value for all the pulmonary vein regions. The recommended AI value seems to be too high for the posterior and inferior walls, but this remains to be proven in future research.
- Open Access
- 29 October 2024
- Published by Wiley
Predictors of delayed pericardial effusion or tamponade after left atrial appendage closure
Background: Delayed pericardial effusion or pericardial tamponade (PE/PT) is a relatively rare complication of left atrial appendage closure (LAAC) but has serious consequences for affected patients. There are limited data related to delayed PE/PT. The aim of this study was to analyze the incidence and potential risks of delayed PE/PT following LAAC. Methods: Patients with nonvalvular atrial fibrillation who were successfully implanted with LAAC devices from October 2014 to July 2023 were screened retrospectively. Patients were divided into two groups according to whether delayed PE/PT occurred. Univariate and multivariate analyses were used to evaluate the potential risks of delayed PE/PT. Results: A total of 1866 patients with successfully implanted LAAC devices were included. Fifty-two patients were excluded because of missing transesophageal echocardiogram or cardiac CT angiogram data 45 days after the procedure, and 106 patients were excluded because of loss to follow-up. Thirty-seven patients were excluded because of acute PE/PT during perioperative period. Among the remaining 1671 patients, 11 developed delayed PE/PT during the follow-up period (2 Watchman2.5, 2 ACP and 7 LAmbre). Ten of these patients recovered after pericardiocentesis, and one patient required cardiac surgery. Patients with delayed PE/PT had a larger left atrial appendage (LAA) orifice diameter, a higher rate of nitinol plug device use and a lower incidence of systolic heart failure. Univariate and multivariate analyses revealed that the use of a nitinol plug device was associated with delayed PE/PT (OR=1.819 95% CI: 0.757-4.343, P=0.018). A larger maximal LAA orifice diameter was also associated with the occurrence of delayed PE/PT (OR=1.473 95% CI: 1.163-1.866, P=0.001). Conclusion: Delayed PE/PT is a rare complication of LAAC device implantation and is related to the use of a nitinol plug device rather than a nitinol cage device. Patients with larger LAA orifice diameters are at greater risk of delayed PE/PT after LAAC.
- 29 October 2024
- Published by Wiley
Open AccessPredictors of delayed pericardial effusion or tamponade after left atrial appendage closure
Background: Delayed pericardial effusion or pericardial tamponade (PE/PT) is a relatively rare complication of left atrial appendage closure (LAAC) but has serious consequences for affected patients. There are limited data related to delayed PE/PT. The aim of this study was to analyze the incidence and potential risks of delayed PE/PT following LAAC. Methods: Patients with nonvalvular atrial fibrillation who were successfully implanted with LAAC devices from October 2014 to July 2023 were screened retrospectively. Patients were divided into two groups according to whether delayed PE/PT occurred. Univariate and multivariate analyses were used to evaluate the potential risks of delayed PE/PT. Results: A total of 1866 patients with successfully implanted LAAC devices were included. Fifty-two patients were excluded because of missing transesophageal echocardiogram or cardiac CT angiogram data 45 days after the procedure, and 106 patients were excluded because of loss to follow-up. Thirty-seven patients were excluded because of acute PE/PT during perioperative period. Among the remaining 1671 patients, 11 developed delayed PE/PT during the follow-up period (2 Watchman2.5, 2 ACP and 7 LAmbre). Ten of these patients recovered after pericardiocentesis, and one patient required cardiac surgery. Patients with delayed PE/PT had a larger left atrial appendage (LAA) orifice diameter, a higher rate of nitinol plug device use and a lower incidence of systolic heart failure. Univariate and multivariate analyses revealed that the use of a nitinol plug device was associated with delayed PE/PT (OR=1.819 95% CI: 0.757-4.343, P=0.018). A larger maximal LAA orifice diameter was also associated with the occurrence of delayed PE/PT (OR=1.473 95% CI: 1.163-1.866, P=0.001). Conclusion: Delayed PE/PT is a rare complication of LAAC device implantation and is related to the use of a nitinol plug device rather than a nitinol cage device. Patients with larger LAA orifice diameters are at greater risk of delayed PE/PT after LAAC.
- Open Access
- 27 May 2024
- Published by Oxford University Press (OUP) in EP Europace
- Vol. 26 (7)
Efficacy and safety of pulsed field ablation for accessory pathways: a pilot study
Aims: Radiofrequency ablation is used as a first-line therapy for accessory pathways (APs). However, data regarding the effects of pulsed field ablation (PFA) on APs are limited. We sought to evaluate the acute procedural and 6-month success and safety of PFA in a cohort of patients with APs. Methods and results: A focal contact force–sensing PFA catheter was used for patients with APs. Pulsed field ablation generator generated a bipolar and biphasic waveform (±1000 V) with a duration of 100 ms from the tip of the PFA catheter. A 100% acute procedural success was achieved in 10 conscious patients with APs (7 left anterolateral, 2 left inferolateral, and 1 right posteroseptal APs) including 6 (60%) patients after an initial application. The average total ablation time was 6.3 ± 4.9 s for 4.7 ± 1.8 ablation sites (ASs), including 3.1 ± 2.4 s at targets and 3.2 ± 2.9 s at 3.2 ± 2 bolus ASs. The mean skin-to-skin time was 59.3 ± 15.5 min, and PFA catheter dwell time was 29.4 ± 7.8 min. One patient encountered transient sinus arrest during PFA due to parasympathetic overexcitation. Sinus rhythm was restored in all patients without any significant adverse events during the short-term follow-up. Conclusion: Pulsed field ablation of APs was feasible, effective, and safe. Its efficiency was remarkable for its ultrarapid termination of AP conduction. Further studies are warranted to prove whether utilization of PFA with current parameters can extend to manifold AP ablation.
- 27 May 2024
- Published by Oxford University Press (OUP) in EP Europace
- Vol. 26 (7)
Open AccessEfficacy and safety of pulsed field ablation for accessory pathways: a pilot study
Aims: Radiofrequency ablation is used as a first-line therapy for accessory pathways (APs). However, data regarding the effects of pulsed field ablation (PFA) on APs are limited. We sought to evaluate the acute procedural and 6-month success and safety of PFA in a cohort of patients with APs. Methods and results: A focal contact force–sensing PFA catheter was used for patients with APs. Pulsed field ablation generator generated a bipolar and biphasic waveform (±1000 V) with a duration of 100 ms from the tip of the PFA catheter. A 100% acute procedural success was achieved in 10 conscious patients with APs (7 left anterolateral, 2 left inferolateral, and 1 right posteroseptal APs) including 6 (60%) patients after an initial application. The average total ablation time was 6.3 ± 4.9 s for 4.7 ± 1.8 ablation sites (ASs), including 3.1 ± 2.4 s at targets and 3.2 ± 2.9 s at 3.2 ± 2 bolus ASs. The mean skin-to-skin time was 59.3 ± 15.5 min, and PFA catheter dwell time was 29.4 ± 7.8 min. One patient encountered transient sinus arrest during PFA due to parasympathetic overexcitation. Sinus rhythm was restored in all patients without any significant adverse events during the short-term follow-up. Conclusion: Pulsed field ablation of APs was feasible, effective, and safe. Its efficiency was remarkable for its ultrarapid termination of AP conduction. Further studies are warranted to prove whether utilization of PFA with current parameters can extend to manifold AP ablation.
- 24 August 2022
- Published by Wolters Kluwer Health in Zhonghua Xin Xue Guan Bing Za Zhi
- Vol. 50 (8) , 780-784
[Feasibility and safety of closing large left atrial appendage using the LAmbre device].
Objective: To evaluate the feasibility and safety of the LAmbre occluder for large-diameter left atrial appendage occlusion. Methods: This study was a retrospective cohort study. Patients with large orifice of the left atrial appendage (≥31 mm) and occlusion with the LAmbre device in the Arrhythmia Center of Ningbo First Hospital were included from June 2018 to March 2020. Baseline data were collected and major perioperative complications of left atrial appendage occlusion (including death, stroke, instrumental embolism, cardiac tamponade, and major bleeding events) were recorded. Patients were followed up 45 days, 6 months and 12 months after surgery. The shunt and device-related thrombosis were recorded by esophageal cardiac ultrasound or pulmonary vein CT, and the occurrence of postoperative thromboembolism, bleeding events, death and other serious adverse events were recorded. Results: The average age and left atrial appendage ostial dimension of 32 patients (37.5% women) included in this research were (70.4±8.4) years old and (34.4±2.9) mm. The LAmbre device was successfully implanted in 31(96.9%) patients. No major complications occurred during the perioperative period. During the 12-month follow-up, pericardial tamponade occurred in 1(3.2%) patient and was recovered after treatment. There was no occluder edge shunt>5 mm in patients followed up by esophageal echocardiography. No significant peri-device leak, device-related thrombus, thromboembolism or death event has occurred. Conclusion: The LAmbre occluder may be feasible and safe for large-diameter left atrial appendage occlusion.
- 24 August 2022
- Published by Wolters Kluwer Health in Zhonghua Xin Xue Guan Bing Za Zhi
- Vol. 50 (8) , 780-784
[Feasibility and safety of closing large left atrial appendage using the LAmbre device].
Objective: To evaluate the feasibility and safety of the LAmbre occluder for large-diameter left atrial appendage occlusion. Methods: This study was a retrospective cohort study. Patients with large orifice of the left atrial appendage (≥31 mm) and occlusion with the LAmbre device in the Arrhythmia Center of Ningbo First Hospital were included from June 2018 to March 2020. Baseline data were collected and major perioperative complications of left atrial appendage occlusion (including death, stroke, instrumental embolism, cardiac tamponade, and major bleeding events) were recorded. Patients were followed up 45 days, 6 months and 12 months after surgery. The shunt and device-related thrombosis were recorded by esophageal cardiac ultrasound or pulmonary vein CT, and the occurrence of postoperative thromboembolism, bleeding events, death and other serious adverse events were recorded. Results: The average age and left atrial appendage ostial dimension of 32 patients (37.5% women) included in this research were (70.4±8.4) years old and (34.4±2.9) mm. The LAmbre device was successfully implanted in 31(96.9%) patients. No major complications occurred during the perioperative period. During the 12-month follow-up, pericardial tamponade occurred in 1(3.2%) patient and was recovered after treatment. There was no occluder edge shunt>5 mm in patients followed up by esophageal echocardiography. No significant peri-device leak, device-related thrombus, thromboembolism or death event has occurred. Conclusion: The LAmbre occluder may be feasible and safe for large-diameter left atrial appendage occlusion.
- Open Access
- 18 July 2022
- Published by MDPI AG in Journal of Cardiovascular Development and Disease
- Vol. 9 (7) , 229
Unipolar Electrogram-Guided versus Lesion Size Index-Guided Catheter Ablation in Patients with Paroxysmal Atrial Fibrillation
Background: This research explores the relationship between the unipolar electrogram (UP-EGM) and lesion size index (LSI) in different regions of continuous circular lesions (CCLs) and to assess the safety and efficacy of UP-EGM-guided versus LSI-guided radiofrequency catheter ablation (RFCA) in patients with paroxysmal atrial fibrillation (PAF). Methods: A total of 120 patients with drug-refractory PAF who underwent index RFCA were scheduled to be consecutively included from March 2020 to April 2021. All the patients were randomly divided 1:1 into two groups: the UP-EGM group and the LSI group. The first-pass PVI rate, acute PVI rate, and the sinus rhythm maintenance rate were compared. Results: A total of 120 patients with PAF were included in the study: the UP-EGM group (n = 60) and the LSI group (n = 60). All the LSI values in the UP-EGM group were less than those in the corresponding regions in the LSI group (all p < 0.001). There were no significant differences in the first-pass PVI rate and acute PVI rate between the two groups. After a mean follow-up period of 11.31 ± 1.70 months, the sinus rhythm maintenance rate in the UP-EGM group was comparable to that in the LSI group (90% vs. 91.7%, p = 0.752). Conclusion: UP-EGM-guided and LSI-guided RFCA are both effective and safe in patients with PAF. However, UP-EGM may be more suitable than LSI for guiding individual RFCA.
- 18 July 2022
- Published by MDPI AG in Journal of Cardiovascular Development and Disease
- Vol. 9 (7) , 229
Open AccessUnipolar Electrogram-Guided versus Lesion Size Index-Guided Catheter Ablation in Patients with Paroxysmal Atrial Fibrillation
Background: This research explores the relationship between the unipolar electrogram (UP-EGM) and lesion size index (LSI) in different regions of continuous circular lesions (CCLs) and to assess the safety and efficacy of UP-EGM-guided versus LSI-guided radiofrequency catheter ablation (RFCA) in patients with paroxysmal atrial fibrillation (PAF). Methods: A total of 120 patients with drug-refractory PAF who underwent index RFCA were scheduled to be consecutively included from March 2020 to April 2021. All the patients were randomly divided 1:1 into two groups: the UP-EGM group and the LSI group. The first-pass PVI rate, acute PVI rate, and the sinus rhythm maintenance rate were compared. Results: A total of 120 patients with PAF were included in the study: the UP-EGM group (n = 60) and the LSI group (n = 60). All the LSI values in the UP-EGM group were less than those in the corresponding regions in the LSI group (all p < 0.001). There were no significant differences in the first-pass PVI rate and acute PVI rate between the two groups. After a mean follow-up period of 11.31 ± 1.70 months, the sinus rhythm maintenance rate in the UP-EGM group was comparable to that in the LSI group (90% vs. 91.7%, p = 0.752). Conclusion: UP-EGM-guided and LSI-guided RFCA are both effective and safe in patients with PAF. However, UP-EGM may be more suitable than LSI for guiding individual RFCA.
- Open Access
- 31 May 2022
- Published by MDPI AG in Journal of Clinical Medicine
- Vol. 11 (11) , 3110
Safety and Efficacy of Left Atrial Catheter Ablation in Patients with Left Atrial Appendage Occlusion Devices
Background: Left atrial appendage occlusion (LAAO) is an alternative to oral anticoagulation for thromboembolic prevention in patients with atrial fibrillation (AF). Left atrial (LA) catheter ablation (CA) in patients with LAAO devices has not been well investigated. Here, we report on the safety and efficacy of LA CA in patients with nitinol cage or plug LAAO devices. Methods: A total of 18 patients (aged 67 ± 11 years; 14 males; 5 paroxysmal AF) with LAAO devices (nitinol cage, n = 10; nitinol plug, n = 8) and symptomatic LA tachyarrhythmias were included. Periprocedural and follow-up data were assessed. Results: A total of 20 LA CA procedures were performed at a median of 130 (63, 338) days after LAAO. The strategy of CA consisted of circumferential pulmonary vein isolation (n = 16), linear lesions (n = 14) and complex fractionated atrial electrogram ablation (n = 6). No major adverse events occurred periprocedurally. Repeated transesophageal echocardiography showed no device-related thrombus, newly developed peridevice leakage or device dislodgement. After a median follow-up period of 793 (376, 1090) days, four patients (22%) experienced LA tachyarrhythmias recurrence and two received redo LA CA. No patients suffered stroke or major bleeding events during follow-up. Conclusions: LA CA in patients with LAAO devices (either nitinol cages or nitinol plugs) seems to be safe and efficient in our single-center experience.
- 31 May 2022
- Published by MDPI AG in Journal of Clinical Medicine
- Vol. 11 (11) , 3110
Open AccessSafety and Efficacy of Left Atrial Catheter Ablation in Patients with Left Atrial Appendage Occlusion Devices
Background: Left atrial appendage occlusion (LAAO) is an alternative to oral anticoagulation for thromboembolic prevention in patients with atrial fibrillation (AF). Left atrial (LA) catheter ablation (CA) in patients with LAAO devices has not been well investigated. Here, we report on the safety and efficacy of LA CA in patients with nitinol cage or plug LAAO devices. Methods: A total of 18 patients (aged 67 ± 11 years; 14 males; 5 paroxysmal AF) with LAAO devices (nitinol cage, n = 10; nitinol plug, n = 8) and symptomatic LA tachyarrhythmias were included. Periprocedural and follow-up data were assessed. Results: A total of 20 LA CA procedures were performed at a median of 130 (63, 338) days after LAAO. The strategy of CA consisted of circumferential pulmonary vein isolation (n = 16), linear lesions (n = 14) and complex fractionated atrial electrogram ablation (n = 6). No major adverse events occurred periprocedurally. Repeated transesophageal echocardiography showed no device-related thrombus, newly developed peridevice leakage or device dislodgement. After a median follow-up period of 793 (376, 1090) days, four patients (22%) experienced LA tachyarrhythmias recurrence and two received redo LA CA. No patients suffered stroke or major bleeding events during follow-up. Conclusions: LA CA in patients with LAAO devices (either nitinol cages or nitinol plugs) seems to be safe and efficient in our single-center experience.
- Open Access
- 1 October 2021
- Published by Oxford University Press (OUP) in European Heart Journal
- Vol. 42 (Supplement)
High precision machine learning-enabled ECG algorithm for predicting sites of idiopathic ventricular arrhythmia origin
Introduction: Radiofrequency catheter ablation (CA) is an efficient antiarrhythmic treatment with a class I indication for idiopathic ventricular arrhythmia (IVA). The accurate prediction of the origins of IVA can significantly increase the procedure success rate, reduce operation duration and decrease the risk of complications. The present work proposes an ECG analysis algorithm to estimate 21 possible origins of idiopathic ventricular arrhythmia at a clinical-grade level accuracy, which include left coronary cusp (LCC), right coronary cusp (RCC), aortomitral continuity (AMC), summit, LCC-RCC commissure, left His bundle, mitral valve (MV), left septal including left anterior fascicle (LAF), left posterior fascicle (LPF), left anterior papillary muscle (LAPM), left posterior papillary muscle (LPPM), anterior cusp (AC), left cusp (LC), right cusp (RC), RVOT septal, free wall, right His bundle, tricuspid valve (TV), and right anterior papillary muscle (RAPM). Method: A total of 18,612 ECG recordings extracted from 545 patients who underwent successful CA to treat IVA were proportionally sampled into training, validation and testing cohorts. We designed four classification schemes responding to different hierarchical levels of the possible IVA origins. The first scheme will help the operators to figure out the origin from epicardium of left ventricular summit, right, and left ventricle. The second one can separate origins from left/right outflow tract and left/right non-out flow tract, respectively. The third one is able to predict 18 anatomical locations, and the fourth scheme can distinguish 21 possible sites. For every classification scheme, we compared 98 distinct machine learning models with optimized hyperparameter values obtained through extensive grid search and reported an optimal algorithm with the highest accuracy scores attained on the validation cohorts. Results: In the first classification scheme used to predict right ventricular endocardium, left ventricular endocardium, and epicardium of left ventricular summit, the model achieved an accuracy of 99.79 (99.41–99.89) and a F1-score of 99.84 (99.6–99.96). For scheme 2, the proposed method reached an accuracy of 99.62 (99.09–99.78) and a F1-score of 99.42 (98.79–99.75). For scheme 3, the model achieved an accuracy of 97.78 (96.76–98.41), a F1-score of 97.74 (94.15–99.73), and an adjusted accuracy of 98.53 (98.33–99.15). For scheme 4 that can distinguish 21 origin sites, the proposed model attained an accuracy of 98.24 (97.36–98.71), a F1-score of 98.56 (97.88–99.12) and an adjusted accuracy of 98.75 (98.35–99.38). Conclusion: The proposed machine learning model can be immediately and effortlessly deployed to electrophysiology labs allowing cardiologists to predict the exact origins of arrhythmia and provide an optimum treatment plan both before and during the CA procedure. This approach will significantly reduce the CA procedure duration and the risk of complications. Funding Acknowledgement: Type of funding sources: Foundation. Main funding source(s): 2020 Natural Science Foundation of Zhengjiang Province Confusion matrix for classification schemes
- 1 October 2021
- Published by Oxford University Press (OUP) in European Heart Journal
- Vol. 42 (Supplement)
Open AccessHigh precision machine learning-enabled ECG algorithm for predicting sites of idiopathic ventricular arrhythmia origin
Introduction: Radiofrequency catheter ablation (CA) is an efficient antiarrhythmic treatment with a class I indication for idiopathic ventricular arrhythmia (IVA). The accurate prediction of the origins of IVA can significantly increase the procedure success rate, reduce operation duration and decrease the risk of complications. The present work proposes an ECG analysis algorithm to estimate 21 possible origins of idiopathic ventricular arrhythmia at a clinical-grade level accuracy, which include left coronary cusp (LCC), right coronary cusp (RCC), aortomitral continuity (AMC), summit, LCC-RCC commissure, left His bundle, mitral valve (MV), left septal including left anterior fascicle (LAF), left posterior fascicle (LPF), left anterior papillary muscle (LAPM), left posterior papillary muscle (LPPM), anterior cusp (AC), left cusp (LC), right cusp (RC), RVOT septal, free wall, right His bundle, tricuspid valve (TV), and right anterior papillary muscle (RAPM). Method: A total of 18,612 ECG recordings extracted from 545 patients who underwent successful CA to treat IVA were proportionally sampled into training, validation and testing cohorts. We designed four classification schemes responding to different hierarchical levels of the possible IVA origins. The first scheme will help the operators to figure out the origin from epicardium of left ventricular summit, right, and left ventricle. The second one can separate origins from left/right outflow tract and left/right non-out flow tract, respectively. The third one is able to predict 18 anatomical locations, and the fourth scheme can distinguish 21 possible sites. For every classification scheme, we compared 98 distinct machine learning models with optimized hyperparameter values obtained through extensive grid search and reported an optimal algorithm with the highest accuracy scores attained on the validation cohorts. Results: In the first classification scheme used to predict right ventricular endocardium, left ventricular endocardium, and epicardium of left ventricular summit, the model achieved an accuracy of 99.79 (99.41–99.89) and a F1-score of 99.84 (99.6–99.96). For scheme 2, the proposed method reached an accuracy of 99.62 (99.09–99.78) and a F1-score of 99.42 (98.79–99.75). For scheme 3, the model achieved an accuracy of 97.78 (96.76–98.41), a F1-score of 97.74 (94.15–99.73), and an adjusted accuracy of 98.53 (98.33–99.15). For scheme 4 that can distinguish 21 origin sites, the proposed model attained an accuracy of 98.24 (97.36–98.71), a F1-score of 98.56 (97.88–99.12) and an adjusted accuracy of 98.75 (98.35–99.38). Conclusion: The proposed machine learning model can be immediately and effortlessly deployed to electrophysiology labs allowing cardiologists to predict the exact origins of arrhythmia and provide an optimum treatment plan both before and during the CA procedure. This approach will significantly reduce the CA procedure duration and the risk of complications. Funding Acknowledgement: Type of funding sources: Foundation. Main funding source(s): 2020 Natural Science Foundation of Zhengjiang Province Confusion matrix for classification schemes
- Open Access
- 1 May 2021
- Published by Europa Digital & Publishing in EuroIntervention
- Vol. 17 (1) , 90-92
Procedural and short-term outcomes of occluding large left atrial appendages with the LAmbre device
The Official Journal of EuroPCR and the European Association of Percutaneous Cardiovascular Interventions (EAPCI)
- 1 May 2021
- Published by Europa Digital & Publishing in EuroIntervention
- Vol. 17 (1) , 90-92
Open AccessProcedural and short-term outcomes of occluding large left atrial appendages with the LAmbre device
The Official Journal of EuroPCR and the European Association of Percutaneous Cardiovascular Interventions (EAPCI)
- Open Access
- 14 December 2020
- Published by Wiley
A High Precision Algorithm to Classify Left and Right Outflow Tract Ventricular Tachycardia
Several algorithms based on 12-lead ECG measurements have been proposed to identify right ventricular outflow tract (RVOT) and left ventricular outflow tract (LVOT) locations from which ventricular tachycardia (VT) and frequent premature ventricular complex (PVC) originated. However, a clinical-grade artificial intelligence algorithm is not available yet, which can automatically analyze characteristics of 12-lead ECGs and predict RVOT to LVOT origins of VT and PVC. We randomly sampled training, validation, and testing datasets from 420 patients who underwent successful catheter ablation (CA) to treat VT or PVCs, containing (340, 80%), (38, 9%), and (42, 10%) patients, respectively. We iteratively trained an AI algorithm that was supplied with 1,600,800 features extracted from 12-lead ECGs of the patients in the training cohort. The area under the curve (AUC) of the receiver operating characteristic (ROC) curve was calculated from the internal validation dataset to choose an optimal discretization cutoff threshold. After running on the testing dataset, the proposed approach attained the following performance metrics and 95% CIs (confidence intervals), accuracy (ACC) of 97.62 (87.44 -99.99), weighted F1-score of 98.46 (90-100), AUC of 98.99 (96.89-100), sensitivity (SE) of 96.97 (82.54-99.89), and specificity (SP) of 100 (62.97-100). The proposed multi-stage diagnostic scheme attained clinical-grade precision of prediction for LVOT and RVOT locations of VT origin with fewer applicability restrictions than prior studies.
- 14 December 2020
- Published by Wiley
Open AccessA High Precision Algorithm to Classify Left and Right Outflow Tract Ventricular Tachycardia
Several algorithms based on 12-lead ECG measurements have been proposed to identify right ventricular outflow tract (RVOT) and left ventricular outflow tract (LVOT) locations from which ventricular tachycardia (VT) and frequent premature ventricular complex (PVC) originated. However, a clinical-grade artificial intelligence algorithm is not available yet, which can automatically analyze characteristics of 12-lead ECGs and predict RVOT to LVOT origins of VT and PVC. We randomly sampled training, validation, and testing datasets from 420 patients who underwent successful catheter ablation (CA) to treat VT or PVCs, containing (340, 80%), (38, 9%), and (42, 10%) patients, respectively. We iteratively trained an AI algorithm that was supplied with 1,600,800 features extracted from 12-lead ECGs of the patients in the training cohort. The area under the curve (AUC) of the receiver operating characteristic (ROC) curve was calculated from the internal validation dataset to choose an optimal discretization cutoff threshold. After running on the testing dataset, the proposed approach attained the following performance metrics and 95% CIs (confidence intervals), accuracy (ACC) of 97.62 (87.44 -99.99), weighted F1-score of 98.46 (90-100), AUC of 98.99 (96.89-100), sensitivity (SE) of 96.97 (82.54-99.89), and specificity (SP) of 100 (62.97-100). The proposed multi-stage diagnostic scheme attained clinical-grade precision of prediction for LVOT and RVOT locations of VT origin with fewer applicability restrictions than prior studies.